SETup:SAUDio
January 16, 2009
GPIB Command Conventions
SETup:SAUDio:CONTinuous
|
Function
|
Sets/queries the trigger arm state for the swept audio measurement. Determines whether measurement will make one measurement then stop (0|OFF (single)), or automatically rearm upon completion of one measurement and repeat the process (1|ON (continuous)).
|
|
Setting
|
Range: 1|ON|0|OFF
|
|
Query
|
Range: 1|0
|
|
*RST Setting
|
0 (off)
|
|
Programming Example
OUTPUT 714;"SETUP:SAUDIO:CONTINUOUS ON" !Turns on continuous triggering.
|
SETup:SAUDio:COUNt[:SNUMber]
|
Function
|
Sets/queries the number of audio measurements to perform at each measurement point in a single multi-measurement sweep. This command sets the count, and automatically sets the multi-measurement STATe command to 1 (on).
|
|
Setting
|
Range: 1 to 999
|
|
Query
|
Range: 1 to 999
|
|
*RST Setting
|
10 |
|
Programming Example
OUTPUT 714;"SETUP:SAUDIO:COUNT:SNUMBER 5" !Turns on multi-measurements and makes 5 measurements at each frequency point in the sweep.
|
SETup:SAUDio:COUNt:NUMBer
|
Function
|
Sets/queries the number of audio measurements to perform at each measurement point in a single multi-measurement sweep when the
SETup:SAUDio:COUNt:STATe
is set to 1 (on).
|
|
Setting
|
Range: 1 to 999
|
|
Query
|
Range: 1 to 999
|
|
*RST Setting
|
10
|
|
Programming Example
OUTPUT 714;"SETUP:SAUDIO:COUNT:NUMBER 5" !Makes 5 measurements at each frequency point in the sweep.
|
SETup:SAUDio:COUNt:STATe
|
Function
|
Selects/queries the swept audio measurement's multi-measurement count state.
|
|
Setting
|
Range: 1|ON|0|OFF
|
|
Query
|
Range: 1|0
|
|
*RST Setting
|
0 (off)
|
|
Programming Example
OUTPUT 714;"SETUP:SAUDIO:COUNT:STATE ON" !Turns on multi-measurement for the swept audio measurement.
|
SETup:SAUDio:COUPling
|
Function
|
Sets/queries the audio generator's coupling to the AUDIO OUT port.
|
|
Setting
|
Range: AC|DC
|
|
Query
|
Range: AC|DC
|
|
*RST Setting
|
DC |
|
Programming Example
OUTPUT 714;"SETUP:SAUDIO:COUPLING AC" !Select AC coupling to the AUDIO OUT port.
|
SETup:SAUDio:DETector[:TYPE]
|
Function
|
Sets/queries the type of level detector used in swept audio measurements.
|
|
Setting
|
Range: RMS|PEAK
|
|
Query
|
Range: RMS|PEAK
|
|
*RST Setting
|
RMS |
|
Programming Example
OUTPUT 714;"SETUP:SAUDIO:DETECTOR:TYPE PEAK" !Select the peak detector for the AF analzyer.
|
SETup:SAUDio:FILTer[:TYPE]
|
Function
|
Sets/queries the type of filter on the sampled data, before any measurements are performed.
The filter choices are: none, TBPass (100 Hz bandwidth bandpass), CMESsage (C-Message), BPASs50 (50 Hz to 15 kHz bandpass), BPASs300 (300 Hz to 15 kHz bandpass)
|
|
Setting
|
Range: NONE|TBPass|CMESsage|BPASs50|BPASs300
|
|
Query
|
Range: NONE|TBP|CMES|BPAS50|BPAS300
|
|
*RST Setting
|
None
|
|
Programming Example
OUTPUT 714;"SETUP:SAUDIO:FILTER:TYPE TBPASS" !Selects the 100 Hz bandpass filter.
|
SETup:SAUDio:FREQuency[:VALue]?
|
Function
|
Returns a comma-separated list of the generated frequency for each frequency point (see
SETup:SAUDio:FREQuency:POINts
).
When you specify the start frequency, stop frequency, and the number of measurement points, the test set calculates the intermediate frequencies that are generated. You cannot directly set the intermediate frequencies using this command, you can only query them.
|
|
Query
|
Range: up to 60 comma separated values
|
|
Programming Example
OUTPUT 714;"SETUP:SAUDIO:FREQUENCY:VALUE?" !Queries the list of generated frequencies.
|
SETup:SAUDio:FREQuency:POINts
|
Function
|
Sets/queries the number of evenly-spaced frequency points in the sweep. If this value is set to 1, only the start frequency is used in the sweep.
|
|
Setting
|
Range: 1 to 60
|
|
Query
|
Range: 1 to 60
|
|
*RST Setting
|
5
|
|
Programming Example
OUTPUT 714;"SETUP:SAUDIO:FREQUENCY:POINTS 20" !Make audio measurements at 20 frequency points in the sweep.
|
SETup:SAUDio:FREQuency:STARt
|
Function
|
Sets/queries the frequency at which to begin the sweep. The default units are Hz.
It is possible to set the start frequency higher than the stop frequency. If this is done, the sweep starts at the higher frequency and stops at the lower frequency.
|
|
Setting
|
Range: 300.0 to 15000.0 Hz
|
|
Query
|
Range: 300.0 to 15000.0 Hz
|
|
*RST Setting
|
300 Hz
|
|
Programming Example
OUTPUT 714;"SETUP:SAUDIO:FREQUENCY:START 450HZ" !Start the frequency sweep at 450 Hz.
|
SETup:SAUDio:FREQuency:STOP
|
Function
|
Sets/queries the frequency at which to end the sweep. The default units are Hz.
It is possible to set the stop frequency lower than the start frequency. If this is done, the sweep starts at the higher frequency and stops at the lower frequency.
|
|
Setting
|
Range: 300.0 to 15000.0 Hz
|
|
Query
|
Range: 300.0 to 15000.0 Hz
|
|
*RST Setting
|
3000 Hz |
|
Programming Example
OUTPUT 714;"SETUP:SAUDIO:FREQUENCY:STOP 1500HZ" !Stop the frequency sweep at 1500 Hz.
|
SETup:SAUDio:ICOunt:MAXimum?
|
Function
|
Queries the total multi-measurement count. This is equal to the number of frequency points in the sweep multiplied by the multi-measurement count set for the swept audio measurement.
|
|
Query
|
Range: 1 to 59940 (60 measurement points multiplied by 999 multi-measurements)
|
|
*RST Setting
|
5 |
|
Programming Example
OUTPUT 714;"SETUP:SAUDIO:ICOUNT:MAXIMUM?" !Queries the total number of measurements made.
|
SETup:SAUDio:PEAK:VOLTage
|
Function
|
Sets/queries the expected peak voltage of the swept audio measurement. This determines the range and resolution of the sampled voltage values.
|
|
Setting
|
Range: 1 mV to 20 V
Resolution: 1 mV
|
|
Query
|
Range: 1 mV to 20 V
|
|
*RST Setting
|
20 V peak
|
|
Programming Example
OUTPUT 714;"SETUP:SAUDIO:PEAK:VOLTAGE 5V"
|
SETup:SAUDio:SETTling[:TIMe]
|
Function
|
Sets/queries how long swept audio measurements wait to start an audio measurement after setting the audio source's frequency. A setting of 0.0 seconds let's swept audio measurements run at maximum speed.
Settling time may be needed when testing a device that has a time delay through the audio path. Path delays can cause the audio generator's signal to change frequencies before the output signal can be analyzed at that frequency. Specifying a settling time ensures that the audio generator stays at each frequency point long enough for the audio analyzer to correctly measure the device's audio output.
|
|
Setting
|
Range: 0.0 to 999 ms
Resolution: 1 ms
|
|
Query
|
Range: 0.0 to 999 ms
|
|
*RST Setting
|
0 ms
|
|
Programming Example
OUTPUT 714;"SETUP:SAUDIO:SETTLING:TIME 20MS"
|
SETup:SAUDio:SDIStortion:STATe
|
Function
|
Sets/queries whether SINAD and distortion measurements are enabled. Turning on SINAD/Distortion measurements will cause the swept audio measurement to run significantly slower because of added sampling and calculation times.
|
|
Setting
|
Range: 1|ON|0|OFF
|
|
Query
|
Range: 1|0
|
|
*RST Setting
|
0 (off)
|
|
Programming Example
OUTPUT 714;"SETUP:SAUDIO:SDISTORTION:STATE ON"
|
SETup:SAUDio:TIMeout[:STIMe]
|
Function
|
Selects/queries the timeout value in seconds that is used for swept audio measurements and sets the timeout state to ON. Units (S|MS) are optional, if no units are specified then units default to S (seconds). |
|
Setting
|
Range: 0.1 to 999 seconds
Resolution: 0.1 seconds
|
|
Query
|
Range: 0.1 to 999 seconds
|
|
*RST Setting
|
10 seconds |
|
Programming Example
OUTPUT 714;"SETUP:SAUDIO:TIMEOUT:STIME 10" !Enables timeouts and sets the value to 10 seconds.
|
SETup:SAUDio:TIMeout:TIME
|
Function
|
Selects/queries the timeout value in seconds that is used for swept audio measurements when the timeout state is set to ON. Units (S|MS) are optional, if no units are specified then units default to S (seconds). |
|
Setting
|
Range: 0.1 to 999 seconds
Resolution: 0.1 seconds
|
|
Query
|
Range: 0.1 to 999 seconds
|
|
*RST Setting
|
10 seconds |
|
Programming Example
OUTPUT 714;"SETUP:SAUDIO:TIMEout:TIME 10" !Sets the timeout value to 10 seconds.
|
SETup:SAUDio:TIMeout:STATe
|
Function
|
Selects/queries the swept audio measurement timeout state. Timeout is set using the
SETup:SAUDio:TIMeout:TIME
command. |
|
Setting
|
Range: 1|ON|0|OFF
|
|
Query
|
Range: 1|0
|
|
*RST Setting
|
0 (off)
|
|
Programming Example
OUTPUT 714;"SETUP:SAUDIO:TIMEOUT:STATE ON" !Enables timeouts for swept audio measurements.
|
SETup:SAUDio:VOLTage:AMPLitude
|
Function
|
Sets/queries the output level (peak voltage) of the audio generator when it is turned on. This command does not turn on the audio generator.
|
|
Setting
|
Range: 0.0 to 9.0 V peak (R
L
> 600 ohms)
Resolution:
-
< 0.5 mV peak (less than or equal to 1 V peak output)
-
< 5 mV peak (greater than 1 V peak output)
|
|
Query
|
Range: 0.0 to 9.0 V peak (R
L
> 600 ohms)
|
|
*RST Setting
|
0.0 V peak
|
|
Programming Example
OUTPUT 714;"SETUP:SAUDIO:VOLTAGE:AMPLITUDE 1V" !Set the audio generator's output to 1Vpeak.
|
|
|