Frequency Modulation (FM) Measurement Description

Last updated: August 12, 2005

Description

How is a frequency modulation measurement made?

The Analog Frequency Modulation measurement measures the frequency deviation and, if enabled, the total harmonic distortion and modulation frequency, for frequency modulated analog signals (such as AMPS transmitters). Measurements are performed after any filtering, de-emphasis, and expansion are performed (selected by the user in the setup parameters).

The FM deviation and distortion measurements are not affected by a static (constant) frequency error on the input signal as long as the input signal is within 2.5 kHz of the expected frequency.

This measurement is not used for digital signals.

Operating Consideration

The distortion value is not returned if the Distortion State is set to OFF. The modulation frequency value is not returned if the Modulation Frequency State is set to OFF.

Triggering is always immediate for this measurement. Trigger Delay and other trigger sources are not used.

When manually controlling the test set from the front panel, all measurements default to using continuous triggering. If you make digital measurements on a DTC using RF Rise or Protocol triggering, and then hand off to an AVC and select an analog measurement, the analog measurement may not trigger and display a result. This is because the digital measurement is still waiting for a burst transmission to trigger it (which it won't get from an analog signal) and therefore the measurements "freeze". To prevent this from happening when making analog measurements, either turn off the digital measurements or set their trigger source to Immediate.

Frequency Modulation Measurement Parameters

  • Multi-Measurement Count (See Statistical Measurement Results ) - Select On to enable averaging of multiple signaling tone measurement results.
  • Distortion State (See SETup:FM:DISTortion:STATe ) - Select On to enable the Distortion State.
  • Distortion Fundamental Frequency (See SETup:FM:DISTortion:FREQuency ) - The Distortion Fundamental Frequency value specifies the demodulated audio frequency to use when making a distortion measurement, and must be entered to make a valid measurement. However, specifying the fundamental frequency does not affect the measurement state.
  • Modulation Frequency State (See SETup:FM:FREQuency:STATe ) - The modulation frequency measurement is enabled and disabled using the Mod Freq State setting. The modulation frequency measurement is made after any filtering is applied, and can be used to make measurements such as SAT frequency.
  • Filter Type ( SETup:FM:FILTer[:TYPE] ) - The following filter are avaialble to make measurements on the demodulated FM (baseband) signal.
    • None - no filtering is provided (default)
    • 50 Hz to 15 kHz
    • 300 Hz to 15 kHz
    • 100 Hz BW Band Pass Filter (Tunable)
    • C-Message
    • 50 Hz to 3 kHz
  • 100 Hz BW BPF Center Frequency ( SETup:FM:FILTer:TBPass[:FREQuency] ) - The setting is available when this filter is selected, and can be set in the range of 300 Hz to 15 kHz
  • Detector Type ( SETup:FM:DETector[:TYPE] ) - Select the type of detector to use when making FM deviation measurements: RMS, Peak+, Peak-, Peak +/- Max and Peak +/- /2.
    • The Peak+ and Peak- detectors measure the greatest positive or negative frequency excursion in carrier frequency during the sample.
    • The RMS detector measures the root mean squared value. This setting does not affect the distortion measurement.
    • The Peak+/- Max detector measures both greatest positive and negative peaks and take the absolute value of each, then return the greater one.
    • The Peak+/- /2 detector measures both greatest positive and negative peaks and take the absolute value of each, then returns average value.