READ

Last updated: May 26, 2009

GPIB Command Conventions

READ:AFANalyzer[:ALL]?

Function

Initiates and fetches an audio frequency analyzer measurement as a sequential operation.

Measurement results are returned in the following order: integrity, average audio level, average SINAD, average audio distortion.

Query

Integrity: see Integrity Indicator

Average audio level

  • Range: 0 to 20 volts, 9.91 E+37
  • Resolution: the lesser of 4 significant digits or 10 uV.

Average SINAD

  • Range: -99.00 to 99.00 dB, 9.91 E+37
  • Resolution: 0.01 dB

Average audio distortion

  • Range: 0.00 to 99.00%, 9.91 E+37
  • Resolution: 0.01%

Requirements

Test or Lab Application Revision: A.01 and above

READ:CFDTune[:ALL]?

Function

Initiates and fetches the fast device tune measurement as a sequential operation.

Results are returned in the following order:

  • Integrity
  • Number of Frequency Steps
  • Number of Power Steps
  • Number of MS TX power steps measured (in most cases, it equals Number of Frequency Steps multiplied by Number of Power Steps )
  • Comma-separated list of 400 MS Tx Power results. 400 results are always returned; if the test set measured fewer than 400 power steps from the mobile, the remaining values are NANs (9.91E+37). The power results for the first frequency in the MS TX Frequency Sequence are returned first, then the power results for the second frequency in the MS TX Frequency Sequence, etc., followed by NANs if fewer than 400 points were measured.

See Fast Device Tune Measurement Results .

Query

Integrity

Number of Frequency Steps

  • Range: 1 to 20, 9.91 E+37

  • Resolution: 1.0

Number of Power Steps

  • Range: 1 to 20, 9.91 E+37

  • Resolution: 1.0

Number of MS Tx Power Steps Measured

  • Range: 1 to 400, 9.91 E+37

  • Resolution: 1.0

MS Tx Power Results Array

  • Range: -100.00 to +100.00 dBm, 9.91 E+37

  • Resolution: 0.01 dBm

Requirement

Test Application Revision: A.09 and above, with the required feature license

Lab Application Revision: C.04 and D.00.04 and above

Programming Example

 
OUTPUT 714;"READ:CFDTune:ALL?"

READ:EVMQpsk[:ALL]?

Function

Initiates and fetches a QPSK EVM measurement as a sequential operation.

Measurement results are returned in the following order: integrity, EVM maximum, frequency error worst case, origin offset maximum, phase error maximum, magnitude error maximum, peak EVM maximum.

If the multi-measurement state is on, each result represents the average of a number of QPSK EVM measurements. Multi-measurement state, by default, is off. To set up multi-measurements, see SETup:EVMQpsk:COUNt[:SNUMber] .

Query

Integrity

  • Range: 0 to 16

  • Resolution: 1

EVM Maximum

  • Range: 0 to 99%, 9.91 E+37

  • Resolution: 0.1

Frequency Error worst case

  • Range-99.99 to 99.99 kHz, 9.91 E+37

  • Resolution: 0.1

Origin Offset Maximum

  • Range: -99 to 0 dB, 9.91 E+37

  • Resolution: 0.1

Phase Error Maximum

  • Range: 0 to 180 degrees, 9.91 E+37

  • Resolution: 0.1

Magnitude Error Maximum

  • Range: 0 to 99%, 9.91 E+37

  • Resolution: 0.1

Peak EVM Maximum

  • Range: 0 to 99%, 9.91 E+37

  • Resolution: 0.1

READ:FSTability[:ALL]?

Function

Initiates and fetches a frequency stability measurement as a sequential operation.

Measurement results are returned in the following order: integrity, worst-case frequency error, average frequency.

Query

Worst-case frequency error

  • Range: -500 to 500 parts per million (ppm)
  • Resolution: 0.01

Average frequency

  • Range: 100 E+6 to 3 E+9 Hz
  • Resolution: 1 Hz

READ:HBLerror[:ALL]?

Function

Initiates and fetches an HSDPA block error ratio measurement as a sequential operation.

Measurement results are returned in the following order: Integrity Indicator, Block Error Ratio in %, Information Bit Throughput in kbps, Positive Acknowledgments, Negative Acknowledgments, Statistical DTXs, Number of Blocks Tested, Median Channel Quality Indicator.

Query

Integrity Indicator - See Integrity Indicator .

Block Error Ratio

  • Range: 0.00 to 100.00%, 9.91E+37
  • Resolution: 0.01%

Information Bit Throughput

  • Range: 0.000 to 14000.000 kbps, 9.91E+37
  • Resolution: 0.001 kbps

Positive Acknowledgments

  • Range: 0 to 99000, 9.91E+37
  • Resolution: 1

Negative Acknowledgments

  • Range: 0 to 99000, 9.91E+37
  • Resolution: 1

Statistical DTXs

  • Range: 0 to 99000, 9.91E+37
  • Resolution: 1

Number of Blocks Tested

  • Range: 0 to 99000, 9.91E+37
  • Resolution: 1

Median Channel Quality Indicator

  • Range: 0 to 30, 9.91E+37
  • Resolution: 1

Requirements

Test Application Revision: A.08 and above, with the required feature license

Lab Application Revision: C.03 and above

READ:HRCQuality[:ALL]?

Function

Initiates and fetches a Reporting of Channel Quality Indicator measurement as a sequential operation.

Measurement results are returned in the following order: Integrity Indicator, Reporting of Channel Quality Indicator Pass/Fail Result.

Query

Integrity Indicator - See Integrity Indicator .

Reporting of Channel Quality Indicator

  • Range: 0 (pass), 1 (fail), 9.91E+37

Requirements

Lab Application Revision: E.00.16 and above

READ:SAUDio[:ALL]?

Function

Initiates and fetches a swept audio measurement as a sequential operation.

Measurement results are returned in the following order: integrity indicator, average swept audio level and average distortion for the first frequency point, average swept audio level and average distortion for the second frequency point, and so on. The number of frequency points is determined with the SETup:SAUDio:FREQuency:POINts command.

For frequencies above 10 kHz, this measurement returns 9.91 E+37.

Query

Integrity indicator: see Integrity Indicator

Average audio level

  • Range: 1 mV to 20 V, 9.91 E+37
  • Resolution: the lesser of 4 significant digits or 10 uV

Average distortion

  • Range: 0 to 99.9%, 9.91 E+37
  • Resolution: 0.01%

READ:SMONitor:TRACe?

Function

Queries (initiates and fetches) one spectrum monitor measurement as a sequential operation.

Returns a comma separated list of 401 points representing the amplitude results of the trace data.

Query

Range:

  • For spans of 125 kHz and 500 kHz

    • frequency of point: within the current span (relative to the center frequency)
    • amplitude of point: -50 to +55 dB
  • For all other spans

    • frequency of point: within the current span (relative to the center frequency)
    • amplitude of point: (reference level - 60 dB) to +55 dB

Requirements

Test Application Revision: A.03 and above

Lab Application Revision: A.01 and above.

READ:WACLeakage[:ALL]?

Function

Initiates and fetches an adjacent channel leakage ratio measurement as a sequential operation.

Measurement results are returned in the following order: integrity, -5 MHz minimum offset power ratio, +5 MHz maximum offset power ratio, -10 MHz minimum offset power ratio,
+10 MHz maximum offset power ratio.

If the measurement state of any offset is set to OFF, the result for that offset will be 9.91 E+37.

Query
  • Range: -100 to 100 dBm
  • Resolution: 0.01 dB

Requirements

Test or Lab Application Revision: A.01 and above

READ:WBERror[:ALL]?

Function

Initiates and fetches a loopback bit error ratio measurement as a sequential operation.

Measurement results are returned in the following order: integrity, bit error ratio, bit error count, bits tested.

Query

Range

  • Integrity: 1 to 23
  • Bit Error Ratio: 0 to 99%, 9.91 E+37
  • Bit Error Count: 0 to 999999, 9.91 E+37
  • Bits Tested: 0 to 999999, 9.91 E+37

Resolution

  • Integrity: 1
  • Bit Error Ratio: 0.0001 dB
  • Bit Error Count: 1
  • Bits Tested: 1

Requirements

Test or Lab Application Revision: A.01 and above

READ:WBLerror[:ALL]?

Function

Initiates and fetches a loopback block error ratio measurement as a sequential operation.

Measurement results are returned in the following order: integrity, block error ratio, block error count, blocks tested, uplink missing blocks.

See Block Error Ratio Measurement Results .

Query

Range

  • Integrity: see Integrity Indicator .
  • Block Error Ratio: 0 to 99%, 9.91 E+37
  • Block Error Count: 0 to 99000, 9.91 E+37
  • Blocks Tested: 0 to 99000, 9.91 E+37
  • Uplink Missing Blocks: 0 to 99000, 9.91 E+37

Resolution

  • Integrity: 1
  • Block Error Ratio: 0.0001%
  • Block Error Count: 1
  • Blocks Tested: 1
  • Uplink Missing Blocks: 1

Requirements

Test Application Revision: A.05 and above.

Lab Application Revision: B.01 and above.

Programming Example

 
OUTPUT 714;"READ:WBLerror?" 

READ:WCDomain[:ALL]?

Function

Initiates and fetches a code domain measurement as a sequential operation.

Measurement results are returned in the following order: integrity, EVM, frequency error, origin offset, phase error, magnitude error, time error, peak code domain error.

Query

Integrity

  • Range: 0 to 23

  • Resolution: 1

EVM Maximum

  • Range: 0 to 99%, 9.91 E+37

  • Resolution: 0.1

Frequency Error worst case

  • Range: -99.99 to 99.99 kHz, 9.91 E+37

  • Resolution: 0.1

Origin Offset Maximum

  • Range: -99 to 0 dB, 9.91 E+37

  • Resolution: 0.1

Phase Error Maximum

  • Range: 0 to 180 degrees, 9.91 E+37

  • Resolution: 0.1

Magnitude Error Maximum

  • Range: 0.0 to 99.0%

  • Resolution: 0.1

Time Error Maximum

  • Range: -99 to 99 µ s, 9.91 E+37

  • Resolution: 1 ns

Peak Code Domain Error Maximum

  • Range: -99.0 to 0.0 dB, 9.91 E+37

  • Resolution: 0.1

Requirements

Test Application Revision:

Lab Application Revision: A.01 and above.

READ:WCPower[:ALL]?

Function

Initiates and fetches a channel power measurement as a sequential operation.

Measurement results are returned in the following order: integrity, average channel power.

If the multi-measurement state is on, the result represents the average of a number of channel power measurements. Multi-measurement state, by default, is off. To set up multi-measurements, see SETup:WCPower:COUNt[:SNUMber] .

Query
  • Range: -99 to 99 dBm
  • Resolution: 0.01 dB

Requirements

Test or Lab Application Revision: A.01 and above

READ:WCTFormat[:ALL]?

Function

Initiates and fetches a change of TFC measurement as a sequential operation.

Measurement results are returned in the following order: Integrity Indicator, Step Down Relative Power (average result if SETup:WCTFormat:COUNt[:SNUMber] is greater than 0), Step Down Error, Step Down Pass/Fail, Step Up Relative Power (average result if SETup:WCTFormat:COUNt[:SNUMber] is greater than 0), Step Up Error, Step Up Pass/Fail.

See Change of TFC Measurement Results .

Query

Integrity Indicator - see Integrity Indicator .

Step Down Relative Power

  • Range: -100.0 to 100.0 dB, 9.91E+37
  • Resolution: 0.1 dB

Step Down Error

  • Range: -100.0 to 100.0 dB, 9.91E+37
  • Resolution: 0.1 dB

Step Down Pass/Fail

  • Range: 0 (pass), 1 (fail), 9.91E+37

Step Up Relative Power

  • Range: -100.0 to 100.0 dB, 9.91E+37
  • Resolution: 0.1 dB

Step Up Error

  • Range: -100.0 to 100.0 dB, 9.91E+37
  • Resolution: 0.1 dB

Step Up Pass/Fail

  • Range: 0 (pass), 1 (fail), 9.91E+37

Requirements

Lab Application Revision: C.01 and above

Programming Example

 
OUTPUT 714;"READ:WCTFormat?"

READ:WDPAnalysis[:ALL]?

Function

Initiates and fetches a dynamic power analysis measurement as a sequential operation.

Measurement results are returned in the following order: integrity indicator, the number of steps measured, and a comma-separated list of average power values for each step measured (in dBm).

See Dynamic Power Analysis Measurement Results .

Query

Integrity Indicator - see Integrity Indicator .

Number of Steps

  • Range: 0 to 200
  • Resolution: 1

Average Power

  • Range: -99.99 to 99.99 dBm, 9.91E+37
  • Resolution: 0.01 dBm

Requirements

Test Application Revision: A.05 and above

Lab Application Revision: B.01 and above

Test Application Revision A.08 and Lab Application C.03, increased Number of Steps range to 0 to 200 (previously it was 0 to 100)

Programming Example

 
OUTPUT 714;"READ:WDPAnalysis?"
! Returns integrity, number of steps measured,
! average power for each step

READ:WEDPower[:ALL]?

Function

Initiates and fetches an extended range dynamic power measurement as a sequential operation.

Measurement results are returned in the following order: integrity indicator, the number of steps measured, and a comma-separated list of average power values for each step measured (in dBm).

See Extended Range Dynamic Power Measurement Results .

Query

Integrity Indicator - See Integrity Indicator .

Number of Steps

  • Range: 1 to 85, 9.91E+37
  • Resolution: 1

Average Power

  • Range: -99 to 99 dBm, 9.91E+37
  • Resolution: 0.01 dBm

Requirements

Test Application Revision: A.10 and above

Lab Application Revision: D.01 and above

Programming Example

 
OUTPUT 714;"READ:WEDPower?"
! Returns integrity, number of steps measured,
! average power for each step
 
OUTPUT 714;"READ:WEDPower:STEPs?"
! Returns number of steps measured

READ:WILPower[:ALL]?

Function

Initiates and fetches an Inner Loop Power control measurement as a sequential operation.

Measurement results are returned in the following order:

Query

Integrity

  • Range: 0 to 23

  • Resolution: 1

Overall pass/fail

  • Range: 0 to 1
  • Resolution: 1

Worst case slot number (based on each adjacent timeslot measurement)

  • Range: 1 to 150, 9.91 E+37

  • Resolution: 1

Absolute power (based on each adjacent timeslot measurement)

  • Range: -99.0 to +99.0 dB, 9.91 E+37

  • Resolution: 0.01

Relative power (based on each adjacent timeslot measurement)

  • Range: -99.0 to +99.0 dB, 9.91 E+37

  • Resolution: 0.01

Worst case slot number (based on Rel Pwr 10 TPC (dB) measurements)

  • Range: 1 to 150, 9.91 E+37

  • Resolution: 1

Absolute power (based on Rel Pwr 10 TPC (dB) measurements)

  • Range: -99.0 to +99.0 dB, 9.91 E+37

  • Resolution: 0.01

Relative power (based on Rel Pwr 10 TPC (dB) measurements)

  • Range: -99.0 to +99.0 dB, 9.91 E+37

  • Resolution: 0.01

Requirements

Test Application Revision: A.03 and above.

Lab Application Revision: A.01 and above.

Programming Example

 
OUTPUT 714;"FETCh:WILPower:ALL?" 

READ:WIQTuning[:ALL]?

Function

Initiates and fetches an IQ tuning measurement as a sequential operation.

Measurement results are returned in the following order: Integrity, RMS EVM, Peak EVM, Frequency Error, Origin Offset, Phase Error, Magnitude Error, Time Error, Peak Code Domain Error, IQ Gain Imbalance, IQ Phase Imbalance.

Query

Integrity

  • Range: 0 to 23

  • Resolution: 1

EVM (RMS)

  • Range: 0.0 to 99.0%, 9.91 E+37

  • Resolution: 0.1

EVM (Peak)

  • Range: 0.0 to 99.0%, 9.91 E+37

  • Resolution: 0.1

Frequency Error

  • Range: -99.0 to 99.0 kHz, 9.91 E+37

  • Resolution: 0.1

Origin Offset

  • Range: -99.0 to 0.0 dB, 9.91 E+37

  • Resolution: 0.1

Phase Error

  • Range: 0 to 180 degrees, 9.91 E+37

  • Resolution: 0.1

Magnitude Error

  • Range: 0.0 to 99.0%, 9.91 E+37

  • Resolution: 0.1

Timing Error

  • Range: -99.99 to 99.99 chips, 9.91 E+37

  • Resolution: 0.01 chips

Peak Code Domain Error

  • Range: -99 to 0 dB, 9.91 E+37

  • Resolution: 0.1

IQ Gain Imbalance

  • Range:

  • Resolution:

IQ Phase Imbalance

  • Range:

  • Resolution:

Requirements

Test Application Revision: A.03 and above.

Lab Application Revision: A.01 and above.

Programming Example

 
OUTPUT 714;"READ:WIQTuning:ALL?" !Queries the IQ tuning measurement results.

READ:WOBWidth[:ALL]?

Function

Initiates and fetches an occupied bandwidth measurement as a sequential operation.

Measurement results are returned in the following order: integrity, average occupied bandwidth.

If the multi-measurement state is on, the result represents the average of a number of channel power measurements. Multi-measurement state, by default, is off. To set up multi-measurements, see SETup:WOBWidth:COUNt[:SNUMber] .

Query
  • Range: 0.0 to 10.0 MHz
  • Resolution:.0001

Requirements

Test Application Revision: A.02 and above.

Lab Application Revision: A.01 and above.

READ:WOOPower[:ALL]?

Function

Initiates and fetches a PRACH transmit on/off power measurement as a sequential operation.

Measurement results are returned in the following order: Integrity, Pre Burst Off Power level, Pre Burst Off Power pass/fail, Burst On Power, Open Loop Power Error value, Open Loop Power Error pass/fail, Post Burst Off Power level, Post Burst Off Power pass/fail (see PRACH Transmit On/Off Power Measurement Results ).

Query

Integrity

  • Range: 0 to 23

  • Resolution: 1

Pre Burst Off Power level

  • Range: -99.99 to 00.00 dBm, 9.91 E+37

  • Resolution: 0.01 dBm

Pre Burst Off Power pass/fail

  • Range: 0 (pass), 1 (fail), 9.91 E+37

Burst On Power

  • Range: -99.99 to 99.99 dBm, 9.91 E+37

  • Resolution: 0.01 dBm

Open Loop Power Error value

  • Range: -99.99 to +99.99 dB
  • Resolution: 0.01 dB

Open Loop Power Error pass/fail

  • Range: 0 (pass), 1 (fail), 9.91 E+37

Post Burst Off Power level

  • Range: -99.99 to 00.00 dBm, 9.91 E+37

  • Resolution: 0.01 dBm

Post Burst Off Power pass/fail

  • Range: 0 (pass), 1 (fail), 9.91 E+37

Requirements

Test Application Revision: A.03 and above.

Lab Application Revision: A.01 and above.

READ:WPDiscon[:ALL]?

Function

Initiates and fetches a phase discontinuity measurement as a sequential operation.

Measurement results are returned as a comma separated list of values: Integrity Indicator, Number of Steps Measured, Worst Case Phase Discontinuity Step Number, Worst Case Phase Discontinuity in degrees, Worst Case EVM Step Number, Worst Case EVM in % rms.

See Phase Discontinuity Measurement Results .

Query

Integrity Indicator - see Integrity Indicator .

Number of Steps Measured

  • Range: 2 to 91, 9.91E+37
  • Resolution: 1

Worst Case Phase Discontinuity Step Number

  • Range: 0 to 90, 9.91E+37
  • Resolution: 1

Worst Case Phase Discontinuity

  • Range: -180.0 to +180.0 degrees, 9.91E+37
  • Resolution: 0.1 degrees

Worst Case EVM Step Number

  • Range: 0 to 90, 9.91E+37
  • Resolution: 1

Worst Case EVM

  • Range: 0.0 to 99.0% rms, 9.91E+37
  • Resolution: 0.1% rms

Requirements

Test Application Revision: A.05 and above

Lab Application Revision: B.01 and above

Programming Example

 
OUTPUT 714;"READ:WPDiscon?"

READ:WPPAnalysis[:ALL]?

Function

Initiates and fetches a PRACH preamble analysis measurement as a sequential operation.

Measurement results are returned in the following order: Integrity Indicator, EVM (rms) in % rms, EVM (peak) in %, Frequency Error in Hz, Origin Offset in dB, Phase Error (rms) in degrees, Magnitude Error in %, Timing Error in chips, IQ Gain Imbalance in dB, IQ Phase Imbalance in degrees, Signature.

See PRACH Preamble Analysis Measurement Results .

Query

Integrity Indicator - see Integrity Indicator .

EVM (rms)

  • Range: 0.0 to 100.0% rms, 9.91E+37
  • Resolution: 0.1% rms

EVM (peak)

  • Range: 0.0 to 100.0%, 9.91E+37
  • Resolution: 0.1%

Frequency Error

  • Range: -9.999 kHz to +9.999 kHz, 9.91E+37
  • Resolution: 0.1 Hz

Origin Offset

  • Range: -100.0 to 0.0 dB, 9.91E+37
  • Resolution: 0.1 dB

Phase Error

  • Range: -180.0 to +180.0 degrees, 9.91E+37
  • Resolution: 0.1 degrees

Magnitude Error

  • Range: 0.0 to 100.0%, 9.91E+37
  • Resolution: 0.1%

Timing Error

  • Range: -383.62 to +383.62 chips, 9.91E+37
  • Resolution: 0.01 chips

IQ Gain Imbalance

  • Range: -100.0 to 100.0 dB, 9.91E+37
  • Resolution: 0.1 dB

IQ Phase Imbalance

  • Range: -180.0 to 180.0 degrees, 9.91E+37
  • Resolution: 0.1 degrees

Signature

  • Range: 0 to 15, 9.91E+37
  • Resolution: 1

Requirements

Lab Application Revision: B.01 and above

Programming Example

 
OUTPUT 714;"READ:WPPAnalysis?"
! Returns Integrity Indicator, EVM (rms) in % rms, EVM (peak) in %,
! Frequency Error in Hz, Origin Offset in dB, Phase Error (rms) in degrees,
! Magnitude Error in %, Timing Error in chips, IQ Gain Imbalance in dB,
! IQ Phase Imbalance in degrees, Signature

READ:WSEMask[:ALL]?

Function

Initiates and fetches a spectrum emissions mask measurement as a sequential operation.

Measurement results are returned in the following order: integrity, overall pass/fail, in-channel power, and pass/fail results for each of the four frequency ranges.

In-channel power is returned in dBm/3.84 MHz.

The four frequency ranges for this measurement are as follows:

  • Range 1: +/- 2.515 to 3.485 MHz
  • Range 2: +/- 4.000 to 7.500 MHz
  • Range 3: +/- 7.500 to 8.500 MHz
  • Range 4: +/- 8.500 to 12.500 MHz

If the multi-measurement state is on, the result represents the average of a number of spectrum emissions measurements. Multi-measurement state, by default, is off. To set up multi-measurements, see SETup:WSEMask:COUNt[:SNUMber] .

Query

Integrity

  • Range: 0 to 23

Overall Pass/Fail

  • Range: 0|1

In-Channel Power (dBm/3.84 MHz)

  • Range: -99.00 to +99.00 dBm, 9.91 E+37
  • Resolution: 0.01

Range Pass/Fail

  • Range: 0|1

Requirements

Test Application Revision: A.02 and above.

Lab Application Revision: A.01 and above.

READ:WSEMask:FAIL?

Function

Initiates and fetches a spectrum emissions mask measurement as a sequential operation.

Measurement results are returned in the following order: integrity, overall pass/fail (0 = Pass, 1 = Fail).

If the multi-measurement state is on, the result represents the average of a number of thermal power measurements. Multi-measurement state, by default, is off. To set up multi-measurements, see SETup:WSEMask:COUNt[:SNUMber] .

Query
  • Range: 0|1

Requirements

Test Application Revision: A.02 and above.

Lab Application Revision: A.01 and above.

READ:WTDPower[:ALL]?

Function

Initiates and fetches a TX dynamic power measurement as a sequential operation.

This query returns 101 comma-separated values, in the order of an integrity value, followed by 100 power results.

This query always returns 101 power results, regardless of how many power measurements were performed by the test set when the measurement was executed. The number of power measurements performed by the test set is equal to the Number of Power Steps setting + 1, and can be queried using the FETCh:WTDPower:COUNt[:STEP]? command. If fewer than 100 power measurements were performed by the test set, the remaining power results are returned as 9.91 E+37.

Query

  • Power

    • Range: -100.00 to 100.00 dBm, 9.91 E+37
    • Resolution: 0.01 dB

Requirements

Test Application Revision: A.07 and above

Lab Application Revision: C.02 and above

READ:WTPower[:ALL]?

Function

Initiates and fetches a thermal power measurement as a sequential operation.

Measurement results are returned in the following order: integrity, thermal transmit power.

If the multi-measurement state is on, the result represents the average of a number of thermal power measurements. Multi-measurement state, by default, is off. To set up multi-measurements, see SETup:WTPower:COUNt[:SNUMber] .

Query
  • Range: -99 to 99 dBm, 9.91 E+37
  • Resolution: 0.01 dB

Requirements

Test or Lab Application Revision: A.01 and above

READ:WWQuality[:ALL]?

Function

Initiates and fetches the waveform quality measurements as a sequential operation.

Measurement results are returned in the following order: integrity, EVM maximum, frequency error worst case, origin offset maximum, phase error maximum, magnitude error maximum.

If the multi-measurement state is on, each result represents the average of a number of waveform quality measurements. Multi-measurement state, by default, is off. To set up multi-measurements, see SETup:WWQuality:COUNt[:SNUMber] .

Query

Integrity

  • Range: 0 to 16

  • Resolution: 1

EVM Maximum

  • Range: 0 to 99%, 9.91 E+37

  • Resolution: 0.1

Frequency Error worst case

  • Range-99.99 to 99.99 kHz, 9.91 E+37

  • Resolution: 0.1

Origin Offset Maximum

  • Range: -99 to 0 dB, 9.91 E+37

  • Resolution: 0.1

Phase Error Maximum

  • Range: 0 to 180 degrees, 9.91 E+37

  • Resolution: 0.1

Magnitude Error Maximum

  • Range: 0 to 99%, 9.91 E+37

  • Resolution: 0.1

Requirements

Test or Lab Application Revision: A.01 and above