The function you requested is not yet available.
|
The test set does not have this capability.
|
|
IQ Calibration completed successfully for modulator <N>
|
<N> is the IQ modulator number that you are attempting to calibrate, <N> is 1 or 2.
|
|
IQ Calibration failed for modulator <N>
|
<N> is the IQ modulator number that you are attempting to calibrate, <N> is 1 or 2.
|
|
The function you requested is not available in this TA.
|
This function is used in another Test Application.
|
|
IQ Calibration for modulator 1 in progress (10 minutes). Call processing disabled.
|
This error is cleared by either the; IQ Calibration completed successfully for modulator <N>, or IQ Calibration failed for modulator <N>.
|
|
IQ Calibration for modulator 2 in progress (6 minutes). Call processing disabled.
|
This error is cleared by either the; IQ Calibration completed successfully for modulator <N>, or IQ Calibration failed for modulator <N>.
|
|
Instrument warning: Audio generator instrument has been closed.
|
The audio generator instrument was closed automatically by the test set.
|
|
Measurement warning: Audio Analyzer instrument has been closed.
|
The audio analyzer instrument was closed automatically by the test set. |
|
Measurement warning: Analog audio measurement has been closed.
|
Analog audio measurements have been closed by the test set. |
|
Protocol warning: Setting change has terminated the link with the DUT
|
Some setting change has caused the call to disconnect.
|
|
Warning; Call processing disabled
|
The call processing functions are not active because the test set is performing calibration operations.
|
|
IQ first modulator calibration X%
|
IQ modulator calibration in progress, X represents the percent complete.
|
|
IQ second modulator calibration X%
|
IQ modulator calibration in progress, X represents the percent complete.
|