FPC Mode under different RCs
When the
Forward Power Control
(
F2
) on the
Call Control 4 of 4
menu is selected, all of the parameters and operations relating to forward power control (FPC) are available from the FPC Info menu.
FPC Mode
(
F3
) is used to specify whether the FPC operation is enabled on the test set and what kind of FPC operating mode will be used on the mobile station (MS). When a call is connected in IS-2000 system with radio configuration (Fwd3, Rvs3), (Fwd4, Rvs3) and (Fwd5, Rvs4) (see
Radio Configuration
), the test set informs the MS of the FPC mode to be used via the FPC_MODE field in the
Power Control Message
. For more information on how the FPC works, see
Forward Power Control Overview
.
FPC Mode
FPC Mode
|
Description
|
Ignore FPC Bits
|
When selected, the test set will ignore the Power Control Bits (PCBs) on the reverse power control sub-channel no matter what is sent from the MS.
|
000 - Primary 800
|
When selected, the test set sets the MS to use FPC mode '000' as specified in the standard. In this mode, the MS transmits the primary reverse power control subchannel at 800 bps data rate and the test set will adjust the forward fundamental channel level according to the PCBs received on the reverse power control sub-channel in the specified power step size.
|
011 - EIB
|
When selected, the test set sets the MS to use FPC mode '011' as specified in the standard. In this mode, the MS transmits the Erasure Indicator Bits (EIBs) on the reverse power control sub-channel. The test set checks the received EIBs on the reverse power control sub-channel against the corresponding frames sent from the test set, and keeps a record of matched/not matched/not received frames in the EIB counters.
|
GPIB Commands:
CALL[:CELL]:FPControl[:NORMal]:MODE
This parameter specifies FPC Mode under radio configuration (fwd11, rev8). (see
Radio Configuration
).When a call is connected in IS-2000 system with radio configuration (Fwd11, Rvs8) (see
Radio Configuration
), the test set informs the MS of the FPC mode to be used via the FPC_MODE field in the
Power Control Message
. For more information on how the FPC works, see
Forward Power Control Overview
.
FPC Mode <RC6>
FPC Mode
|
Description
|
Ignore FPC Bits
|
When selected, the test set will ignore the Power Control Bits (PCBs) on the reverse power control sub-channel no matter what is sent from the MS.
|
000 - Primary 400
|
When selected, the test set sets the MS to use FPC mode '000' as specified in the standard. In this mode, the MS transmits the primary reverse power control bits in PCGs 1, 3, 5, 7, 9, 11, 13 and 15 at 400 bps data rate and the test set will adjust the forward fundamental channel level according to the PCBs received on the reverse power control sub-channel in the specified power step size.
|
011 - Primary 200
|
When selected, the test set sets the MS to use FPC mode '011' as specified in the standard. In this mode, the MS transmits the primary reverse power control bits in PCGs 3, 7, 11 and 15 at 200 bps data rate and the test set will adjust the forward fundamental channel level according to the PCBs received on the reverse power control sub-channel in the specified power step size.
|
GPIB Commands:
CALL[:CELL]:FPControl:SLOW:MODE