Verifying the Test Set

Last updated: 2014-11-28

Description

If you are experiencing problems with your test set, first ensure that you have performed the recommended calibrations (see the Calibrating the Test Set). If you are still experiencing problems, run the following verifications to determine if a hardware failure has occurred in your test set.

Self Test

The Self Test verification checks the integrity of the test set's RF Interface module.

You do not have to perform the Self Test verification from each application/format in your test set; perform it from one application/format verifies the test set's RF Interface module.

To initiate the Self Test verification from the front panel, press the SYSTEM CONFIG key, press the right More ( 1 of 2 ) key, and select Self Test ( F8 ). The Self Test verification takes approximately 15 seconds to execute, and returns pass or fail.

Thermal Power Null Adjustment

The Thermal Power Null Adjustment is a periodic calibration that you must perform on your test set. However, if you are experiencing problems with your test set, you can also use the Thermal Power Null Adjustment to ensure that your test set's Thermal Power Detector auto-null circuit has not drifted such that it is no longer able to null the power detector. If this occurs, the Thermal Power Null Adjustment returns a fail.

Related Topics

GPIB Commands: SYSTem:INSTrument:THERmal:POWer:NULL:ADJust?