Manual User Error Messages

Manual User Error Messages

Last updated: June 20, 2006

Description

These messages are intended to be displayed on the manual user interface only, and are not entered into the Error/Event Queue.

Error Message

Description

The function you requested is not yet available.

The test set does not have this capability.

IQ Calibration completed successfully for modulator <N>

<N> is the IQ modulator number that you are attempting to calibrate, <N> is 1 or 2.

IQ Calibration failed for modulator <N>

<N> is the IQ modulator number that you are attempting to calibrate, <N> is 1 or 2.

The function you requested is not available in this TA.

This function is used in another Test Application.

IQ Calibration for modulator 1 in progress (10 minutes). Call processing disabled.

This error is cleared by either the; IQ Calibration completed successfully for modulator <N>, or IQ Calibration failed for modulator <N>.

IQ Calibration for modulator 2 in progress (6 minutes). Call processing disabled.

This error is cleared by either the; IQ Calibration completed successfully for modulator <N>, or IQ Calibration failed for modulator <N>.

Instrument warning: Audio generator instrument has been closed.

The audio generator instrument was closed automatically by the test set.

Measurement warning: Audio Analyzer instrument has been closed.

The audio analyzer instrument was closed automatically by the test set.

Measurement warning: Analog audio measurement has been closed.

Analog audio measurements have been closed by the test set.

Protocol warning: Setting change has terminated the link with the DUT

Some setting change has caused the call to disconnect.

Warning; Call processing disabled

The call processing functions are not active because the test set is performing calibration operations.

IQ first modulator calibration X%

IQ modulator calibration in progress, X represents the percent complete.

IQ second modulator calibration X%

IQ modulator calibration in progress, X represents the percent complete.

Device Specific MUI Errors

TA2000 MUI Errors

Error Message

Description

MUI100 "Call processing disabled" This error, during calibration, is part of the test set's normal operation.

If this error occurs during mobile station testing, an internal error has occurred in the test set's firmware.

CDMA MUI Errors 

Error Message

Description

MUI200 "Call processing disabled" This error, during calibration, is part of the test set's normal operation.

If this error occurs during mobile station testing, an internal error has occurred in the test set's firmware.

MUI210 "Channel power calibration completed successfully"

This error message is displayed for information purposes only.

MUI211 "Channel power calibration failed due to temperature drift. Wait for temperature to stabilize"

This error message occurs when the internal temperature of the test set has drifted more that +/- 10 deg C since the last calibration. Normally, letting the test set fully warm up before calibrating will eliminate this problem.

MUI212 "Channel power calibration failed due to oven out of range"

The only user requirement for performing calibration is to make sure the test set is warmed up and to remove RF power to the front panel RF IN/OUT connector. If these requirements have been met and this message appears a test set failure condition exists.

MUI213 "Channel power calibration failed due to loopback switch problem" The only user requirement for performing calibration is to make sure the test set is warmed up and to remove RF power to the front panel RF IN/OUT connector. If these requirements have been met and this message appears a test set failure condition exists.

MUI214 "Channel power calibration failed due to under range condition"

The only user requirement for performing calibration is to make sure the test set is warmed up and to remove RF power to the front panel RF IN/OUT connector. If these requirements have been met and this message appears a test set failure condition exists.

MUI215 "Channel power calibration failed due to over range condition"

The only user requirement for performing calibration is to make sure the test set is warmed up and to remove RF power to the front panel RF IN/OUT connector. If these requirements have been met and this message appears a test set failure condition exists.

MUI216 "Channel power calibration failed due to correlation problem"

The only user requirement for performing calibration is to make sure the test set is warmed up and to remove RF power to the front panel RF IN/OUT connector. If these requirements have been met and this message appears a test set failure condition exists.

MUI217 "Channel power calibration failed"

The only user requirement for performing calibration is to make sure the test set is warmed up and to remove RF power to the front panel RF IN/OUT connector. If these requirements have been met and this message appears a test set failure condition exists.

MUI218 "Channel power calibration in progress. Do not apply RF power during calibration."

This error message is displayed for information purposes only.

MUI219 "Channel power calibration<x>%."

This error message is displayed for information purposes only.

MUI220 "Channel power calibration failed due to source level problem."

This error message is displayed for information purposes only.

MUI221 "Channel power calibration failed due to IF filter coefficient problem."

This error message is displayed for information purposes only.

MUI222 "Channel power calibration failed due to IF frequency response problem."

This error message is displayed for information purposes only.

MUI223 "Channel power calibration failed due to waveform quality problem."

This error message is displayed for information purposes only.

MUI230 "Digital average power calibration completed successfully. Cycle power to continue."

This error message is displayed for information purposes only.

MUI231 "Digital average power calibration failed due to temperature drift. Wait for temperature to stabilize."

This error message occurs when the internal temperature of the test set has drifted more that +/- 10 deg C since the last calibration. Normally, letting the test set fully warm up before calibrating will eliminate this problem.

MUI232 "Digital average power calibration failed due to oven out of range

This error message occurs when the internal temperature of the test set has drifted more that +/- 10 deg C since the last calibration. Normally, letting the test set fully warm up before calibrating will eliminate this problem.

MUI233 "Digital average power calibration failed due to loopback switch problem

The only user requirement for performing calibration is to make sure the test set is warmed up and to remove RF power to the front panel RF IN/OUT connector. If these requirements have been met and this message appears a test set failure condition exists.

MUI234 "Digital average power calibration failed"

The only user requirement for performing calibration is to make sure the test set is warmed up and to remove RF power to the front panel RF IN/OUT connector. If these requirements have been met and this message appears a test set failure condition exists.

MUI235 "Digital average power calibration in progress. Do not apply RF power during calibration"

This error message is displayed for information purposes only.

MUI236 "Digital average power calibration <x>%."

This error message is displayed for information purposes only.

MUI241 "Digital average power measurement has been closed"

This error message is displayed for information purposes only.

MUI242 "Waveform quality measurement has been closed"

This error message is displayed for information purposes only.

MUI243 "Channel power measurement has been closed"

This error message is displayed for information purposes only.

MUI244 "Frame error rate measurement has been closed"

This error message is displayed for information purposes only.

MUI245 "Access probe power measurement has been closed"

This error message is displayed for information purposes only.

MUI246 "Time response of open loop power control measurement has been closed"

This error message is displayed for information purposes only.

MUI247 "Gated power measurement has been closed"

This error message is displayed for information purposes only.

MUI248 "Tx spurious emissions measurement has been closed"

This error message is displayed for information purposes only.

MUI249 "CDMA measurement warning; Graphical access probe power measurement has been closed."

This error message is displayed for information purposes only.

MUI250 "CDMA measurement warning; TX dynamic power measurement has been closed."

This error message is displayed for information purposes only.

MUI251 "CDMA measurement warning; Multi-tone audio measurement has been closed."

This error message is displayed for information purposes only.

MUI252 CDMA measurement warning; Fast Device Tune measurement has been closed

MUI253 Frequency and Power Sequence tables built successfully

MUI261 "CDMA measurement warning; TX spurious emissions measurement requires MS to transmit in non-gated mode. Verify loopback service option and full traffic data rate"

Select "Full" in the Traffic Data Rate field before attempting TX Spurious measurements.

MUI262 "CDMA measurement warning; Digital average power measurement is not accurate for gated MS transmissions in radio configurations greater than 2"

The test set can accurately measure gated MS transmissions in RC 1 and 2, but not in RC 3, 4 and 5. In RC 3, 4 and 5 it is sometimes impossible for the test set to know whether the MS is transmitting gated or not. So when this warning is seen, the user must decide whether or not their current setup leads to MS gated output or not.

See R-FCH Gating .

Digital2000 MUI Errors

Error Message

Description

MUI300 "Digital2000 warning; Call processing disabled"

This error, during calibration, is part of the test set's normal operation.

If this error occurs during mobile station testing, an internal error has occurred in the test set's firmware.

MUI341 "Digital2000 warning; Handoff waveform quality measurement has been closed"

This error message is displayed for information purposes only.

MUI342 "Digital2000 warning; Code channel time/phase measurement has been closed"

This error message is displayed for information purposes only.

MUI343 "Digital2000 warning; TDSO frame error rate measurement has been closed"

This error message is displayed for information purposes only.

Digital95 MUI Errors

Error Message

Description

MUI400 "Digital2000 warning; Call processing disabled"

This error, during calibration, is part of the test set's normal operation.

If this error occurs during mobile station testing, an internal error has occurred in the test set's firmware.

Related Topics


Error Message Log

Error Messages