CALL:RESelection

Last updated: May 26, 2009

This section is only applicable to the lab application.

GPIB Command Conventions

CALL[:CELL]:RESelection:GSM:SHCS[:SVALue]

Function

Sets/queries the Shcs,rat (GSM), and sets its state to ON.

This setting controls the Shcs,rat parameter in SIB3. This parameter determines the serving cell RSCP level at which the UE does not need to make inter-RAT measurements.

This setting only affects SIB3 if both the "Inter-RAT (GSM) and "Shcs,rat (GSM) State" settings are On.

Setting

Range: - 105 to 91 (in increments of 2)

Query

Range: - 105 to 91 (in increments of 2)

*RST Setting

0

Requirements

Lab Application Revision: F.00.19 and above

Programming Example

 
OUTPUT 714;"CALL:RESelection:GSM:SHCS 0" 
!Sets the Shcs,rat (GSM) to 0 and turns it on.

CALL[:CELL]:RESelection:GSM:SHCS:STATe

Function

Sets/queries the Shcs,rat (GSM) State.

This setting controls whether the Shcs,rat (GSM) parameter will be included in SIB3.

Setting

Range: 1|ON|0|OFF

Query

Range: 1|0
*RST Setting 0 (off)
Requirements Lab Application Revision: F.00.19 and above

Programming Example

 
OUTPUT 714;"CALL:RESelection:GSM:SHCS:STATe OFF"
! Sets the Shcs,rat (GSM) State to Off.

CALL[:CELL]:RESelection:GSM:SHCS:VALue

Function

Sets/queries the Shcs,rat (GSM).

This setting controls the Shcs,rat parameter in SIB3. This parameter determines the serving cell RSCP level at which the UE does not need to make inter-RAT measurements.

This setting only affects SIB3 if both the "Inter-RAT (GSM) and "Shcs,rat (GSM) State" settings are On.

Setting

Range: - 105 to 91 (in increments of 2)

Query

Range: - 105 to 91 (in increments of 2)

*RST Setting

0

Requirements

Lab Application Revision: F.00.19 and above

Programming Example

 
OUTPUT 714;"CALL:RESelection:GSM:SHCS:VALue 0" 
!Sets the Shcs,rat (GSM) to 0.

CALL[:CELL]:RESelection:GSM:SLIMit

Function

Sets/queries the Slimit,searchrat (GSM).

This parameter controls the Slimit,searchrat (GSM) parameter in SIB3.

Setting

Range: -32 to 20 (in increments of 2)

Query

Range: -32 to 20 (in increments of 2)

*RST Setting

0

Requirements

Lab Application Revision: F.00.19 and above

Programming Example

 
OUTPUT 714;"CALL:RESelection:GSM:SLIMit" 
!Sets the Slimit,searchrat (GSM) to 0.

CALL[:CELL]:RESelection:GSM:SSEarch

Function

Sets/queries the Ssearch,rat (GSM).

This setting controls the Ssearch,rat parameter in SIB3. This broadcast parameter determines the serving cell CPICH Ec/No level at which the UE does not need to make inter-RAT measurements.

Setting

Range: -32 to 20 (in increments of 2)

Query

Range: -32 to 20 (in increments of 2)

*RST Setting

0

Requirements

Lab Application Revision: F.00.19 and above

Programming Example

 
OUTPUT 714;"CALL:RESelection:GSM:SSEarch 0" 
!Sets the Ssearch,rat (GSM) to 0.

CALL[:CELL]:RESelection:GSM:STATe

Function

Sets/queries the Inter-RAT (GSM) 'S' Parameters State.

This setting controls whether any of the inter-RAT cell selection/reselection parameters are included in SIB3.

Setting

Range: 1|ON|0|OFF

Query

Range: 1|0
*RST Setting 0 (off)
Requirements Lab Application Revision: C.03 and above

Programming Example

 
OUTPUT 714;"CALL:RESelection:GSM:STATe OFF"
! Sets Inter-RAT (GSM) 'S' Parameters State to Off.

CALL[:CELL]:RESelection:HYSTeresis[1]

Function

Sets/queries Hysteresis 1 (Qhyst1) . This setting is only applicable when Quality Measure is set to CPICH RSCP .

This setting can only be changed when call status is Idle . If BCCH Update Page is set to Auto and you change this setting while your UE is camped to the test set, the test set performs a Broadcast Channel (BCCH) Update Procedure (SIB Contents) . If BCCH Update Page is set to Inhibit (the default value) and you change this setting while your UE is camped to the test set, you must force your UE to re-read the BCCH to observe the change (see BCCH Update Page for more details).

Setting

Range: 0 dB to +40 dB

Resolution: 1 dB

Query

Same as setting

*RST Setting

0 dB

Requirements

Lab Application Revision: D.00.04 and above

Programming Example

 
OUTPUT 714;"CALL:RESelection:HYSTeresis 3"
! Sets the Hysteresis 1 (Qhyst1) to 3 dB.

CALL[:CELL]:RESelection:HYSTeresis2

Function

Sets/queries Hysteresis 2 (Qhyst2) . This setting is only applicable when Quality Measure is set to CPICH Ec/N0 .

This setting can only be changed when call status is Idle . If BCCH Update Page is set to Auto and you change this setting while your UE is camped to the test set, the test set performs a Broadcast Channel (BCCH) Update Procedure (SIB Contents) . If BCCH Update Page is set to Inhibit (the default value) and you change this setting while your UE is camped to the test set, you must force your UE to re-read the BCCH to observe the change (see BCCH Update Page for more details).

Setting

Range: 0 dB to +40 dB

Resolution: 1 dB

Query

Same as setting

*RST Setting

0 dB

Requirements

Lab Application Revision: D.00.04 and above

Programming Example

 
OUTPUT 714;"CALL:RESelection:HYSTeresis2 3"
! Sets Hysteresis 2 (Qhyst2) to 3 dB.

CALL[:CELL]:RESelection:QUALity:LMINimum

Function

Sets/queries the Minimum Required Quality Level (Qqualmin) .

This setting can only be changed when call status is Idle . If BCCH Update Page is set to Auto and you change this setting while your UE is camped to the test set, the test set performs a Broadcast Channel (BCCH) Update Procedure (SIB Contents) . If BCCH Update Page is set to Inhibit (the default value) and you change this setting while your UE is camped to the test set, you must force your UE to re-read the BCCH to observe the change (see BCCH Update Page for more details).

Setting

Range: -24 dB to 0 dB

Resolution: 1 dB

Query

Same as setting

*RST Setting

-24 dB

Requirements

Lab Application Revision: D.00.04 and above

Programming Example

 
OUTPUT 714;"CALL:RESelection:QUALity:LMINimum -10"
! Sets Minimum Required Quality Level (Qqualmin) to -10 dB.

CALL[:CELL]:RESelection:QUALity:MEASure

Function

Sets/queries the Quality Measure .

This setting can only be changed when call status is Idle . If BCCH Update Page is set to Auto and you change this setting while your UE is camped to the test set, the test set performs a Broadcast Channel (BCCH) Update Procedure (SIB Contents) . If BCCH Update Page is set to Inhibit (the default value) and you change this setting while your UE is camped to the test set, you must force your UE to re-read the BCCH to observe the change (see BCCH Update Page for more details).

Setting

Range:

  • ECN0 - sets Quality Measure (Qmeas) to CPICH Ec/N0
  • ECNO - sets Quality Measure (Qmeas) to CPICH Ec/N0
  • RSCP - sets Quality Measure (Qmeas) to CPICH RSCP

Query

Range: RCSP|ECN0

*RST Setting

ECN0

Requirements

Lab Application Revision: D.00.04 and above

Programming Example

 
OUTPUT 714;"CALL:RESelection:QUALity:MEASure RSCP"
! Sets Quality Measure (Qmeas) to CPICH RSCP.

CALL[:CELL]:RESelection:RLMinimum

Function

Sets/queries the Minimum Required Rx Level (Qrxlevmin) .

This setting can only be changed when call status is Idle . If BCCH Update Page is set to Auto and you change this setting while your UE is camped to the test set, the test set performs a Broadcast Channel (BCCH) Update Procedure (SIB Contents) . If BCCH Update Page is set to Inhibit (the default value) and you change this setting while your UE is camped to the test set, you must force your UE to re-read the BCCH to observe the change (see BCCH Update Page for more details).

Setting

Range: -115 dBm to -25 dBm

Resolution: 1 dB

Query

Same as setting

*RST Setting

-115 dBm

Requirements

Lab Application Revision: D.00.04 and above

Programming Example

 
OUTPUT 714;"CALL:RESelection:RLMinimum -100"
! Sets Minimum Required Rx Level (Qrxlevmin) to -100 dBm.

CALL[:CELL]:RESelection:TIMer

Function

Sets/queries the Timer (Treselection) .

This setting can only be changed when call status is Idle . If BCCH Update Page is set to Auto and you change this setting while your UE is camped to the test set, the test set performs a Broadcast Channel (BCCH) Update Procedure (SIB Contents) . If BCCH Update Page is set to Inhibit (the default value) and you change this setting while your UE is camped to the test set, you must force your UE to re-read the BCCH to observe the change (see BCCH Update Page for more details).

Setting

Range: 0 s to 31 s

Resolution: 1 s

Query

Same as setting

*RST Setting

0 s

Requirements

Lab Application Revision: D.00.04 and above

Programming Example

 
OUTPUT 714;"CALL:RESelection:TIMer 10"
! Sets Timer (Treselection) to 10 seconds.