CALL:UPLink
Last updated: May 26, 2009
GPIB Command Conventions
CALL:UPLink:CHANnel[:MCHannel]
|
Function
|
The setting form of this command performs two functions: (1) sets
Uplink Channel
to a specific UARFCN (see
CALL:UPLink:CHANnel:CHANnel
) and (2) sets uplink channel control to off (manual) (see
CALL:UPLink:CHANnel:CONTrol:AUTO
).
The query form of this command returns the uplink UARFCN that is used by the test set when
CALL:UPLink:CHANnel:CONTrol:AUTO
is set to off (manual). Note, when
CALL:UPLink:CHANnel:CONTrol:AUTO
is set to on (auto), the uplink UARFCN used by the test set is determined by
CALL:CHANnel
.
The uplink UARFCN being used by the test set at any given time can be determined by
CALL:STATus:UPLink:CHANnel?
.
This setting cannot be changed while in the active cell operating mode. Attempting to change this setting while in this mode results in an error issued and the test set will reject the change. Use the cell off operating mode when changing this value. See
CALL:OPERating
to select operating modes.
If the
Uplink Channel
and
DL Channel
settings have an uplink/downlink offset of less than 30 MHz, the warning message "+217, Performance not specified for UL/DL frequency offset < 30 MHz" will be issued. The setting will be used as entered if not changed by the user.
|
|
Setting
|
Range: see
CALL:UPLink:CHANnel:CHANnel
|
|
Query
|
Same as setting |
|
*RST Setting
|
Uplink UARFCN = 9750, uplink channel control = on (auto)
|
|
Requirements
|
Test or Lab Application Revision: A.01 and above
Test Application Revision A.05.28 and Lab Application Revision B.01.29, added 812 - 837
Test Application Revision A.08 and Lab Application Revision C.03, changed channel ranges to match 3GPP TS 25.101 v6.7.0 (previously, channels supported were: 12-287, 812-837, 4095-4270, 4375-4600, 8525-8950 and 9225-9925, resolution = 1).
Test Application Revision A.09 and Lab Application Revisions C.04 and D.00, changed channel ranges to match 3GPP TS 25.101 v7.2.0 (previously channels supported were: 12, 37, 62, 87, 112, 137, 162, 187, 212, 237, 262, 287, 782, 787, 807, 812, 837, 862, 1162, 1187, 1212, 1237, 1262, 1287, 1312, 1337, 1362, 4132-4233 -- resolution = 1, 8562-8913 -- resolution = 1, 9262-9538 -- resolution = 1, 9612-9888 -- resolution = 1).
Test Application Revision A.10 and Lab Application Revision D.01, added 2012 - 2338, 2362, 2387, 2412, 2437, 2462, 2487, 2512, 2537, 2562, 2587, 2612, 2637, 2662, 2687
Test Application Revision A.11 and Lab Application Revisions D.02 and E.00, added 2887 to 3163, 3187, 3212, 3237, 3262, 3287, 3312, 3337, 3362, 3387, 3412, 3437, 3462
Lab Application Revision E.00.34 and above, added 3487 - 3587.
Test Application Revision A.14 and Lab Application Revision F.00 and above, added 3612 - 3678, 3702, 3707, 3732, 3737, 3762, 3767, 3792 - 3818, 3842, 3867, 3892 - 3918, 3942, 3967
|
|
Programming Example
OUTPUT 714;"CALL:UPLink:CHANnel 9755"
! Sets the uplink channel UARFCN to 9755 and selects manual control mode for uplink channel.
|
CALL:UPLink:CHANnel:CHANnel
|
Function
|
The setting form of this command sets
Uplink Channel
to a specific UARFCN.
Note, this uplink UARFCN is only used by the test set when
CALL:UPLink:CHANnel:CONTrol:AUTO
is set to off (manual). When
CALL:UPLink:CHANnel:CONTrol:AUTO
is set to on (auto), the uplink UARFCN used by the test set is determined by
CALL:CHANnel
.
The query form of this command returns the uplink UARFCN that is used by the test set when
CALL:UPLink:CHANnel:CONTrol:AUTO
is set to off (manual).
The uplink UARFCN being used by the test set at any given time can be determined by
CALL:STATus:UPLink:CHANnel?
.
This setting cannot be changed while in the active cell operating mode. Attempting to change this setting while in this mode results in an error issued and the test set will reject the change. Use the cell off operating mode when changing this value. See
CALL:OPERating
to select operating modes.
If the
Uplink Channel
and
DL Channel
settings have an uplink/downlink offset of less than 30 MHz, the warning message "+217, Performance not specified for UL/DL frequency offset < 30 MHz" will be issued. The setting will be used as entered if not changed by the user.
|
|
Setting
|
Range:
-
12, 37, 62, 87, 112, 137, 162, 187, 212, 237, 262, 287
-
782, 787, 807, 812, 837, 862
-
937 - 1288, resolution = 1
-
1312 - 1513, resolution = 1
-
1662, 1687, 1712, 1737, 1762, 1787, 1812, 1837, 1862
-
2012 - 2338, resolution = 1
-
2362, 2387, 2412, 2437, 2462, 2487, 2512, 2537, 2562, 2587, 2612, 2637, 2662, 2687
-
2712 - 2863, resolution = 1
-
2887 - 3163, resolution = 1
-
3187, 3212, 3237, 3262, 3287, 3312, 3337, 3362, 3387, 3412, 3437, 3462
-
3487 - 3587, resolution = 1
-
3612 - 3678, resolution = 1
-
3702, 3707, 3732, 3737, 3762, 3767
-
3792 - 3818, resolution = 1
-
3842, 3867
-
3892 - 3918, resolution = 1
-
3942, 3967
-
4132 - 4233, resolution = 1
-
8762 - 8912, resolution = 1
-
9262 - 9538, resolution = 1
-
9612 - 9888, resolution = 1
|
|
Query
|
Same as setting |
|
*RST Setting
|
9750
|
|
Requirements
|
Test Application Revision: A.06 and above
Lab Application Revision: C.01 and above
Test Application Revision A.08 and Lab Application Revision C.03, changed channel ranges to match 3GPP TS 25.101 v6.7.0 (previously, channels supported were: 12-287, 812-837, 4095-4270, 4375-4600, 8525-8950 and 9225-9925, resolution = 1).
Test Application Revision A.09 and Lab Application Revisions C.04 and D.00, changed channel ranges to match 3GPP TS 25.101 v7.2.0 (previously channels supported were: 12, 37, 62, 87, 112, 137, 162, 187, 212, 237, 262, 287, 782, 787, 807, 812, 837, 862, 1162, 1187, 1212, 1237, 1262, 1287, 1312, 1337, 1362, 4132-4233 -- resolution = 1, 8562-8913 -- resolution = 1, 9262-9538 -- resolution = 1, 9612-9888 -- resolution = 1).
Test Application Revision A.10 and Lab Application Revision D.01, added 2012 - 2338, 2362, 2387, 2412, 2437, 2462, 2487, 2512, 2537, 2562, 2587, 2612, 2637, 2662, 2687
Test Application Revision A.11 and Lab Application Revisions D.02 and E.00, added 2887 to 3163, 3187, 3212, 3237, 3262, 3287, 3312, 3337, 3362, 3387, 3412, 3437, 3462
Lab Application Revision: E.00.34 and above, added 3487 - 3587.
Test Application Revision A.14 and Lab Application Revision F.00 and above, added 3612 - 3678, 3702, 3707, 3732, 3737, 3762, 3767, 3792 - 3818, 3842, 3867, 3892 - 3918, 3942, 3967
|
|
Programming Example
OUTPUT 714;"CALL:UPLink:CHANnel:CHANnel 9755"
! Sets the uplink channel UARFCN to 9755.
|
CALL:UPLink:CHANnel:CONTrol:AUTO
|
Function
|
Sets/queries the
Uplink Channel
control mode.
The uplink UARFCN being used by the test set at any given time can be determined by
CALL:STATus:UPLink:CHANnel?
.
This setting cannot be changed while in the active cell operating mode. Attempting to change this setting while in this mode results in an error issued and the test set will reject the change. Use the cell off operating mode when changing this value. See
CALL:OPERating
to select operating modes.
|
|
Setting
|
Range:
-
1|ON (auto)
-
0|OFF (manual)
|
|
Query
|
Range: 1|0 |
|
*RST Setting
|
1 (auto)
|
|
Requirements
|
Test Application Revision: A.06 and above
Lab Application Revision: C.01 and above
|
|
Programming Example
OUTPUT 714;"CALL:UPLink:CHANnel:CONTrol:AUTO OFF"
! Sets uplink channel control mode to manual.
|
CALL:UPLink:CHANnel:SEParation
|
Function
|
THIS COMMAND IS OBSOLETE. The
Uplink Separation (Band)
setting has been replaced by the
Band Arbitrator
setting (
CALL[:CELL]:BARBitrator
). Setting CALL:UPLink:CHANnel:SEParation to VMHZ45 sets
Band Arbitrator
to
Band 5
. Setting CALL:UPLink:CHANnel:SEParation to any other value sets
Band Arbitrator
to
Band 6
.
Sets/queries
Uplink Separation (Band)
.
When
CALL:UPLink:CHANnel:CONTrol:AUTO
is set to on (auto), the uplink UARFCN used by the test set is determined by this
Uplink Separation
setting and by
CALL:CHANnel
.
If
CALL:UPLink:CHANnel:CONTrol:AUTO
is set to on (auto) and you attempt to set
Uplink Separation
to a value that would result in an invalid uplink channel based on the current
CALL:CHANnel
setting, an error is posted. With
CALL:UPLink:CHANnel:CONTrol:AUTO
set to off (manual), set
CALL:CHANnel
and
Uplink Separation
to a valid combination, then set
CALL:UPLink:CHANnel:CONTrol:AUTO
to on.
This setting cannot be changed while in active cell operating mode. Attempting to change this setting while in this mode results in an error issued and the test set will reject the change. Use the cell off operating mode when changing this value. See
CALL:OPERating
to select operating modes.
|
|
Setting
|
Range:
-
MHZ45 - 45 MHz (Band 6)
-
VMHZ45 - 45 MHz (Band 5)
-
MHZ80 - 80 MHz (Band 2)
-
MHZ95 - 95 MHz (Band 3)
-
MHZ190 - 190 MHz (Band 1)
-
MHZ400 - 400 MHz (Band 4)
|
|
Query
|
Same as setting |
|
*RST Setting
|
MHZ190 |
|
Requirements
|
Test Application Revision: A.06 and above
Test Application Revision A.08 and Lab Application Revision C.03, added VMHZ45
Lab Application Revision: C.01 and above
|
|
Programming Example
OUTPUT 714;"CALL:UPLink:CHANnel:SEParation MHZ80"
|
CALL:UPLink:DCCHannel:DDATa
|
Function
|
Sets/queries the
Uplink Dummy DCCH Data
state.
|
|
Setting
|
Range: 1|ON|0|OFF
|
|
Query
|
Range: 1|0
|
|
*RST setting |
0 (off) |
|
Requirements
|
Test Application Revision: A.04 and above
Lab Application Revision: A.02 and above
|
|
Programming Example
OUTPUT 714;"CALL:UPLink:DCCHannel:DDATa OFF"
|
CALL:UPLink:DPCChannel:POWer:INITial?
|
Function
|
Queries the expected initial power level for the uplink dedicated physical control channel (see
Initial DPCCH TX Power
).
|
|
Query
|
Range: -50 to 24 dBm and 9.91E+37 (NAN)
|
|
Requirements
|
Test Application Revision: A.02.25 and above.
Lab Application Revision: A.01 and above.
|
|
Programming Example
OUTPUT 714;"CALL:UPLink:DPCChannel:POWer:INITial?"
|
CALL:UPLink:DPCChannel:SLOT:FORMat
|
Function
|
Sets/queries the the uplink DPCCH slot format to UE for UL DTX or UL DTX+DL DRX operation.
|
|
Setting
|
Range: 1|4
|
|
Query
|
Range: 1|4
|
|
*RST setting |
1 |
|
Requirements
|
Lab Application Revision: F.00.19 and above.
|
|
Programming Example
OUTPUT 714;"CALL:UPLink:DPCChannel:SLOT:FORMat 1"
|
CALL:UPLink:DPCHannel:BETA:AUTo
|
Function
|
Sets/queries the power gain control mode for the uplink dedicated physical channel (DPCH) which consists of the dedicated physical data channel (DPDCH) and the dedicated physical control channel (DPCCH). See
Uplink DPCH Bc/Bd Control
for details.
ON = automatic
OFF = manual
This setting cannot be changed while in the active cell operating mode. Attempting to change this setting while in this mode results in an error issued and the test set will reject the change. Use the cell off or FDD test operating mode when changing this value. See
CALL:OPERating
to select operating modes.
|
|
Setting
|
Range: 1|ON|0|OFF
|
|
Query
|
Range: 1|0
|
|
*RST setting |
1 (on) |
|
Requirements
|
Test Application Revision: A.03 and above.
Lab Application Revision: A.01 and above.
Test Application Revision A.07 and above, this command applies to FDD test operating mode.
Lab Application Revision C.02 and above, this command applies to FDD test operating mode.
|
|
Programming Example
OUTPUT 714;"CALL:UPLink:DPCHannel:BETA:AUTo OFF"
! Enables manual control of the DPCH power gain.
|
CALL:UPLink:DPCHannel:MANual:CBETa
|
Function
|
Sets/queries the power gain of the dedicated physical control channel (DPCCH) when the DPCH power gain control mode is MANual (see
CALL:UPLink:DPCHannel:BETA:AUTo
). See
Manual Uplink DPCH Bc
.
This setting cannot be changed while in the active cell operating mode. Attempting to change this setting while in this mode results in an error issued and the test set will reject the change. Use the
Cell Off
or
FDD Test
operating mode when changing this value. See
CALL:OPERating
to select operating modes.
|
|
Setting
|
Range: 1 to 15
Resolution: 1
|
|
Query
|
Same as setting |
|
*RST setting |
8 |
|
Requirements
|
Test Application Revision: A.03 and above.
Lab Application Revision: A.01 and above.
Test Application Revision A.07 and Lab Application Revision C.02, this command applies to FDD test operating mode.
Lab Application Revision: D.00, range is 1 to 15 (but 1 is not recommended)
Lab Application Revision D.01, range is 2 to 15
Test Application Revision A.11 and Lab Application Revisions D.02 and E.00, range is 1 to 15 (vs. 2 to 15)
|
|
Programming Example
OUTPUT 714;"CALL:UPLink:DPCHannel:MANual:CBETa 10"
! Sets manual control of power gain for the DPCCH to 10.
|
CALL:UPLink:DPCHannel:MANual:DBETa
|
Function
|
Sets/queries the power gain of the dedicated physical data channel (DPDCH) when the DPCH power gain control mode is MANual (see
CALL:UPLink:DPCHannel:BETA:AUTo
). See
Manual Uplink DPCH Bd
.
This setting cannot be changed while in the active cell operating mode. Attempting to change this setting while in this mode results in an error issued and the test set will reject the change. Use the
Cell Off
or
FDD Test
operating mode when changing this value. See
CALL:OPERating
to select operating modes.
|
|
Setting
|
Range: 0 to 15
Resolution: 1
|
|
Query
|
Range: 0 to 15
|
|
*RST setting |
15 |
|
Requirements
|
Test Application Revision: A.03 and above.
Lab Application Revision: A.01 and above.
Test Application Revision A.07 and above, this command applies to FDD test operating mode.
Lab Application Revision C.02 and above, this command applies to FDD test operating mode.
|
|
Programming Example
OUTPUT 714;"CALL:UPLink:DPCHannel:MANual:DBETa 5"
! Sets manual control of power gain for the DPDCH to 5.
|
CALL:UPLink:DPCHannel:SCODe
|
Function
|
Sets/queries the uplink dedicated physical channel's scrambling code. See
Uplink Parameters
for details.
This setting cannot be changed while in the active cell operating mode. Attempting to change this setting while in this mode results in an error issued and the test set will reject the change. Use the cell off operating mode when changing this value. See
CALL:OPERating
to select operating modes.
|
|
Setting
|
Range: 0 to 16777215
|
|
Query
|
Same as setting |
|
*RST Setting
|
0
|
|
Requirements
|
Test Application Revision: A.02 and above.
Lab Application Revision: A.01 and above.
|
|
Programming Example
OUTPUT 714;"CALL:UPLink:DPCH:SCODe 10"
! Sets the uplink DPCH scrambling code to 10.
|
CALL:UPLink:DPCChannel:SLOT:FORMat
|
Function
|
Sets/queries the UL DPCCH Slot Format.
This setting configures the uplink DPCCH slot format for UL DTX or ULDTX+DL DRX operation in the UE. These two UL DPCCH slot formats are intended for use when the DPDCH is not configured in the uplink. The bit information for each uplink DPCCH slot format is decscribed in 3GPP 25.211 v760 Table 2. |
|
Setting
|
Range: 1|4
|
|
Query
|
Range: 1|4
|
|
*RST Setting
|
1
|
|
Requirements
|
Lab Application Revision: F.00.19 and above |
|
Programming Example
OUTPUT 714;"CALL:UPLink:DPCChannel:SLOT:FORMat 1"
! Sets the the UL DPCCH Slot Format to 1.
|
CALL:UPLink:PRAChannel:ASUBchannels
|
Function
|
Sets/queries the
Available Subchannels (Bit Mask)
.
This setting can only be changed when call status is
Idle
. If
BCCH Update Page
is set to
Auto
and you change this setting while your UE is camped to the test set, the test set performs a
Broadcast Channel (BCCH) Update Procedure (SIB Contents)
. If
BCCH Update Page
is set to
Inhibit
(the default value) and you change this setting while your UE is camped to the test set, you must force your UE to re-read the BCCH to observe the change (see
BCCH Update Page
for more details).
|
|
Setting
|
Range: 0 to 111111111111
(1 to 12 characters from the set 0 and 1).
|
|
Query
|
Range: "000000000000" to "111111111111"
(The returned string is always 12 characters from the set 0 and 1).
|
|
*RST Setting
|
"000000000001"
|
|
Requirements
|
Test Application Revision: A.03 and above.
Lab Application Revision: A.01 and above.
|
|
Programming Example
OUTPUT 714;"CALL:UPLink:PRAChannel:ASUBchannels '111111111111'"
! Sets all twelve PRACH access subchannels avai1able for use.
|
CALL:UPLink:PRAChannel:BETA:AUTo
|
Function
|
Sets/queries the power gain control mode for the message portion of the uplink physical random access channel (PRACH). See
Uplink Parameters
for details.
ON = automatic
OFF = manual
This setting can only be changed when call status is
Idle
. If
BCCH Update Page
is set to
Auto
and you change this setting while your UE is camped to the test set, the test set performs a
Broadcast Channel (BCCH) Update Procedure (SIB Contents)
. If
BCCH Update Page
is set to
Inhibit
(the default value) and you change this setting while your UE is camped to the test set, you must force your UE to re-read the BCCH to observe the change (see
BCCH Update Page
for more details).
|
|
Setting
|
Range: 1|ON|0|OFF
|
|
Query
|
Range: 1|0
|
|
*RST setting |
1 (automatic) |
|
Requirements
|
Test Application Revision: A.03 and above.
Lab Application Revision: A.01 and above.
|
|
Programming Example
OUTPUT 714;"CALL:UPLink:PRAChannel:BETA:AUTo OFF"
! Enables manual control of the PRACH power level.
|
CALL:UPLink:PRAChannel:MANual:CBETa
|
Function
|
Sets/queries the
Manual PRACH Bc
.
This command is only applicable when
CALL:UPLink:PRAChannel:BETA:AUTo
) is set to auto.
This setting can only be changed when call status is
Idle
. If
BCCH Update Page
is set to
Auto
and you change this setting while your UE is camped to the test set, the test set performs a
Broadcast Channel (BCCH) Update Procedure (SIB Contents)
. If
BCCH Update Page
is set to
Inhibit
(the default value) and you change this setting while your UE is camped to the test set, you must force your UE to re-read the BCCH to observe the change (see
BCCH Update Page
for more details).
|
|
Setting
|
Range: 2 to 15
Resolution: 1
|
|
Query
|
Range: 2 to 15
|
|
*RST setting |
15 |
|
Requirements
|
Test Application Revision: A.03 and above.
Lab Application Revision: A.01 and above.
|
|
Programming Example
OUTPUT 714;"CALL:UPLink:PRAChannel:MANual:CBETa 10"
! Sets manual control of power gain for the control data
! of the message portion of the PRACH to 10.
|
CALL:UPLink:PRAChannel:MANual:DBETa
|
Function
|
Sets/queries the
Manual PRACH Bd
.
This command is only applicable when
CALL:UPLink:PRAChannel:BETA:AUTo
) is set to auto.
This setting can only be changed when call status is
Idle
. If
BCCH Update Page
is set to
Auto
and you change this setting while your UE is camped to the test set, the test set performs a
Broadcast Channel (BCCH) Update Procedure (SIB Contents)
. If
BCCH Update Page
is set to
Inhibit
(the default value) and you change this setting while your UE is camped to the test set, you must force your UE to re-read the BCCH to observe the change (see
BCCH Update Page
for more details).
|
|
Setting
|
Range: 0 to 15
Resolution: 1
|
|
Query
|
Range: 0 to 15
|
|
*RST setting |
15 |
|
Requirements
|
Test Application Revision: A.03 and above.
Lab Application Revision: A.01 and above.
|
|
Programming Example
OUTPUT 714;"CALL:UPLink:PRAChannel:MANual:DBETa 5"
! Sets manual control of power gain for message data of
! the message portion of the PRACH to 5.
|
CALL:UPLink:PRAChannel:POWer:INITial?
|
Function
|
Queries the expected initial power level for the uplink physical random access channel (see
Initial PRACH TX Power
).
|
|
Query
|
Range: -50 to 24 dBm and 9.91E+37 (NAN)
|
|
Requirements
|
Test Application Revision: A.02.25 and above.
Lab Application Revision: A.01 and above.
|
|
Programming Example
OUTPUT 714;"CALL:UPLink:PRAChannel:POWer:INITial?"
|
CALL:UPLink:PRAChannel:POWer[:RAMP]:STEP[:LEVel]
|
Function
|
Sets/queries the
PRACH Power Ramp Step
.
This setting can only be changed when call status is
Idle
. If
BCCH Update Page
is set to
Auto
and you change this setting while your UE is camped to the test set, the test set performs a
Broadcast Channel (BCCH) Update Procedure (SIB Contents)
. If
BCCH Update Page
is set to
Inhibit
(the default value) and you change this setting while your UE is camped to the test set, you must force your UE to re-read the BCCH to observe the change (see
BCCH Update Page
for more details).
|
|
Setting
|
Range: 1 to 8
Resolution: 1
|
|
Query
|
Range: 1 to 8
|
|
*RST Setting
|
3
|
|
Requirements
|
Test Application Revision: A.02 and above.
Lab Application Revision: A.01 and above.
|
|
Programming Example
OUTPUT 714;"CALL:UPLink:PRAChannel:POWer:STEP:LEVel 5"
|
CALL:UPLink:PRAChannel:PREambles:NUMBer
|
Function
|
Sets/queries the
PRACH Preambles
.
This setting can only be changed when call status is
Idle
. If
BCCH Update Page
is set to
Auto
and you change this setting while your UE is camped to the test set, the test set performs a
Broadcast Channel (BCCH) Update Procedure (SIB Contents)
. If
BCCH Update Page
is set to
Inhibit
(the default value) and you change this setting while your UE is camped to the test set, you must force your UE to re-read the BCCH to observe the change (see
BCCH Update Page
for more details).
|
|
Setting
|
Range: 1 to 64
Resolution: 1
|
|
Query
|
Range: 1 to 64
|
|
*RST Setting
|
64
|
|
Requirements
|
Test Application Revision: A.03 and above.
Lab Application Revision: A.01 and above.
|
|
Programming Example
OUTPUT 714;"CALL:UPLink:PRAChannel:PREambles:NUMBer 15"
! Sets the maximum number of preambles per
! PRACH Ramping Cycle to 15.
|
CALL:UPLink:PRAChannel:PREambles:RCYCles[:MMAX]
|
Function
|
Sets/queries
PRACH Ramping Cycles (MMAX)
.
This setting can only be changed when call status is
Idle
. If
BCCH Update Page
is set to
Auto
and you change this setting while your UE is camped to the test set, the test set performs a
Broadcast Channel (BCCH) Update Procedure (SIB Contents)
. If
BCCH Update Page
is set to
Inhibit
(the default value) and you change this setting while your UE is camped to the test set, you must force your UE to re-read the BCCH to observe the change (see
BCCH Update Page
for more details).
|
|
Setting
|
Range: 1 to 32
Resolution: 1
|
|
Query
|
Range: 1 to 32
|
|
*RST Setting
|
2
|
|
Requirements
|
Test Application Revision: A.03 and above.
Lab Application Revision: A.01 and above.
|
|
Programming Example
OUTPUT 714;"CALL:UPLink:PRAChannel:PREambles:RCYCles 5"
! Sets the PRACH ramping cycles per PRACH access request to 5.
|
CALL:UPLink:PRAChannel:SCODe
|
Function
|
Sets/queries the
PRACH Scrambling Codeword
.
This parameter can not be set in active cell operating mode.
|
|
Setting
|
Range: 0 to 15
Resolution: 1
|
|
Query
|
Range: 0 to 15
|
|
*RST Setting
|
0
|
|
Requirements
|
Test Application Revision: A.02 and above.
Lab Application Revision: A.01 and above.
|
|
Programming Example
OUTPUT 714;"CALL:UPLink:PRAChannel:SCODe 5"
|
CALL:UPLink:PRAChannel:SIGNature
|
Function
|
This command is obsolete, please use the
CALL:UPLink:PRAChannel:SMASk
command instead.
Sets/queries the uplink physical random access channel's preamble signature. Changing this setting changes the
CALL:UPLink:PRAChannel:SMASk
setting as well.
While in the active cell operating mode, this setting can only be changed when the call status is Idle.
|
|
Setting
|
Range: 0 to 15
Resolution: 1
|
|
Query
|
Same as setting |
|
*RST Setting
|
0
|
|
Requirements
|
Test Application Revision: A.02 and above, obsolete as of A.05.
|
|
Programming Example
OUTPUT 714;"CALL:UPLink:PRAChannel:SIGNature 11"
! Set the PRACH signature to 11.
|
CALL:UPLink:PRAChannel:SMASk
|
Function
|
Sets/queries the
PRACH Signature
.
This setting can only be changed when call status is
Idle
. If
BCCH Update Page
is set to
Auto
and you change this setting while your UE is camped to the test set, the test set performs a
Broadcast Channel (BCCH) Update Procedure (SIB Contents)
. If
BCCH Update Page
is set to
Inhibit
(the default value) and you change this setting while your UE is camped to the test set, you must force your UE to re-read the BCCH to observe the change (see
BCCH Update Page
for more details).
|
|
Setting
|
Range: 0 to 1111111111111111
(1 to 16 characters from the set 0 and 1)
|
|
Query
|
Range: "0000000000000000" to "1111111111111111"
(The return string is always 16 characters from the set 0 and 1)
|
|
*RST Setting
|
"0000000000000001" |
|
Requirements
|
Test Application Revision: A.05 and above
Lab Application Revision: A.02 and above
|
|
Programming Example
OUTPUT 714;"CALL:UPLink:PRAChannel:SMASk '0000000000000001'"
|
CALL:UPLink:PRAChannel:TIMing[:OFFSet]
|
Function
|
Sets/queries
Uplink Timing Offset
.
This setting cannot be changed while in the active cell operating mode. Attempting to change this setting while in this mode results in an error issued and the test set will reject the change. Use the cell off operating mode when changing this value. See
CALL:OPERating
to select operating modes.
|
|
Setting
|
Range: -256 to +256
|
|
Query
|
Same as setting |
|
*RST Setting
|
0 chips
|
|
Requirements
|
Test Application Revision: A.02 and above
Lab Application Revision: A.01 and above
|
|
Programming Example
OUTPUT 714;"CALL:UPLink:TIMing 16"
|
CALL:UPLink:TXPower:LEVel:MAXimum
|
Function
|
Sets/queries
Maximum Uplink Transmit Power Level
.
This setting can only be changed when call status is
Idle
. If
BCCH Update Page
is set to
Auto
and you change this setting while your UE is camped to the test set, the test set performs a
Broadcast Channel (BCCH) Update Procedure (SIB Contents)
. If
BCCH Update Page
is set to
Inhibit
(the default value) and you change this setting while your UE is camped to the test set, you must force your UE to re-read the BCCH to observe the change (see
BCCH Update Page
for more details).
|
|
Setting
|
Range: -50 to +33 dBm
|
|
Query
|
Same as setting |
|
*RST Setting
|
33 dBm |
|
Requirements
|
Test Application Revision: A.05 and above
Lab Application Revision: B.01 and above
|
|
Programming Example
OUTPUT 714;"CALL:UPLink:TXPower:LEVel:MAXimum 24"
! Sets Maximum Uplink Transmit Power Level to +24 dBm.
|
|
|