SETup:CFDTune
Last updated: June 20, 2007
GPIB Command Conventions
SETup:CFDTune:FREQuency:STEP:COUNt
|
Function
|
Sets/queries the
Number of Frequency Steps
.
|
|
Setting
|
Range: 1 to 20
Resolution: 1
|
|
Query
|
Range: 1 to 20
|
|
*RST Setting
|
5
|
|
Requirement
|
Test Application Revision: B.11 and above, with the required feature license
Lab Application Revision: B.04 and above
|
|
Programming Example
OUTPUT 714;"SETup:CDFTune:FREQuency:STEP:COUNt 10"
!Sets the number of frequency steps to 10.
|
SETup:CFDTune:MS:RX:FREQuency:STEP[:VALues]
|
Function
|
Sets/queries the
MS RX Frequency Step Values
.
If you set fewer than 40 values, the remaining values are set to NAN (9.91E+37).
|
|
Setting
|
Range:
Although you can enter any frequencies between 421 and 2170 MHz, when running the Fast Device Tune measurement in the cdma2000 application/format, you can only test frequencies within a single cdma2000 frequency band range:
-
420 to 493.48 MHz (Band: NMT-450)
-
832 to 960 MHz (Bands: US Cellular, Japan CDMA, China Cellular, and Secondary 800)
-
1840 to 1870 MHz (Band: Korean PCS)
-
2110 to 2170 MHz (Band: IMT-2000)
-
1930 to 1995 MHz (Bands: US PCS and US PCS 1900)
-
2110 to 2155 MHz (Band: AWS)
Resolution: 1 Hz
|
|
Query
|
Same as setting |
|
*RST Setting
|
872.0MHz, 872.2MHz, 872.4MHz, 872.6MHz, 872.8MHz, 873.0MHz, 873.2MHz, 873.4MHz, 873.6MHz, 873.8MHz, 874.0MHz, 874.2MHz, 874.4MHz, 874.6MHz, 874.8MHz, 875.0MHz, 875.2MHz, 875.4MHz, 875.4MHz, 875.6MHz, 875.8MHz, 1933.0MHz, 1933.2MHz, 1933.4MHz, 1933.6MHz, 1933.8MHz, 1934.0MHz, 1934.2MHz, 1934.4MHz, 1934.6MHz, 1934.8MHz, 1935.0MHz, 1935.2MHz, 1935.4MHz, 1935.6MHz, 1935.8MHz, 1936.0MHz, 1936.2MHz, 1936.4MHz, 1936.6MHz, 1936.8MHz
|
|
Requirement
|
Test Application Revision: B.11 and above, with the required feature license
Lab Application Revision: B.04 and above
|
|
Programming Example
OUTPUT 714;"SETup:CFDTune:MS:RX:FREQuency:STEP 1932 MHz, 1933 MHz, 1934 MHz, 1935 MHz"
! Sets MS RX Frequency Step Values to 1932 MHz, 1933 MHz, 1934 MHz, 1935 MHz.
|
SETup:CFDTune:MS:RX:FREQuency:STEP:STARt:[INDex]
|
Function
|
Sets/queries the
MS RX Frequency Step Start Index
. |
|
Setting
|
Range: 0 to 39
Resolution: 1
|
|
Query
|
Same as setting |
|
*RST Setting
|
0
|
|
Requirement
|
Test Application Revision: B.11 and above, with the required feature license
Lab Application Revision: B.04 and above
|
|
Programming Example
OUTPUT 714;"SETup:CDFTune:MS:RX:FREQuency:STEP:STARt 6"
!Sets the MS RX Frequency Step Start Index to 6.
|
SETup:CFDTunee:MS:RX:POWer:STEP[:LEVels]
|
Function
|
Sets/queries the
MS RX Power Step Levels
.
If you set fewer than 40 values, the remaining values are set to NAN (9.91E+37).
|
|
Setting
|
Range:
Although you can enter any power levels between -165 and 37 dBm, when running the Fast Device Tune measurement in the cdma2000 application/format, the test set can only transmit test power levels in the range of -115 to -13 dBm (
RF IN/OUT
) or -115 to -5 dBm (
RF OUT ONLY
). These ranges assume an amplitude offset of 0; the allowable ranges change as you change amplitude offsets,, see
Cell Power, AWGN Power and Total RF Power Ranges
).
Resolution: 0.01
|
|
Query
|
Same as setting |
|
*RST Setting
|
-15.0dBm, -20.0dBm, -25.0dBm, -30.0dBm, -35.0dBm, -40.0dBm, -45.0dBm, -50.0dBm, -55.0dBm, -60.0dBm, -65.0dBm, -70.0dBm, -75.0dBm, -80.0dBm, -85.0dBm, -85.0dBm, -85.0dBm, -85.0dBm, -85.0dBm, -85.0dBm, -85.0dBm, -85.0dBm, -85.0dBm, -85.0dBm, -85.0dBm, -85.0dBm, -85.0dBm, -85.0dBm, -85.0dBm, -85.0dBm, -85.0dBm, -85.0dBm, -85.0dBm, -85.0dBm, -85.0dBm, -85.0dBm, -85.0dBm, -85.0dBm, -85.0dBm, -85.0dBm |
|
Requirement
|
Test Application Revision: B.11 and above, with the required feature license
Lab Application Revision: B.04 and above
Test Application Revision B.12 and Lab Application B.05, added support for RF OUT ONLY port.
|
|
Programming Example
OUTPUT 714;"SETup:CFDTune:MS:RX:POWer:STEP -16, -20, -30, -40, -50,-60, -70, -80, -90, -100"
! Sets MS RX Power Step Levels to -16, -20, -30, -40, -50,-60, -70, -80, -90, -100 dBm.
|
SETup:CFDTune:MS:RX:POWer:STEP:STARt:[INDex]
|
Function
|
Sets/queries the
MS RX Power Step Start Index
. |
|
Setting
|
Range: 0 to 39
Resolution: 1
|
|
Query
|
Same as setting |
|
*RST Setting
|
0
|
|
Requirement
|
Test Application Revision: B.11 and above, with the required feature license
Lab Application Revision: B.04 and above
|
|
Programming Example
OUTPUT 714;"SETup:CDFTune:MS:RX:POWer:STEP:STARt 6"
!Sets the MS RX Power Step Start Index to 6.
|
SETup:CFDTune:MS:TX:FREQuency:STEP[:VALues]
|
Function
|
Sets/queries the
MS TX Frequency Step Values
.
If you set fewer than 40 values, the remaining values are set to NAN (9.91E+37).
|
|
Setting
|
Range:
Although you can enter any frequencies between 411 and 1980 MHz, when running the Fast Device Tune measurement in the cdma2000 application/format, you can only test frequencies within a single cdma2000 frequency band range:
-
410 to 483.48 MHz (Band: NMT-450)
-
806 to 925 MHz (Bands: US Cellular, Japan CDMA, China Cellular, and Secondary 800)
-
1750 to 1780 MHz (Band: Korean PCS)
-
1920 to 1980 MHz (Band: IMT-2000)
-
1850 to 1915 MHz (Bands: US PCS and US PCS 1900)
-
1710 to 1755 MHz (Band: AWS)
Resolution: 1 Hz
|
|
Query
|
Same as setting |
|
*RST Setting
|
827.0MHz, 827.2MHz, 827.4MHz, 827.6MHz, 827.8MHz, 828.0MHz, 828.2MHz, 828.4MHz, 828.6MHz, 828.8MHz, 829.0MHz, 829.2MHz, 829.4MHz, 829.6MHz, 829.8MHz, 830.0MHz, 830.2MHz, 830.4MHz, 830.6MHz, 830.8MHz, 1853.0MHz, 1853.2MHz, 1853.4MHz, 1853.6MHz, 1853.8MHz, 1854.0MHz, 1854.2MHz, 1854.4MHz, 1854.6MHz, 1854.8MHz, 1855.0MHz, 1855.2MHz, 1855.4MHz, 1855.6MHz, 1855.8MHz, 1856.0MHz, 1856.2MHz, 1856.4MHz, 1856.6MHz, 1856.8MHz
|
|
Requirement
|
Test Application Revision: B.11 and above, with the required feature license
Lab Application Revision: B.04 and above
|
|
Programming Example
OUTPUT 714;"SETup:CFDTune:MS:TX:FREQuency:STEP 1852 MHz, 1853 MHz, 1854 MHz, 1855 MHz"
! Sets MS TX Frequency Step Values to 1852 MHz, 1853 MHz, 1854 MHz, 1855 MHz.
|
SETup:CFDTune:MS:TX:FREQuency:STEP:STARt:[INDex]
|
Function
|
Sets/queries the
MS TX Frequency Step Start Index
. |
|
Setting
|
Range: 0 to 39
Resolution: 1
|
|
Query
|
Same as setting |
|
*RST Setting
|
0
|
|
Requirement
|
Test Application Revision: B.11 and above, with the required feature license
Lab Application Revision: B.04 and above
|
|
Programming Example
OUTPUT 714;"SETup:CDFTune:MS:TX:FREQuency:STEP:STARt 6"
!Sets the MS TX Frequency Step Start Index to 6.
|
SETup:CFDTune:MS:TX:POWer:STEP[:LEVels]
|
Function
|
Sets/queries the
MS TX Power Step Levels
.
If you set fewer than 40 values, the remaining values are set to NAN (9.91E+37).
|
|
Setting
|
Range:
Although you can enter any power levels between -129 and +80 dBm (because this measurement is common to the W-CDMA/HSDPA, 1xEV-DO and cdma2000 formats), when running the Fast Device Tune measurement in the cdma2000 application/format, you can only measure power levels in the range of -79 to +30 dBm and the measurement is only specified for power levels in the range of -61 to +30 dBm (these ranges assume an amplitude offset of 0; the ranges change as you change amplitude offsets, see
Amplitude Offset
).
Resolution: 0.01
|
|
Query
|
Same as setting |
|
*RST Setting
|
12.0dBm, 7.0dBm, 2.0dBm, -8.0dBm, -13.0dBm, -18.0dBm, -23.0dBm, -28.0dBm, -33.0dBm, -38.0dBm, -43.0dBm, -48.0dBm, -53.0dBm, -58.0dBm, -63.0dBm, -63.0dBm, -63.0dBm, -63.0dBm, -63.0dBm, -63.0dBm, -63.0dBm, -63.0dBm, -63.0dBm, -63.0dBm, -63.0dBm, -63.0dBm, -63.0dBm, -63.0dBm, -63.0dBm, -63.0dBm, -63.0dBm, -63.0dBm, -63.0dBm, -63.0dBm, -63.0dBm, -63.0dBm, -63.0dBm, -63.0dBm, -63.0dBm, -63.0dBm |
|
Requirement
|
Test Application Revision: B.11 and above, with the required feature license
Lab Application Revision: B.04 and above
|
|
Programming Example
OUTPUT 714;"SETup:CDFTune:MS:RX:POWer:STEP 20, 16, 12, 8, 4, 0, -4, -8, -12, -16, -20, -30, -40, -50"
!Sets Rx Power Step Sequency values to 20, 16, 12, 8, 4, 0, -4, -8, -12, -16, -20, -30, -40, -50.
|
SETup:CFDTune:MS:TX:POWer:STEP:STARt:[INDex]
|
Function
|
Sets/queries the
MS TX Power Step Start Index
. |
|
Setting
|
Range: 0 to 39
Resolution: 1
|
|
Query
|
Same as setting |
|
*RST Setting
|
0
|
|
Requirement
|
Test Application Revision: B.11 and above, with the required feature license
Lab Application Revision: B.04 and above
|
|
Programming Example
OUTPUT 714;"SETup:CDFTune:MS:TX:POWer:STEP:STARt 6"
!Sets the MS TX Power Step Start Index to 6.
|
SETup:CFDTune:POWer:STEP:COUNt
|
Function
|
Sets/queries the
Number of Power Steps
.
|
|
Setting
|
Range: 1 to 20
Resolution: 1
|
|
Query
|
Range: 1 to 20
|
|
*RST Setting
|
10
|
|
Requirement
|
Test Application Revision: B.11 and above, with the required feature license
Lab Application Revision: B.04 and above
|
|
Programming Example
OUTPUT 714;"SETup:CDFTune:POWer:STEP:COUNt 15"
!Sets the number of power steps to 15.
|
SETup:CFDTune:TIMeout[:STIMe]
|
Function
|
The setting form of this command performs two functions: (1) sets the Fast Device Tune measurement's timeout (see
Timeouts
) value, and (2) sets the Fast Device Tune measurement's timeout state to 1 (on).
The query form returns the Fast Device Tune measurement's timeout value.
See
Fast Device Tune Measurement Parameters
.
|
|
Setting
|
Range: 0.1 to 999 seconds
Resolution: 0.1 second
|
|
Query
|
Range: 0.1 to 999
|
|
*RST Setting
|
10 seconds, off
|
|
Requirement
|
Test Application Revision: B.11 and above, with the required feature license
Lab Application Revision: B.04 and above
|
|
Programming Example
OUTPUT 714;"SETup:CDFTune:TIMeout 5S"
!Sets timeout state to ON and sets the timeout time to 5 seconds.
|
SETup:CFDTune:TIMeout:STATe
|
Function
|
Sets/queries Fast Device Tune measurement's timeout state (see
Timeouts
). |
|
Setting
|
Range: 1|ON|0|OFF
|
|
Query
|
Range: 1|0
|
|
*RST Setting
|
0 (off)
|
|
Requirement
|
Test Application Revision: B.11 and above, with the required feature license
Lab Application Revision: B.04 and above
|
|
Programming Example
OUTPUT 714;"SETup:CDFTune:TIMeout:STATe ON"
|
SETup:CFDTune:TIMeout:TIME
|
Function
|
Sets/queries Fast Device Tune measurement's timeout (see
Timeouts
) setting to be used when the timeout state is set to ON. |
|
Setting
|
Range: 0.1 to 999 seconds
Resolution: 0.1 second
|
|
Query
|
Range: 0.1 to 999
|
|
*RST Setting
|
10 seconds
|
|
Requirement
|
Test Application Revision: B.11 and above, with the required feature license
Lab Application Revision: B.04 and above
|
|
Programming Example
OUTPUT 714;"SETup:CDFTune:TIMeout:TIME 5S"
|
|
|