How Do I Perform MEID Verification?
Last updated: June 20, 2006
In the
Active Cell
operating mode and
IS-2000
system, the test set supports to test the MEID-equipped mobile stations (MS). This example is to verify that an MEID-equipped MS can:
-
correctly tell the test set that it supports MEID (via
Registration Message
with MEID support indicator set to 1 in the Station Class Mask (SCM) field).
-
successfully respond to an MEID information request from an MEID-enabled test set. (Setting the
MEID Support
to on will enable the test set to support MEID. The test set will sends the
Status Request Message
to retrieve the MEID value from the MEID-equipped MS.)
-
successfully establish a call with an MEID-enabled test set using the public long code mask (PLCM) determined by the specified
PLCM Type
.
-
successfully perform an implicit handoff by changing the
PLCM Type
.
-
successfully establish a call with an MEID-disabled test set using the public long code mask (PLCM) derived from the pseudo-ESN. (Setting the
MEID Support
to off will disable the test set to support the MEID. The test set does not sends the
Status Request Message
to retrieve the MEID value from the MEID-equipped MS. The MEID-equipped MS only registers to the test set using pseudo-ESN so that PLCM for a call can only derive from the pseudo-ESN.)
Step A: Configure Call Parameters
-
Press the
Call Setup
key.
-
Verify that the
Operating Mode
(
F1
) is set to
Active Cell
and the
System Type
(
F2
) is set to
IS-2000
.
-
Set the parameters that are required for your MS to acquire the IS-2000 system. For example,
-
Set the
Cell Band
(
F8
) on
Call Parms 1 of 4
) and
Channel
(
F9
on
Call Parms 1 of 4
) to match your MS's configuration for IS-2000 system.
-
Set the
SID
,
NID
as required for your MS from the Cell Parameters menu (see
B. Cell Parameters
for detailed procedure).
Step B: Enable Test Set to Support MEID/Enhanced PLCM
-
Press the left
More
key until you reach the
Call Control 3 of 3
(in Test Application) or the
Call Control 3 of 5
(in Lab Application).
-
Select the
MEID/Enhanced PLCM
(
F3
).
-
Set the
MEID Support
(
F1
) to
On
. This is to enable the test set to support MEID tests.
-
Set the
PLCM Type
(
F2
) to
MEID Based
.
Step C: Register the Mobile Station.
-
If there is data shown on Mobile Station Information screen, clear these data by pressing the
F5
(
Clear MS & Capability Info
) on
Call Control 2 of 3
(in Test Application) or
Call Control 2 of 5
(in Lab Application).
-
Power on the MS and wait for it to perform a power-up registration.
-
When a registration has completed, the registration data (including MEID related information) reported/retrieved from the MS will be displayed on the
Mobile Reported Information
screen. During registration, the MS sends a
Registration Message
with SCM bit 4 (MEID support indicator) set to 1 (indicating that it supports MEID). Then the test set sends a
Status Request Message
with RECORD_TYPE = '00100111' to request MEID value. Verify that the MS responds with
Status Response Message
with RECORD_TYPE = '00100111', and set MEID_LEN and MEID fields properly.
Step D: Establish a Call.
-
If required, set the service option for the radio configuration (for example,
SO2
for
Radio Config (Fwd3, Rvs3)
) in which you'd like to bring up the call.
-
Select
Originate Call
(
F3
) on
MEID/PLCM 2 of 2
or make a MS originated call (by entering a phone number and pressing SEND on the MS).
-
Check for
Connected
in the
Active Cell
status area indicating a call is successfully established. The test set sends
MEID Extended Channel Assignment Message
with PLCM_TYPE set to '0100' (corresponding to
MEID Based
you've set during Step B). When the call is successfully established, it indicates that the MS can use the PLCM determined by the specified
PLCM Type
.
Step E: Perform an Implicit Handoff.
-
Change the
PLCM Type
(
F2
) you've set in Step B to a different value (for example,
ESN Based
).
-
Observe the
Active Cell
status area. You should see the status changes from
Connected
to
Handoff
to
Connected
. The test set sends
MEID Extended Universal Handoff Direction Message
with PLCM_TYPE set to '0000' (corresponding to
ESN Based
you've changed to). The call is maintained, which indicates that the MS can use the PLCM determined by the specified
PLCM Type
.
Step F: Disable the Test Set to Support MEID and Verify the MS Behavior.
-
End the call.
-
Change the
MEID Support
(
F1
) to
Off
.This is to disable the test set to support MEID. You should see that
PLCM Type
(
F2
) and
Query MEID
(
F3
) become invalid.
-
Clear the registration data on Mobile Station Information screen by pressing the
F5
(
Clear MS & Capability Info
) on
Call Control 2 of 3
(in Test Application) or
Call Control 2 of 5
(in Lab Application).
-
Wait for the MS making a timer-based registration or press the
Register Mobile
(
F6
) on
Call Control 1 of 3
(in Test Application) or
Call Control 1of 5
(in Lab Application) to force MS making a registration.
-
When a registration has completed, you should see that no MEID value is displayed on the
Mobile Reported Information
screen although the MS reports that it supports MEID. This is because the test set does not send the
Status Request Message
to an MEID-equipped MS to retrieve its MEID value. Only pseudo-ESN is reported during MS registration. The MS sends a
Registration Message
with SCM bit 4 (MEID support indicator) set to 1 (indicating that it supports MEID), but no
Status Request Message
with RECORD_TYPE = '00100111' (MEID) is sent by the test set to retrieve MEID value.
-
Establish a call and verify that a call is connected. This indicates that an MEID-equipped MS can establish a call with an MEID-disabled test set but use the PLCM derived from the pseudo-ESN instead of the specified PLCM Type. The test set sends
Extended Channel Assignment Message
(without any PLCM information included) during call establishment.
Step H: End the Test.
-
If desired, end the call.
-
Verify that the
MEID Support
(
F1
) is set to
Off
if you plan to test an ESN-equipped MS. Otherwise, the call can not be established.
|