SETup:WOBWidth
Last updated: January 16, 2009
GPIB Command Conventions
SETup:WOBWidth:CONTinuous
|
Function
|
Sets/queries the trigger arm state for the occupied bandwidth measurement. The trigger arm state determines whether the occupied bandwidth measurement will be made once then stop (0|OFF (single)), or automatically re-arm upon completion of one measurement and repeat the process (1|ON (continuous)).
|
|
Setting
|
Range: 1|ON|0|OFF
|
|
Query
|
Range: 1|0
|
|
*RST Setting
|
0 (off)
|
|
Requirements
|
Test Application Revision: A.02 and above.
Lab Application Revision: A.01 and above.
|
|
Programming Example
OUTPUT 714;"SETUP:WOBWIDTH:CONTINUOUS OFF"
!Sets measurement to single trigger mode.
|
SETup:WOBWidth:PERcent:TOTal:POWer
|
Function
|
Sets/queries the percentage of total power parameter used in the occupied bandwidth measurement.
|
|
Setting
|
Range: 70.0 to 99.0 percent
|
|
Query
|
Range: 70.0 to 99.0 percent
|
|
*RST Setting
|
99 percent
|
|
Requirements
|
Test Application Revision: A.03 and above.
Lab Application Revision: A.01 and above.
|
|
Programming Example
OUTPUT 714;"SETup:WOBWidth:PERcent:TOTal:POWer 99"
!Sets percentage of total power to 99%.
|
SETup:WOBWidth:COUNt[:SNUMber]
|
Function
|
The setting form of this command performs two functions: (1) sets the number of occupied bandwidth measurements to be made in a multi-measurement, and (2) sets the multi-measurement STATe command to 1 (on).
The query form returns the number of occupied bandwidth multi-measurements to be made when multi-measurement state is ON.
|
|
Setting
|
Range: 1 to 999
|
|
Query
|
Range: 1 to 999
|
|
*RST Setting
|
Off (10 when STATe = On) |
|
Requirements
|
Test Application Revision: A.02 and above.
Lab Application Revision: A.01 and above.
|
|
Programming Example
OUTPUT 714;"SETUP:WOBWIDTH:COUNT:SNUMBER 5"
!Turns on multi-measurements and sets the number of measurements to 5.
|
SETup:WOBWidth:COUNt:NUMBer
|
Function
|
Sets/queries the number of occupied bandwidth measurements to be made when the multi-measurement state is ON.
|
|
Setting
|
Range: 1 to 999
|
|
Query
|
Range: 1 to 999
|
|
*RST Setting
|
10 |
|
Requirements
|
Test Application Revision: A.02 and above.
Lab Application Revision: A.01 and above.
|
|
Programming Example
OUTPUT 714;"SETUP:WOBWIDTH:COUNT:NUMBER 5"
!Sets the number of occupied bandwidth multi-measurements to 5.
|
SETup:WOBWidth:COUNt:STATe
|
Function
|
Sets/queries the occupied bandwidth multi-measurement state.
|
|
Setting
|
Range: 1|ON|0|OFF
|
|
Query
|
Range: 1|0
|
|
*RST Setting
|
0 (off)
|
|
Requirements
|
Test Application Revision: A.02 and above.
Lab Application Revision: A.01 and above.
|
|
Programming Example
OUTPUT 714;"SETUP:WOBWIDTH:COUNT:STATE ON" !Turns on occupied bandwidth multi-measurements.
|
SETup:WOBWidth:TIMeout[:STIMe]
|
Function
|
The setting form of this command performs two functions: (1) sets the occupied bandwidth measurement's timeout (see
Measurement Timeouts
) value, and (2) sets the occupied bandwidth measurement's timeout STATe to 1 (on).
The query form returns the occupied bandwidth measurement's timeout value.
|
|
Setting
|
Range: 0.1 to 999.9 seconds
Resolution: 0.1 second
|
|
Query
|
Range: 0.1 to 999.9
|
|
*RST Setting
|
10 seconds, off
|
|
Requirements
|
Test Application Revision: A.02 and above.
Lab Application Revision: A.01 and above.
|
|
Programming Example
OUTPUT 714;"SETUP:WOBWIDTH:TIMEOUT:STIME 5 S"
!Sets timeout state to ON and sets the timeout time to 5 seconds.
|
SETup:WOBWidth:TIMeout:STATe
|
Function
|
Sets/queries the occupied bandwidth measurement's timeout (see
Measurement Timeouts
) state.
|
|
Setting
|
Range: 1|ON|0|OFF
|
|
Query
|
Range: 1|0
|
|
*RST Setting
|
0 (off)
|
|
Requirements
|
Test Application Revision: A.02 and above.
Lab Application Revision: A.01 and above.
|
|
Programming Example
OUTPUT 714;"SETUP:WOBWIDTH:TIMEOUT:STATE ON"
|
SETup:WOBWidth:TIMeout:TIME
|
Function
|
Sets/queries the occupied bandwidth measurement's timeout (see
Measurement Timeouts
) setting to be used when the timeout state is set to ON.
|
|
Setting
|
Range: 0.1 to 999.9 seconds
Resolution: 0.1 second
|
|
Query
|
Range: 0.1 to 999.9
|
|
*RST Setting
|
10 seconds
|
|
Requirements
|
Test Application Revision: A.02 and above.
Lab Application Revision: A.01 and above.
|
|
Programming Example
OUTPUT 714;"SETUP:WOBWIDTH:TIMEOUT:TIME 5 S"
|
SETup:WOBWidth:TRIGger:DELay
|
Function
|
Sets/queries the occupied bandwidth measurement's trigger delay setting.
This setting can be used to delay the measurement trigger in order to start a measurement during a specific timeslot in the frame. For example, a 1.333 ms delay would cause the trigger to wait until after the first two timeslots have elapsed.
|
|
Setting
|
Range: -10 to 10 milliseconds
Resolution: 1 nanosecond
|
|
Query
|
Range: -10 to 10 milliseconds
|
|
*RST Setting
|
0 seconds
|
|
Requirements
|
Test Application Revision: A.02 and above.
Lab Application Revision: A.01 and above.
|
|
Programming Example
OUTPUT 714;"SETUP:WOBWIDTH:TRIGGER:DELAY 1MS"
|
SETup:WOBWidth:TRIGger:SOURce
|
Function
|
Sets/queries the occupied bandwidth measurement's trigger source.
|
|
Setting
|
Range: IMMediate|PROTocol|AUTO
|
|
Query
|
Range: IMM|PROT|AUTO
|
|
*RST Setting
|
AUTO |
|
Requirements
|
Test Application Revision: A.02 and above.
Lab Application Revision: A.01 and above.
|
|
Programming Example
OUTPUT 714;"SETUP:WOBWIDTH:TRIGGER:SOURCE IMMEDIATE"
|
|