SETup:WSEMask
Last updated: January 16, 2009
GPIB Command Conventions
SETup:WSEMask:BOUNdary:ADDitional[1]
|
Function
|
Sets/queries the range boundary between range 1 and range 2 for additional SEM requirements.
|
|
Setting
|
Range:
-
KHZ2600 (2.6 MHz)
-
KHZ3500 (3.5 MHz)
|
|
Query
|
Range:
-
KHZ2600 (2.6 MHz)
-
KHZ3500 (3.5 MHz)
|
|
*RST Setting
|
KHZ3500
|
|
Requirements
|
Test Application Revision: A.14 and above.
|
|
Programming Example
OUTPUT 714;"SETup:WSEMask:BOUNdary:ADDitional KHZ3500"
!Sets the range boundary between range 1 and 2 for additional SEM requirements.
|
SETup:WSEMask:BWIDth:ADDitional2
|
Function
|
Sets/queries the measurement bandwidth of the second range of the additional SEM requirements, either 100 kHz or 1 MHz (as defined in TS 25.101 (v8.3.0, 2008-07)).
|
|
Setting
|
Range:
|
|
Query
|
Range:
|
|
*RST Setting
|
KHZ100
|
|
Requirements
|
Test Application Revision: A.14 and above.
|
|
Programming Example
OUTPUT 714;"SETup:WSEMask:BWIDth:ADDitional2 KHZ100"
!Sets the the measurement bandwidth of the second range of the
!additional SEM requirements to 100 kHz.
|
SETup:WSEMask:CONTinuous
|
Function
|
Sets/queries the trigger arm state for the spectrum emission mask measurement. The trigger arm state determines whether the spectrum emission mask measurement will be made once then stop (0|OFF (single)), or automatically re-arm upon completion of one measurement and repeat the process (1|ON (continuous)).
See
Spectrum Emission Mask Measurement Description
.
|
|
Setting
|
Range: 1|ON|0|OFF
|
|
Query
|
Range: 1|0
|
|
*RST Setting
|
0 (off)
|
|
Requirements
|
Test Application Revision: A.02 and above.
Lab Application Revision: A.01 and above.
|
|
Programming Example
OUTPUT 714;"SETUP:WSEMASK:CONTINUOUS OFF"
!Sets measurement to single trigger mode.
|
SETup:WSEMask:COUNt[:SNUMber]
|
Function
|
The setting form of this command performs two functions: (1) sets the number of spectrum emission mask measurements to be made in a multi-measurement, and (2) sets the multi-measurement STATe command to 1 (on).
The query form returns the number of spectrum emission mask multi-measurements to be made when multi-measurement state is ON.
See
Spectrum Emission Mask Measurement Description
.
|
|
Setting
|
Range: 1 to 999
|
|
Query
|
Range: 1 to 999
|
|
*RST Setting
|
Off (10 when STATe = On) |
|
Requirements
|
Test Application Revision: A.02 and above.
Lab Application Revision: A.01 and above.
|
|
Programming Example
OUTPUT 714;"SETUP:WSEMASK:COUNT:SNUMBER 5"
!Turns on multi-measurements and sets the number of measurements to 5.
|
SETup:WSEMask:COUNt:NUMBer
|
Function
|
Sets/queries the number of spectrum emission mask measurements to be made when the multi-measurement state is ON.
See
Spectrum Emission Mask Measurement Description
.
|
|
Setting
|
Range: 1 to 999
|
|
Query
|
Range: 1 to 999
|
|
*RST Setting
|
10 |
|
Requirements
|
Test Application Revision: A.02 and above.
Lab Application Revision: A.01 and above.
|
|
Programming Example
OUTPUT 714;"SETUP:WSEMASK:COUNT:NUMBER 5"
!Sets the number of spectrum emission mask multi-measurements to 5.
|
SETup:WSEMask:COUNt:STATe
|
Function
|
Sets/queries the spectrum emission mask multi-measurement state.
See
Spectrum Emission Mask Measurement Description
.
|
|
Setting
|
Range: 1|ON|0|OFF
|
|
Query
|
Range: 1|0
|
|
*RST Setting
|
0 (off)
|
|
Requirements
|
Test Application Revision: A.02 and above.
Lab Application Revision: A.01 and above.
|
|
Programming Example
OUTPUT 714;"SETUP:WSEMASK:COUNT:STATE ON"
! Turns on spectrum emission mask multi-measurements.
|
SETup:WSEMask:LIMit:ADDitional[1]
|
Function
|
Sets/queries the first additional limit, which is applied to the first range for additional SEM requirements.
|
|
Setting
|
Range: -61.0 dBm to +28 dBm
Resolution: 0.01 dBm
|
|
Query
|
Same as setting
|
|
*RST Setting
|
-15 dBm
|
|
Requirements
|
Test Application Revision: A.14 and above.
|
|
Programming Example
OUTPUT 714;"SETup:WSEMask:LIMit:ADDitional -15DBM"
! Sets the first additional limit to -15 dBm.
|
SETup:WSEMask:LIMit:ADDitional2
|
Function
|
Sets/queries the second additional limit, which is applied to the second range for additional SEM requirements.
|
|
Setting
|
Range: -61.0 dBm to +28 dBm
Resolution: 0.01 dBm
|
|
Query
|
Same as setting
|
|
*RST Setting
|
-13 dBm
|
|
Requirements
|
Test Application Revision: A.14 and above.
|
|
Programming Example
OUTPUT 714;"SETup:WSEMask:LIMit:ADDitional2 -13DBM"
! Sets the first additional limit to -13 dBm.
|
SETup:WSEMask:LIMit:REQuirement
|
Function
|
Sets/queries the spectrum emission mask (SEM) fail limit mask.
When set to BASic (Basic Limit Only), SEM is tested for the basic SEM requirement only, and results are returned only for ranges 1 to 4 of the basic limit.
When set to BADDitional (Basic AND Additional Limit), SEM is tested for both basic and additional SEM requirements. Results are returned for range 1 to 4 and the additional range 1 to 2.
|
|
Setting
|
Range: BASic | BADDitional
|
|
Query
|
Range: BAS | BADD
|
|
*RST Setting
|
BASic
|
|
Requirements
|
Test Application Revision: A.14 and above.
|
|
Programming Example
OUTPUT 714;"SETup:WSEMask:LIMit:REQuirement BASic"
! Sets the spectrum emission mask (SEM) fail limit mask to basic limit only.
|
SETup:WSEMask:TIMeout[:STIMe]
|
Function
|
The setting form of this command performs two functions: (1) sets the spectrum emission mask measurement's timeout (see
Measurement Timeouts
) value, and (2) sets the spectrum emission mask measurement's timeout STATe to 1 (on).
The query form returns the spectrum emission mask measurement's timeout value.
See
Spectrum Emission Mask Measurement Description
.
|
|
Setting
|
Range: 0.1 to 999.9 seconds
Resolution: 0.1 second
|
|
Query
|
Range: 0.1 to 999.9
|
|
*RST Setting
|
10 seconds, off
|
|
Requirements
|
Test Application Revision: A.02 and above.
Lab Application Revision: A.01 and above.
|
|
Programming Example
OUTPUT 714;"SETUP:WSEMASK:TIMEOUT:STIME 5 S"
! Sets timeout state to ON and sets the timeout time to 5 seconds.
|
SETup:WSEMask:TIMeout:STATe
|
Function
|
Sets/queries the spectrum emission mask measurement's timeout (see
Measurement Timeouts
) state.
See
Spectrum Emission Mask Measurement Description
.
|
|
Setting
|
Range: 1|ON|0|OFF
|
|
Query
|
Range: 1|0
|
|
*RST Setting
|
0 (off)
|
|
Requirements
|
Test Application Revision: A.02 and above.
Lab Application Revision: A.01 and above.
|
|
Programming Example
OUTPUT 714;"SETUP:WSEMASK:TIMEOUT:STATE ON"
|
SETup:WSEMask:TIMeout:TIME
|
Function
|
Sets/queries the spectrum emission mask measurement's timeout (see
Measurement Timeouts
) setting to be used when the timeout state is set to ON.
See
Spectrum Emission Mask Measurement Description
.
|
|
Setting
|
Range: 0.1 to 999.9 seconds
Resolution: 0.1 second
|
|
Query
|
Range: 0.1 to 999.9
|
|
*RST Setting
|
10 seconds
|
|
Requirements
|
Test Application Revision: A.02 and above.
Lab Application Revision: A.01 and above.
|
|
Programming Example
OUTPUT 714;"SETUP:WSEMASK:TIMEOUT:TIME 5 S"
|
SETup:WSEMask:TRIGger:ALIGnment:HSDPcchannel:SUBFrame
|
Function
|
Sets/queries the SEM measurement's
HS-DPCCH Trigger Subframe Alignment
setting.
This command is only applicable when
SETup:WSEMask:TRIGger:SOURce
is set to
HSDPcchannel
.
|
|
Setting
|
Range: 0 to 5
Resolution: 1
|
|
Query
|
Same as setting |
|
*RST Setting
|
0 (subframe 0) |
|
Requirements
|
Test Application Revision: A.08 and above with the required feature license
Lab Application Revision: C.03 and above
|
|
Programming Example
OUTPUT 714;"SETup:WSEMask:TRIGger:ALIGnment:HSDPcchannel:SUBFrame 1"
! Sets HS-DPCCH Trigger Subframe Alignment to subframe 1.
|
SETup:WSEMask:TRIGger:DELay
|
Function
|
Sets/queries the spectrum emission mask measurement's
Trigger Delay
setting.
This setting can be used to delay the measurement trigger in order to start a measurement during a specific timeslot in the frame. For example, a 1.333 ms delay would cause the trigger to wait until after the first two timeslots have elapsed.
|
|
Setting
|
Range: -12 to 12 ms
Resolution: 1 ns
|
|
Query
|
Same as setting |
|
*RST Setting
|
0 seconds
|
|
Requirements
|
Test Application Revision: A.02 and above
Test Application Revision A.08 and Lab Application C.03, increased range to -12 to 12 ms (previously it was -10 ms to 10 ms)
Lab Application Revision: A.01 and above
|
|
Programming Example
OUTPUT 714;"SETUP:WSEMASK:TRIGGER:DELAY 1MS"
|
SETup:WSEMask:TRIGger:SOURce
|
Function
|
Sets/queries the spectrum emission mask measurement's
Trigger Source
.
|
|
Setting
|
Range:
-
AUTO
-
EXTernal
-
HSDPcchannel (feature-licensed test application only)
-
IMMediate
-
PROTocol
|
|
Query
|
Range:
-
AUTO
-
EXT
-
HSDP (feature-licensed test application only)
-
IMM
-
PROT
|
|
*RST Setting
|
AUTO |
|
Requirements
|
Test Application Revision: A.02 and above
Test Application Revision A.08 with the required feature license and Lab Application C.03, added HSDPcchannel
Lab Application Revision: A.01 and above
|
|
Programming Example
OUTPUT 714;"SETUP:WSEMASK:TRIGGER:SOURCE IMMEDIATE"
|
SETup:WSEMask:LIMit:BWIDTH30
|
Function
|
Sets/queries the spectrum emission mask measurement's pass/fail limits for the 30 kHz bandwidth filter frequency offsets.
There are two 30 kHz bandwidth filter pass/fail settings, one for the +/-2.5 MHz offset and one for the +/-3.5 MHz offset.
See
Spectrum Emission Mask Measurement Description
.
|
|
Setting
|
Range: -90 to 0 dBc
Resolution: 0.01 dBc
|
|
Query
|
Range: -90 to 0
|
|
*RST Setting
|
-35 dBc (+/-2.5 MHz offset), -50 dBc(+/-3.5 MHz offset) |
|
Requirements
|
Test Application Revision: A.02 and above.
Lab Application Revision: A.01 and above.
|
|
Programming Example
OUTPUT 714;"SETUP:WSEMASK:LIMit:BWIDTH30 -35,-34"
! Sets the +/-2.5 MHz frequency offset pass/fail limit to -35
! and sets the +/-3.5 MHz frequency offset to -34
|
SETup:WSEMask:LIMit:BWIDTH1000
|
Function
|
Sets/queries the spectrum emission mask measurement's pass/fail limits for the 1 MHz bandwidth filter frequency offsets.
There are four 1 MHz bandwidth filter pass/fail settings
-
+/-3.5 MHz offset
-
+/-7.5 MHz offset
-
+/-8.5 MHz offset
-
+/-12.5 MHz offset
See
Spectrum Emission Mask Measurement Description
.
|
|
Setting
|
Range: -90 to 0 dBc
Resolution: 0.01 dBc
|
|
Query
|
Range: -90 to 0
|
|
*RST Setting
|
-35 dBc (+/-3.5 MHz offset), -39 dBc (+/-7.5 MHz offset), -49 dBc (+/-8.5 MHz offset), -49 dBc(+/-12.5 MHz offset) |
|
Requirements
|
Test Application Revision: A.02 and above.
Lab Application Revision: A.01 and above.
|
|
Programming Example
OUTPUT 714;"SETUP:WSEMASK:LIMit:BWIDTH1000 -35,-34,-45,-45"
! Sets the pass/fail limits for the four SEM 1 MHz filter frequency offsets.
|
|
|