Phase Discontinuity Troubleshooting

Last updated: January 16, 2009

Possible Setup Issues

  • In Active Cell Operating Mode , you must be on a call before initiating the measurement.
  • This measurement is not supported on HSPA connections.
  • When on a non-HSDPA call, Trigger Source must be set to Protocol or External . When on an HSDPA call ( lab application or feature-licensed test application only ), Trigger Source must be set to HS-DPCCH .
  • In FDD Test Operating Mode , you must ensure that the UE is synchronized to the downlink frame clock before running the measurement.
  • During front panel operation of this measurement, you must press the START SINGLE key to initiate the measurement.
  • Ensure that you have met the Phase Discontinuity Measurement Input Signal Requirements .
  • Before performing a phase discontinuity measurement, ensure that the measurement has been properly calibrated (see Calibrating the Test Set ).

Other Considerations

  • No other measurements can be running when a phase discontinuity measurement is performed. If any measurements are running when a phase discontinuity measurement is initiated, they are closed and a message is displayed to indicate that they were closed.
  • When Trigger Source is set to HS-DPCCH ( lab application or feature-licensed test application only ), when the measurement is initiated the test set checks to see that UL CL Power Ctrl Algorithm is set to Two and UL CL Power Ctrl Stepsize is set to 1 dB . If not, the test set sends the UE a Physical Channel Reconfiguration message with these values.
  • When Power Control Pattern is set to 5UP/4DOWN , 5DOWN/4UP , 18 UP or 18 DOWN , when the measurement is initiated the test set checks to see that UL CL Power Ctrl Algorithm is set to One and UL CL Power Ctrl Stepsize is set to 1 dB . If not, the test set sends the UE a Physical Channel Reconfiguration message with these values.
  • Absolute measurement results are affected by any amplitude offsets that are entered (relative measurement results are not). See Amplitude Offset .
  • The sum of HS-DPCCH Trigger Subslot Alignment and HS-DPCCH Measurement Step Interval cannot exceed 1.0 slot ( lab application or feature-licensed test application only ).
  • The Phase (in degrees) result is always zero for step 0, the first step measured (the Reference Timeslot for non-HSDPA phase discontinuity).
  • No Phase Discontinuity (in degrees) result is reported for step 0, the first step measured (the Reference Timeslot for non-HSDPA phase discontinuity).
  • When Power Control Pattern is set to 5UP/4DOWN , 5DOWN/4UP , 18 UP or 18 DOWN , when the measurement completes, in an attempt to maintain the connection with the UE, the test set sets UE Target Power to the UE's output power in the final measured slot and sets UL CL Power Control Mode to Active Bits when in Active Cell Operating Mode or to Alternating Bits when in FDD Test Operating Mode .
  • While on a faded channel (see Digital Bus Information ), the measurement will only return valid results if Trigger Source is set to External .