Agilent Technologies Home Page 8960 Series 10 Wireless Communications Test Set
- +
Home | Product Web Site | Contact Us
+
- -
+ -
8960 1xEV-DO Online User's Guide
E1966A, E1976A, E6706B
Search this guide for     
search tips



Feedback
Did the information on this web page help answer your question or solve a problem?

Yes
No
Did not apply

Please provide additional comments about this page.
(Optional, 255 character max)


Email address (optional):

Privacy Notice: We will not sell or give away your email address to any third party.

 



Spectrum Monitor Troubleshooting

Last updated: December 13, 2002

Possible Setup Issues

During remote operation of the Spectrum Monitor you should configure the trigger arm to single, see SETup:SMONitor .

If the trigger arm is not set to single, the Spectrum Monitor may not return a result. When trigger arm is continuous the Spectrum Monitor rearms itself and starts another measurement cycle, during remote operation the fetch query may not be synchronized to the measurement cycle, see Measurement States .

If trigger source immediate is used for burst modulated signals the results returned will be unreliable. Burst modulated signals should be measured with Trigger Source set to RF Rise or Auto. For RF Rise triggering to work correctly, the signal should be no less than 20 dB below the reference and greater than -20 dBm.

Interpreting Integrity Indicator values

See Integrity Indicator .



Top of pagetop of page     

+ +