Programming a Waveform Quality/Code Domain Measurement
Last updated: October 28, 2004
This section provides an example of how to make a waveform quality measurement via GPIB.
The following program example assumes that the mobile station's antenna output is connected to the RF IN/OUT connector, and the mobile station has found service.
The basic procedure to measure waveform quality is as follows:
-
Make a call.
-
Configure the waveform quality measurement setup parameters using the
SETup:WQUality
subsystem.
-
Start the waveform quality measurement using the
INITiate
subsystem.
-
Use the
FETCh:WQUality
command to obtain waveform quality measurement results.
-
Use the
FETCh:WQUality:CDPower:ICHannel[:ALL]?
or
FETCh:WQUality:CDPower:QCHannel[:ALL]?
commands to obtain code domain power measurement results.
Programming Example
10 RE-SAVE "c:\roadhog\programs\wqu.txt"
20 !This program measures waveform quality
30 !Refer to "Waveform Quality and Frequency Accuracy" in EIA/TIA-IS-98D for standards and
specifications
40 CLEAR SCREEN
50 A=714
60 !***Specify call parameters***
70 Systype$="DIGITAL2000" !Cell System Type
80 Band$="USCELLULAR" !Cell band
90 Channel=384 !RF Channel
100 Sid=1 !System Identification
110 Nid=1 !Network Identification
120 Radio_config$="F3R3" !Radio Configuration
130 Service_opt$="SO2" !Loopback service option
140 !
150 !***Specify test parameters***
160 Loss_frequency$="851 MHZ,896 MHZ" !Reverse channel, forward channel frequencies for
amplitude offset (path loss)
170 Expected_loss$="-2,-2" !Path loss at reverse channel, forward channel frequencies
180 Ior=-50!Cell power
190 Pilot=-7 !Pilot channel power level
200 Traffic=-7.4 !Traffic channel power level
210 !
220 ON TIMEOUT 7,20 CALL Timeout ! Calls "Timeout" routine
230 OUTPUT A;"SETUP:WQU:TIMEOUT:STIME 10 S" !Sets waveform quality measurement timeout to 10
seconds
240 OUTPUT A;"SYST:COMM:GPIB:DEB:STAT ON"!Turns debug state on
250 !
260 !***Set up call parameters***
270 OUTPUT A;"CALL:OPER:MODE CALL" !Sets Operating Mode to Active Cell
280 OUTPUT A;"CALL:SYSTEM ";Systype$ !Sets System Type
290 OUTPUT A;"CALL:BAND ";Band$ !Sets Cell Band for selected System Type
300 OUTPUT A;"CALL:CHAN ";Channel !Sets RF Channel for selected System Type and Cell Band
310 OUTPUT A;"CALL:POW -35" !Sets and turns on Cell Power
320 OUTPUT A;"CALL:SID ";Sid! Sets System Identification
330 OUTPUT A;"CALL:NID ";Nid! Sets Network Identification
340 OUTPUT A;"CALL:RCON ";Radio_config$!Sets Radio configuration
350 OUTPUT A;"CALL:SOPTION ";Service_opt$ !Sets service option for selected System Type and
Radio Configuration
360 OUTPUT A;"CALL:PROT PREV6" !Sets Protocol Revision for the selected System Type
370 OUTPUT A;"CALL:PAG:DRAT FULL" !Sets Paging data rate to full rate
380 OUTPUT A;"CALL:PIL -7" !Sets pilot to default
390 OUTPUT A;"CALL:SYNC -16" !Sets sync to default
400 OUTPUT A;"CALL:PAG -12" !Sets paging to default
410 OUTPUT A;"CALL:FCH -15.6" !Sets traffic to default
420 !
430 PRINT "Turn on mobile station...press continue when it has found service and registered"
440 PAUSE
450 CLEAR SCREEN
460 OUTPUT A;"CALL:REGISTER:TIMER:STATE OFF" !Turns off timer-based registrations
470 WAIT 2 ! Waits for any timer-based registrations to clear out
480 !
490 !***Page the mobile station***
500 DISP "Paging mobile station..."
510 OUTPUT A;"CALL:ORIG" !Pages the mobile station520 OUTPUT A;"CALL:CONNECTED:STATE?" !Queries for connected/idle state...hangs until state change detector is disarmed
530 ENTER A;Call_connected
540 IF NOT Call_connected THEN
550 DISP "Call attempt failed"
560 STOP
570 ELSE
580 DISP "Call connected"
590 END IF
600 !
610 !***Set up measurement parameters***
620 OUTPUT A;"SETUP:WQUALITY:CONTINUOUS OFF" !Sets trigger arm state to single
630 OUTPUT A;"CALL:POW ";Ior !Sets Cell Power for the selected System Type
640 OUTPUT A;"CALL:CLPCONTROL:REVERSE:MODE ALT" !Alternating up/down power control bits
650 OUTPUT A;"CALL:PIL ";Pilot !Sets Pilot channel level
660 OUTPUT A;"CALL:FCH ";Traffic !Sets Fundamental (Traffic) channel level
661 OUTPUT A;"SETUP:WQUALITY:CDPOWER:IQINactive:LIMIT -23"
670 !
680 !***Obtain and display waveform quality measurement results***
690 WAIT 1
700 OUTPUT A;"INITIATE:WQUALITY" !Initiates a Waveform Quality measurement
760 OUTPUT A;"FETCh:WQUALITY?" !Queries the test set for waveform quality measurement
results
770 ENTER A;Integrity,Rho,Freq_error,Time_error,Carr_feed,Phase_err,Mag_err,Evm
780 PRINT "Integrity indicator = ";Integrity
790 PRINT "Rho = ";Rho,"Frequency error = ";Freq_error,"Time error = ";Time_error,"Carrier
feedthrough = ";Carr_feed
800 PRINT "Phase error = ";Phase_err,"Magnitude error = ";Mag_err,"Error vector magnitude =
";Evm
801 !
802 !***Obtain and display code domain power pass/fail measurement results***
803 OUTPUT A;"FETCh:WQUality:CDPower:IQINactive?"
804 ENTER A;Iq_passfail
805 SELECT Iq_passfail
806 CASE 0
807 PRINT "Code domain power passed"
808 CASE 1
809 PRINT "Code domain power failed"
811 END SELECT
813 !***Obtain and display code domain power measurement results (I-channel)***
814 DIM Ichannel$[1000]
815 OUTPUT A;"FETCh:WQUALITY:CDPower:ICH?"
816 ENTER A;Ichannel$
830 INTEGER I,Comma,Nextcomma !I represents the 16 bins, Comma is the current substring position
840 Comma=1 !Initialize Comma to the first character of the Ichannel$ ascii character string
860 PRINT "I-Channel active/inactive = ";Ichannel$[Comma;16]
870 Comma=Comma+17
890 PRINT "I-channel bin 0 walsh channel number = ";Ichannel$[Comma;16]
891 Comma=Comma+17
900 PRINT "I-channel bin 0 spread factor = ";Ichannel$[Comma;16]
905 Comma=Comma+17
910 PRINT "I-channel bin 0 power level = ";Ichannel$[Comma;16]
920 DISP "Test complete"
930 !
940 !***Return to default call processing settings***
950 OUTPUT A;"CALL:CLPCONTROL:REVERSE:MODE ACTIVE" !Return to active power control bits
960 OUTPUT A;"CALL:POW:DIG2000 -50" ! Returns cell power to -50 dBm/1.23 MHz
970 OUTPUT A;"CALL:END"
980 END
990 Timeout: SUB Timeout ! Global timeout handler (from Step 1)
1000 PRINT "Program timed out"
1010 CLEAR 7
1020 CLEAR 714