CALCulate:SMASk

Last Updated: June 19, 2009

These commands provide data processing functions for the Spectral Mask (SMAS) measurement.

CALCulate:SMASk:FMASk[?]

Sets/queries whether or not the fast mask is used. Fast mask is a mode for spectral mask that decreases the time taken to perform the measurement by around 50%. If this command is set to ON a value of 108.2 kHz is used for the residual bandwidth setting when making the spectral mask measurement. This gives a much faster measurement but slightly less accurate.

Setting Range: 0 | OFF | 1 | ON

Query Range: 0 | 1

*RST Setting: O | OFF

Test Set Requirements: N8300A Wireless Networking Test Set Version A.02 and above

Example:

CALCulate:SMASk:FMASk ON

Sets the fast mask on for the Spectral Mask measurement.

CALCulate:SMASk:RLEVel[?]

Sets/queries the reference level for the Spectral Mask measurement. The reference level can be set manually using this command or automatically using CALCulate:SMASk:RLEVel:AUTO[?].

Setting Range: -999 to 100 dB

Query Range: -999 to 100 dB

Resolution: 0.00000001

*RST Setting: -999 dB

Test Set Requirements: N8300A Wireless Networking Test Set Version A.02 and above

Example:

CALCulate:SMASk:RLEVel -30

Sets the reference level to -30 dB for the Spectral Mask measurement.

CALCulate:SMASk:RLEVel:AUTO[?]

Set/queries whether or not the reference level is set automatically for the Spectral Mask measurement. If this command is set to:

Setting Range: 0 | OFF | 1 | ON

Query Range: 0 | 1

*RST Setting: 1 | ON

Test Set Requirements: N8300A Wireless Networking Test Set Version A.02 and above

Example:

CALCulate:SMASk:RLEVel:AUTO ON

Sets the reference level to be the peak of the analyzed signal for the Spectral Mask measurement.

CALCulate:SMASk:SPAN[?]

Set/queries the measurement span size used for the Spectral Mask measurement.

Setting Range: SPAN22 | SPAN44 | SPAN66 | SPAN120

Query Range: SPAN22 | SPAN44 | SPAN66 | SPAN120

*RST Setting: SPAN22  

Test Set Requirements: N8300A Wireless Networking Test Set Version A.02 and above

Example:

CALCulate:SMASk:SPAN SPAN66

Sets the measurement span to 66 MHz for the Spectral Mask measurement.

CALCulate:SMASk:WTYPe[?]

Set/queries the measurement windowing type for the Spectral Mask measurement.

Setting Range: GAUSsian | FLATtop | HANNing | UNIForm

Query Range: GAUS | FLAT | HANN | UNIF

*RST Setting: GAUSsian

Test Set Requirements: N8300A Wireless Networking Test Set Version A.02 and above

Example:

CALCulate:SMASk:WTYPe HANNing  

Sets the windowing type to hanning for the Spectral Mask measurement.