Gated Power Measurement Troubleshooting

Last updated: October 28, 2004

Possible Setup Issues

If the Access Probe Power measurement is currently selected and the Trigger Arm setting is Continuous, all other measurements are disabled. Use the SETup:CAPPower:CONTinuous command to set the trigger arm state to single and retry this measurement.

During remote operation of this measurement, you should configure the Trigger Arm to Single. See SETup:GPOWer:CONTinuous . If single triggering is used during manual operation, press the START SINGLE key to trigger each measurement.

The test set receiver's expected signal amplitude and channel number (or frequency) must be set up correctly for accurate measurements. This is normally done automatically (see Active Cell Operating Mode ) but can also be done manually (see Receiver Control ) .

The following settings are required for the gated power measurement:

  • Operating Mode must be Active Cell
  • System Type must be IS-95 or IS-2000 with Radio Config set to (Fwd1, Rvs1) or (Fwd2, Rvs2)
  • FCH Service Option Setup must be set to SO2 or SO9 (loopback service option)
  • Traffic Data Rate must be set to Eighth.

The input signal must conform to Input Signal Requirements .