Agilent N4010A WLAN Test Suite

MeasurementEVM11b Class

IEEE 802.11b Error Vector measurement

For a list of all members of this type, see MeasurementEVM11b Members.

System.Object
   Agilent.N4010.WLanTest.Measurements.MeasurementBaseFrontEnd
      Agilent.N4010.WLanTest.Measurements.MeasurementBaseSingleResultFrontEnd
         Agilent.N4010.WLanTest.Measurements.MeasurementEVM11b

[Visual Basic]
Public Class MeasurementEVM11b
    Inherits MeasurementBaseSingleResultFrontEnd
[C#]
public class MeasurementEVM11b : MeasurementBaseSingleResultFrontEnd

Thread Safety

Public static (Shared in Visual Basic) members of this type are safe for multithreaded operations. Instance members are not guaranteed to be thread-safe.

Remarks

PICS Item: HRDS28

EVM (error vector magnitude) is the magnitude of the error vector at the measured chip location. The error vector is the vector difference between the I/Q measured-signal phasor and the I/Q reference-signal phasor.

EVM (%rms) is the Root Mean Square of the error vectors, within the measurement interval, computed and expressed as a percentage of the square root of the mean power of an ideal signal.

The capture length over which EVM is measured is controlled by the instrument's MaxPacketLength and FramesUsed and settings:

Requirements

Namespace: Agilent.N4010.WLanTest.Measurements

Assembly: Agilent.N4010.WLanTest (in Agilent.N4010.WLanTest.dll)

See Also

MeasurementEVM11b Members | Agilent.N4010.WLanTest.Measurements Namespace