These messages are intended to be displayed on the manual user interface only, and are not entered into the Error/Event Queue.
MUI200 "Call processing disabled" |
This error, during calibration, is part of the test set's normal operation. If this error occurs during mobile station testing, an internal error has occurred in the test set's firmware. |
MUI210 "Channel power calibration completed successfully" |
This error message is displayed for information purposes only. |
MUI211 "Channel power calibration failed due to temperature drift. Wait for temperature to stabilize" |
This error message occurs when the internal temperature of the test set has drifted more that +/- 10 deg C since the last calibration. Normally, letting the test set fully warm up before calibrating will eliminate this problem. |
MUI212 "Channel power calibration failed due to oven out of range" |
The only user requirement for performing calibration is to make sure the test set is warmed up and to remove RF power to the front panel RF IN/OUT connector. If these requirements have been met and this message appears a test set failure condition exists. |
MUI213 "Channel power calibration failed due to loopback switch problem" | The only user requirement for performing calibration is to make sure the test set is warmed up and to remove RF power to the front panel RF IN/OUT connector. If these requirements have been met and this message appears a test set failure condition exists. |
MUI214 "Channel power calibration failed due to under range condition" |
The only user requirement for performing calibration is to make sure the test set is warmed up and to remove RF power to the front panel RF IN/OUT connector. If these requirements have been met and this message appears a test set failure condition exists. |
MUI215 "Channel power calibration failed due to over range condition" |
The only user requirement for performing calibration is to make sure the test set is warmed up and to remove RF power to the front panel RF IN/OUT connector. If these requirements have been met and this message appears a test set failure condition exists. |
MUI216 "Channel power calibration failed due to correlation problem" |
The only user requirement for performing calibration is to make sure the test set is warmed up and to remove RF power to the front panel RF IN/OUT connector. If these requirements have been met and this message appears a test set failure condition exists. |
The only user requirement for performing calibration is to make sure the test set is warmed up and to remove RF power to the front panel RF IN/OUT connector. If these requirements have been met and this message appears a test set failure condition exists. |
|
MUI218 "Channel power calibration in progress. Do not apply RF power during calibration." |
This error message is displayed for information purposes only. |
MUI219 "Channel power calibration<x>%." |
This error message is displayed for information purposes only. |
MUI220 "Channel power calibration failed due to source level problem." |
This error message is displayed for information purposes only. |
MUI221 "Channel power calibration failed due to IF filter coefficient problem." |
This error message is displayed for information purposes only. |
MUI222 "Channel power calibration failed due to IF frequency response problem." |
This error message is displayed for information purposes only. |
MUI223 "Channel power calibration failed due to waveform quality problem." |
This error message is displayed for information purposes only. |
MUI230 "Digital average power calibration completed successfully. Cycle power to continue." |
This error message is displayed for information purposes only. |
MUI231 "Digital average power calibration failed due to temperature drift. Wait for temperature to stabilize." |
This error message occurs when the internal temperature of the test set has drifted more that +/- 10 deg C since the last calibration. Normally, letting the test set fully warm up before calibrating will eliminate this problem. |
MUI232 "Digital average power calibration failed due to oven out of range |
This error message occurs when the internal temperature of the test set has drifted more that +/- 10 deg C since the last calibration. Normally, letting the test set fully warm up before calibrating will eliminate this problem. |
MUI233 "Digital average power calibration failed due to loopback switch problem |
The only user requirement for performing calibration is to make sure the test set is warmed up and to remove RF power to the front panel RF IN/OUT connector. If these requirements have been met and this message appears a test set failure condition exists. |
MUI234 "Digital average power calibration failed" |
The only user requirement for performing calibration is to make sure the test set is warmed up and to remove RF power to the front panel RF IN/OUT connector. If these requirements have been met and this message appears a test set failure condition exists. |
MUI235 "Digital average power calibration in progress. Do not apply RF power during calibration" |
This error message is displayed for information purposes only. |
MUI236 "Digital average power calibration <x>%." |
This error message is displayed for information purposes only. |
MUI241 "Digital average power measurement has been closed" |
This error message is displayed for information purposes only. |
MUI242 "Waveform quality measurement has been closed" |
This error message is displayed for information purposes only. |
MUI243 "Channel power measurement has been closed" |
This error message is displayed for information purposes only. |
MUI244 "Frame error rate measurement has been closed" |
This error message is displayed for information purposes only. |
MUI245 "Access probe power measurement has been closed" |
This error message is displayed for information purposes only. |
MUI246 "Time response of open loop power control measurement has been closed" |
This error message is displayed for information purposes only. |
MUI247 "Gated power measurement has been closed" |
This error message is displayed for information purposes only. |
MUI248 "Tx spurious emissions measurement has been closed" |
This error message is displayed for information purposes only. |
MUI249 "CDMA measurement warning; Graphical access probe power measurement has been closed." |
This error message is displayed for information purposes only. |
MUI250 "CDMA measurement warning; TX dynamic power measurement has been closed." |
This error message is displayed for information purposes only. |
MUI251 "CDMA measurement warning; Multi-tone audio measurement has been closed." |
This error message is displayed for information purposes only. |
MUI252 CDMA measurement warning; Fast Device Tune measurement has been closed |
|
MUI253 Frequency and Power Sequence tables built successfully |
|
MUI261 "CDMA measurement warning; TX spurious emissions measurement requires MS to transmit in non-gated mode. Verify loopback service option and full traffic data rate" |
Select "Full" in the Traffic Data Rate field before attempting TX Spurious measurements. |
MUI262 "CDMA measurement warning; Digital average power measurement is not accurate for gated MS transmissions in radio configurations greater than 2" |
The test set can accurately measure gated MS transmissions in RC 1 and 2, but not in RC 3, 4 and 5. In RC 3, 4 and 5 it is sometimes impossible for the test set to know whether the MS is transmitting gated or not. So when this warning is seen, the user must decide whether or not their current setup leads to MS gated output or not. See R-FCH Gating . |