Step 3: Set Measurement Parameters
Last updated: April 14, 2005
In this step you set up the measurement parameters which are fixed throughout testing, such as measurement timeouts and trigger arm. It is recommended that you set all measurements to single trigger (see
Trigger Arm (Single or Continuous) Description
).
Click on the "More Details on this Step" icon below for additional information on this step.
You set up other less general measurement parameters while performing specific C.S0033 tests as discussed in
Step 5: INITiate and FETCh Measurements
.
An excerpt from the comprehensive example program is shown below:
1030 !**************************************************************************
1040 ! STEP 3: SET MEASUREMENT PARAMETERS
1050 !**************************************************************************
2780 OUTPUT Testset;"SETUP:CONTINUOUS:OFF"
2790 ! Set all measurements to single trigger
5110 OUTPUT Testset;"SETUP:DOWQ:TIMEOUT 10"
5120 ! Set waveform quality measurement timeout to 10 seconds
5910 OUTPUT Testset;"SETUP:DAPOWER:TIMEOUT 1"
5920 ! Set DIGITAL Power timeout to 1 sec
5930 OUTPUT Testset;"SETUP:CTXS:TIMEOUT 5"
5940 ! Set TX Spur Emissions timeout to 5 sec
7140 OUTPUT Testset;"SETUP:CPOWER:TIMEOUT .5"
7150 ! Set Channel Power timeout to 0.5 sec
8060 OUTPUT Testset;"SETUP:CAPP:TIMEOUT 10"
8070 ! Set access probe power timeout to 10 seconds
9090 OUTPUT Testset;"SETUP:TROP:TIMEOUT 5"
9100 ! Set TROLPC measurement timeout to 5 s
10060 OUTPUT Testset;"SETUP:CPERROR:TIMEOUT 10"
10070 ! Set PER measurement timeout to 10 s
|