When a measurement is started using INITiate commands, a new measurement cycle is started. If the selected measurement is currently in a measurement cycle, it is aborted. If a timeout is specified, the timeout period begins when a measurement is initiated.
NOTE |
|
The INITiate command is used to start measurements.
INITiate commands allow multiple measurements to be started without waiting for other measurement processes to complete. For example, the following program example starts the Digital Average Power and Waveform Quality measurements, and then uses the INITiate:DONE? command in a loop to query the completion status of these measurements. See Measurement Event Synchronization .
As each measurements completed, the FETCh command can be used to acquire the results, and the results entered into variables in the controlling application.
This program ends when the INITiate:DONE? command returns the string "NONE" indicating that all initiated measurements have gone through the measuring state. See Measurement States .
NOTE |
|
NOTE |
|
10 OUTPUT 714;"SETup:ALL:CONTinuous:OFF" 20 ! select single trigger mode for all measurements. 30 OUTPUT 714;"INITiate:DAPower;WQUality" 40 ! start digital average power and waveform quality measurements. 50 LOOP 60 OUTPUT 714;"INITiate:DONE?" 70 ! query to find out if any measurements are done. 80 ENTER 714;Meas_complete$ 90 SELECT Meas_complete$ 100 CASE "DAP" 110 ! test for the string "DAP" indicating 120 ! digital average power measurement is done. 130 OUTPUT 714;"FETCh:DAPower:POWer?" 140 ! query digital average power measurement. 150 ENTER 714;Dapower 160 CASE "WQU" 170 ! test for the string "WQU" indicating 180 ! that the waveform quality measurement is done. 190 OUTPUT 714;"FETCh:WQU:RHO?" 200 ! query rho. 210 ENTER 714;Rho 220 END SELECT 230 EXIT IF Meas_complete$="NONE" 240 END LOOP 250 PRINT Dapower 260 PRINT Rho 270 END
|
|