How Do I Set Up DUT IP Parameters?
Last updated: April 9, 2009
This section is only applicable to the lab application.
-
Press the
CALL SETUP
key.
-
Press the left
More
key until
3 of 3
is displayed.
-
Press the
Data Channel Info
(
F2
) key.
-
Press the left
More
key until
2 of 2
is displayed.
-
Press the
IP Addressing Info
(
F5
) key.
-
Press the
DUT IP Setup
(
F2
) key.
-
Turn the knob to highlight a parameter and enter/select a value as desired.
For Example: How to configure DUT IPv6 DNS Server IP Address
-
Turn the knob to highlight the
DUT IPv6 DNS Server Prefix, Upper(Hex)
parameter and press the knob, a "A-F" softkeys are shown in the right menu for Hex digit input.
-
Input hex strings with 8 characters, as the picture shows.
-
Set the
DUT IPv6 DNS Server Prefix, Lower(Hex)
,
DUT IPv6 DNS Server IID, Upper(Hex)
,
DUT IPv6 DNS Server IID, Lower(Hex)
to complete the DUT IPv6 DNS Server IP address setting.
DUT IPv6 DNS Server IP address Parameter
When manually inputting the DUT IPv6 DNS Server address and the DUT IPv6 P-CSCF Server address, each address is divided into 4 parts to simplify the input. However when you are using the GPIB command to query or setting the two parameters, each address is set or queried once as a whole.
|