Last updated: November 12, 2008
Tx Spurious
Setup (
F1
) key to access the
Tx Spurious Emissions
Setup menu.
Adjacent Limit
and
Alternate Limit
which are available when
Mask Control
is set to
Manual
.)
Multi-Measurement Count
parameter and press the
ON
key or directly enter the number of measurement counts. The number of averages will default to 10.
single
is selected in the
Trigger Arm
field, press the
START SINGLE
key to trigger each measurement. If the
Trigger Arm
is set to
Continuous
the measurement began executing as soon as you initialize the measurement in step 4.
Close Menu
(
F6
) key. The numeric results will display by default.
Graph
(
F3
) key to display a graphical measurement results.Definition: Conducted spurious emissions are emissions at frequencies that are outside the assigned CDMA Channel, measured at the mobile station antenna connector. This test measures the spurious emissions during continuous transmission.
The test set does not support the Enhanced Access Channel tests. The following test procedure is for Access Channel and assumes that the mobile station supports Band Class 1 (US PCS). You should change the settings as specified for other configurations. Unless otherwise stated, leave all parameter settings to the defaults.
Cell Band
(
F8
on the
Call Parms 1 of 3
) to
US PCS
.
Radio Config
(
F11
on the
Call Parms 1 of 3
) to
(Fwd1, Rvs1)
.
(Call Parms 1 of 3)
, set
Service Option for Fwd1, Rvs1
to
SO2 (Loopback)
.
Traffic Data Rate
(
F12
on the
Call Parms 2 of 3
) to
Full
.
Originate Call
(
F3
) to initiate a call. Verify that the
Connected
is in the
Active Cell status
field.
Cell Power
(Î
or
) to
-104 dBm/1.23 MHz
(
F7
on the
Call Parms 1 of 3
).
Rvs Power Ctrl
to
All Up bits
(
F7
on the
Call Parms 2 of 3
menu) to transmit continuous '0' power control bits.
Tx Spurious Emissions
measurement.
Tx Spurious
Setup (
F1
) key to access the
Tx Spurious Emissions
Setup menu. Set the measurement parameters as needed.
single
is selected in the
Trigger Arm
field, press the
START SINGLE
key to trigger each measurement. If the
Trigger Arm
is set to
Continuous
, skip this step.
Graph
(
F3
) key. When viewing the graphical display, you can select
Marker Position
(
F4
) and turn the knob to display the spurious emissions levels at each of the frequency offsets.Tx Spurious Emissions Measurement Description