|
|
|
|
GPRS Mobile Measurements
Last updated: May 8, 2002
This section is
not
applicable to GSM.
The following step-by-step procedures explain how to perform specific measurements available on the GPRS Test Application. A detailed description of each measurement is given in the
Reference
information on the documentation CD-ROM and at the Agilent 8960 support website (http://www.agilent.com/find/8960support/).
Setup parameters for each measurement are assumed to be set to their default (Full preset) values unless otherwise stated.
The following measurement procedures are provided:
Measuring Transmit Power
-
Connect the mobile.
-
Turn the mobile on and wait for
Attached
in the
Active Cell:
field.
Note: For mobiles that don't perform a GPRS attach automatically, set the mobile to data mode.
-
Press the
Start Data Connection
(
F3
) key and watch for the
Active Cell:
field changing to
Transferring
.
-
Press the
Measurement selection
key.
-
Select the
Transmit Power
measurement.
-
Press the
Transmit Power Setup
(
F1
) key.
-
Set the measurement parameters as needed for your measurement situation, including:
Measurement Timeout = 5.0 s
A typical measurement result is shown above.
Measuring Power versus Time
This measurement procedure assumes that your mobile supports two adjacent uplink timeslots.
-
Press the
CALL SETUP
key.
-
Press the
More
key (bottom right) to obtain the Call Parms (3 of 3) screen.
-
Press the
Multislot Config
(
F7
) key and select
2 down, 2 up
.
-
Connect the mobile.
-
Turn the mobile on and wait for
Attached
in the
Active Cell:
field.
Note: For mobiles that don't perform a GPRS attach automatically, set the mobile to data mode.
-
Press the
Start Data Connection
(
F3
) key and watch for the
Active Cell:
field changing to
Transferring
.
-
Press the
Measurement selection
key.
-
Select the
Power vs Time
measurement.
-
Press the
Power vs Time Setup (
F1
)
key.
-
Press the
Measurement Setup (
F1
)
key.
-
Set the measurement parameters as needed for your measurement situation, including:
Measurement Timeout = 10 s
-
Press the
Burst1 Meas Offsets
(
F2
) key and set up your required time offset values for burst 1.
For statistical analysis the test set allows you to set up to 12 time markers. These markers
do not
define the mask but allow you to obtain measurement results from specified points on the mask.
-
Press the
Burst2 Meas Offsets
(
F3
) key and set up your required time offset values for burst 2.
Offsets are referenced to bit zero in a normal burst. To obtain results from measurement points before bit 0, enter negative offset values.
-
Press the
Close Menu
(
F6
) key.
A typical summary results screen is shown above.
-
Press the
Return
(
F6
) key.
-
Press the
Change View
(
F2
) key.
-
Press the
Burst 1 numeric Results
(
F2
) key to view the offset measurement results for burst 1.
A typical numeric measurement results screen for offsets is shown above.
-
Press the
Burst 2 numeric Results
(
F3
) key to view the offset measurement results for burst 2.
-
Press the
Graph
(
F4
) key to access the full graph view of the two adjacent uplink bursts.
The structure of the Power versus Time graph menus are slightly different depending on the number of uplink bursts in the multislot configuration. If you require details on how to navigate through the Power versus Time graph menus, see either:
A typical full graph view of two adjacent uplink bursts is shown above.
Two types of pass/fail results are provided:
-
An overall measurement pass/fail indicator is positioned in the bottom left corner of the display:
-
The text "Pass" is shown in a green-colored font if the measurement has passed the entire mask.
-
The text "Fail" is shown in a red-colored font if the measurement has failed any part of the mask.
-
Individual pass/fail indicators are provided for each section of the mask. On the display, text labels corresponding to the following sections change color to indicate a pass or fail (green indicates a pass; red indicates a fail):
-
Rise
section of burst 1
-
Active (useful) section of
Burst 1
-
Guard
period between burst 1 and 2 (Note: this indicator is only available if you are measuring two uplinks)
-
Active (useful) portion of
Burst 2 (Note: this is only available if you are measuring two uplinks)
-
Fall
section
When the Multi-Measurement Count parameter is set to a number greater than 1, it is possible for some, or all of the individual pass/fail indicators to show a pass while the overall pass/fail indicator shows a fail. This is because the individual pass/fail indicators relate to the last of the multi-measurements made, whereas the overall mask pass/fail indicator is an accumulative result from all the multi-measurements made.
You can zoom in to individual sections of the graph by pressing the
Guard Period
(
F2
),
Burst 1 Views
(
F3
), or
Burst 2 Views
(
F4
) keys. In addition, you can control a marker or change axis values by pressing the
Graph Control
(
F5
) key.
A typical graph view of the falling edge of burst 2 is shown above. You can access this view by pressing the
Burst 2 Views
(
F4
) key then the
Falling Edge
(
F3
) key.
In this case, the marker is turned on and positioned on the falling edge. The marker can be turned on by pressing the
Graph Control
(
F5
) key then the
Marker Position
(
F2
) key. Set the required marker position using the knob or the numeric entry keys. The signal level at the marker and its position along the x-axis are shown at the top of the display above the graph.
Power versus Time Graph Menus (one uplink burst)
Power versus Time Graph Menus (two uplink bursts)
Measuring Phase and Frequency Error
-
Connect the mobile.
-
Turn the mobile on and wait for
Attached
in the
Active Cell:
field.
Note: For mobiles that don't perform a GPRS attach automatically, set the mobile to data mode.
-
Press the
Start Data Connection
(
F3
) key and watch for the
Active Cell:
field changing to
Transferring
.
-
Press the
Measurement selection
key.
-
Select the
Phase & Frequency Error
measurement.
-
Press the
Phase & Freq. Setup
(
F1
) key.
-
Set the measurement parameters as needed for your measurement situation, including:
Measurement Timeout = 10 s
A typical phase and frequency error measurement result is shown above.
-
Press the
Change View
(
F2
) key.
-
Press the
Graph
(
F2
) key to access the peak phase error graph.
A typical peak phase error graph is shown above.
The marker can be turned on by pressing the
Graph Control
(
F5
) key then the
Marker Position
(
F2
) key. Set the required marker position using the knob or the numeric entry keys. In addition, the axis values can be changed by pressing the
Axis Control
(
F1
) key. This allows you to zoom in or out to look at particular sections of the graph in more detail.
A typical peak phase error graph is shown above. This graph is zoomed in to show bits ranging from 100.0 to 120.0 at a 1.0 degrees per division setting. The Y-axis reference point is changed to 5.0 degrees.
In this case, the marker is turned on and positioned at bit 104. The phase error at the marker and its bit position are shown at the top of the display above the graph.
If you require details on how to navigate through the phase and frequency error graph menus, see
Phase and Frequency Error Graph Menus
.
Phase and Frequency Error Graph Menus
Measuring Output RF Spectrum
-
Connect the mobile.
-
Turn the mobile on and wait for
Attached
in the
Active Cell:
field.
Note: For mobiles that don't perform a GPRS attach automatically, set the mobile to data mode.
-
Press the
Start Data Connection
(
F3
) key and watch for the
Active Cell:
field changing to
Transferring
.
-
Press the
Measurement selection
key.
-
Select the
Output RF Spectrum
measurement.
-
Press the
ORFS Setup
(
F1
) key.
-
Press the
Measurement Setup
(
F1
) key.
-
Set the measurement parameters as needed for your measurement, including:
Measurement Timeout = 20 s.
-
Press the
Modulation Frequencies
(
F2
) key and set up your required offset frequencies for ORFS due to modulation.
-
Press the
Switching Frequencies
(
F3
) key and set up your required offset frequencies for ORFS due to switching.
-
Press the
Close Menu
(
F6
) key.
-
Press the
Return
(
F6
) key.
-
Press the
Change View
(
F2
) key.
-
Press the
Modulation Numeric 1 of 3
(
F1
) key.
The measurement results for the first eight positions in the ORFS due to modulation table are displayed. You can view the results for the remaining 14 positions by pressing the
Modulation Numeric 2 of 3
(
F2
) key or the
Modulation Numeric 3 of 3
(
F3
) key. Note that the results are in dB relative to the
30 kHz BW Power
shown on the screen.
-
Press the
Switching Numeric
(
F4
) key to display the measurement results for ORFS due to switching.
A typical measurement results screen for ORFS due to switching is shown above.
-
Press the
Graph
(
F5
) key then the
Modulation
(
F1
) key to access the ORFS due to modulation bar graph.
A typical ORFS due to modulation graph is shown above. On the display, the bar representing each offset is shown in yellow and the center frequency is indicated by a white vertical line. There are two limit lines displayed; the absolute limit line is green and the relative limit line is blue. The pass/fail result is relevant to the displayed measurement.
-
Press the
Switching
(
F2
) key to access the ORFS due to switching bar graph.
A typical ORFS due to switching graph is shown above. On the display, the bar representing each offset is shown in blue and the center frequency is indicated by a white vertical line.
In this case, the marker is turned on and positioned at the 600 kHz offset. The signal level and frequency of the offset where the marker is positioned, are shown at the top of the display above the graph.
The marker can be turned on by pressing the
Graph Control
(
F5
) key then the
Marker Position
(
F2
) key. Set the required marker position using the knob or the numeric entry keys. In addition, the axis values can be changed by pressing the
Axis Control
(
F1
) key. This allows you to zoom in or out to look at particular sections of the graph in more detail.
-
Press the
Modulation and Switching
(
F3
) key to access a screen displaying both graphs, as shown in the example below.
Note that each of the graphs have their own graph controls on
F4
and
F5
.
Any failing results are displayed in red.
If you require details on how to navigate through the Output RF Spectrum graph menus, see
Output RF Spectrum Graph Menus
.
-
To set your own limits for each offset use the Measurement Setup menu and set
Limit Source
to
Manual
. You can then use
Modulation Limits
(
F4
) and
Switching Limits
(
F5
) to set your required limits for each offset.
Output RF Spectrum Graph Menus
Measuring Bit Error
To make bit error measurements your mobile must be fitted with a Test SIM card and have ETSI Type B connection capability.
-
Press the
CALL SETUP
key.
-
Press the
More
key (bottom right) to obtain the Call Parms (3 of 3) screen.
-
Press the
Multislot Config
(
F7
) key and select
1 down, 1 up
.
-
Press the
Data Conn Type
(
F4
) key and select
ETSI Type B
.
-
Connect the mobile.
-
Turn the mobile on and wait for
Attached
in the
Active Cell:
field.
For mobiles that don't perform GPRS attach automatically, set the mobile to data mode.
-
Press the
Start Data Connection
(
F3
) key and watch for the
Active Cell:
field changing to
Transferring
.
-
Press the
Measurement selection
key.
-
Select the
Bit Error
measurement.
-
Press the
Bit Error Setup
(
F1
) key.
-
Set the measurement parameters as needed for your measurement situation, including:
Measurement Timeout = 13.0 s
A typical bit error ratio measurement result is shown above.
To see the effect of cell power on the bit error ratio, press the
Cell Power
(
F7
) key on the Call Parms (1 of 3) screen and slowly reduce the power while observing the Bit Error display.
Measuring Block Error
Block error (BLER) measurements cannot be run concurrently with bit error (BER) measurements. Starting a BLER measurement while a BER measurement is currently running, closes the BER measurement.
-
Press the
CALL SETUP
key.
-
Press the
More
key (bottom right) to obtain the Call Parms (3 of 3) screen.
-
Connect the mobile.
-
Turn the mobile on and wait for
Attached
in the
Active Cell:
field.
For mobiles that don't perform GPRS attach automatically, set the mobile to data mode.
-
Press the
Start Data Connection
(
F3
) key and watch for the
Active Cell:
field changing to
Transferring
.
-
Press the
Measurement selection
key.
-
Select the
Block Error
measurement.
-
Press the
Block Error Setup
(
F1
) key.
-
Set the measurement parameters as needed for your measurement situation, including:
Measurement Timeout = 13.0 s
Number of blocks to test = 100
A typical block error measurement result is shown above.
To see the effect of cell power on the block error, press the
Cell Power
(
F7
) key on the Call Parms (1 of 3) screen and slowly reduce the power while observing the Block Error Ratio display.
Back to Top
|