How Do I Perform Multi-unit Synchronization?

Last updated: November 21, 2005

This section is only applicable to the lab application.

Multiple Sync Clients can synchronize to the same Sync Server at the same time. The following procedure provides an example of how to make a synchronization between two test sets. It assumes that the test set running the E6702B cdma2000 lab application acts as a Sync Client and another one running the E6706A 1xEV-DO lab application acts as a Sync Server.

Prior to the following procedure, make sure you have finished the synchronization connection between the two test sets (see Synchronization Connection Setup ).

Step 1: Configure the Sync Server

  1. Press the SYSTEM CONFIG key.
  2. Press the Instrument Setup ( F1 ) key and set up the LAN IP Address as required.
  3. Press the CALL SETUP key.
  4. Verify that the Operating Mode ( F1 ) is set to Active Cell and System Type ( F2 )is set to IS-856 for this example.
  5. Verify the External Trigger is set to 2 s. Go to Call Control 2 of 3, press the Generator Info ( F3 ), and then press the External Trigger ( F4 ). For details, see How Do I Change Generator Information? .
  6. Set the CDMA System Time (Date) and CDMA System Time (Time) as desired. For detailed procedure, see How Do I Change CDMA System Time Parameters? ).

Step 2: Configure the Sync Client

  1. Press the CALL SETUP key.
  2. Verify that the Operating Mode ( F1 ) is set to Active Cell and System Type ( F2 ) is set to IS-2000 for this example.
  3. Make sure that test set is in idle state.
  4. Press the More key until Call Control 2 of 5 is displayed.
  5. Press Cell Info ( F2 ) key.
  6. Press CDMA System Time/Sync Info ( F5 ) key.
  7. Press Clear Sync Indications ( F4 ) to clear the last synchronization information.
  8. Set the Ext 8960 IP Add r ( F2 ) to the same value of the LAN IP Address of the Sync Server that you have set in Step 1: Configure the Sync Server .

Step 3: Perform the synchronization between two test sets

  1. On the test set that acts as Sync Client, press the Sync To External Server ( F3 ) which initiates the synchronization.
  2. The synchronization takes 2-4 seconds. Once the synchronization successfully completes, the CDMA system time on the sync client is same as that of the sync server. Observe the CDMA System Time and Synchronization Information on both test sets, a typical display on the Sync Client and the Sync Server is shown below.