Channel Power Measurement Troubleshooting
Last updated: May 24, 2006
If the access probe power measurement is on and waiting to measure a probe, all other measurements (except digital average power) will not be able to execute. It is therefore recommended that you turn off the access probe power measurement before attempting any other measurements.
Troubleshooting Techniques
-
Check to see if an integrity indicator other than 0 (normal) was returned with a measurement result (see
Integrity Indicator
).
If Over Range (5) is returned; the input power has exceeded +37 dBm or is greater than +9 dB above the expected power level of the receiver.
If Under Range (6) is returned; the input power is below -61 dBm or is less than -9 dB below the expected power level of the receiver.
Channel Power Under Range/Over Range Integrity Indicators
|