E6701T User Documentation

E6701T User Documentation

Last updated: December 2, 2008

The features and functionality present in the E6701T Special GSM/GPRS/EGPRS Lab Application are identical to those offered in the E6701E GSM/GPRS/EGPRS Lab Application, with the exceptions described in this document. For information on how to use the features that are common to the E6701E and E6701T, see the GSM/GPRS/EGPRS online user's guide located at:

http://wireless.keysight.com/rfcomms/refdocs/gsmgprs/default.php

The E6701T can only be loaded onto an E5515C with Special High Data Rate Hardware (or a test set with the equivalent hardware upgrades). Contact your local Keysight Sales Representative for more information, or visit www.keysight.com/find/8960upgrades.

In a test set with Special High Data Rate Hardware, the standard Protocol Processor is replaced with a High Data Rate Protocol Processor, and the test set's rear panel is modified to include a new LAN 2 PORT , which you must use for sending data to the test set (previously the LAN 1 PORT was used for this purpose).

LAN 2 PORT

The Special High Data Rate Hardware required to run the E6701T includes a modified rear panel with a new LAN 2 PORT . To use the E6701T Data Channels functionality (to establish an Packet-Switched Data connection, for example), you must connect your data server to LAN 2 PORT , rather than to LAN 1 PORT as in the past. LAN 1 PORT is still used for all other LAN operations such as protocol logging, firmware downloads and screen captures.

The LAN 2 PORT connects directly to the test set's High Data Rate protocol processor, as shown in the block diagram below.

You must configure the address of the LAN 2 PORT using the LAN IP Address 2 (Data+) setting (see Configuring the Test Set's LAN for general information about LAN IP addresses). LAN IP Address 2 (Data+) must not be the identical to the LAN IP Address setting, and must be on the same subnet as the LAN IP Address setting. If LAN IP Address or Subnet Mask is changed such that LAN IP Address 2 (Data+) is identical to or is no longer on the same subnet as LAN IP Address , then LAN IP Address 2 (Data+) is set to 0.0.0.0 and a warning is posted.

LAN IP Address 2 (Data+) is accessible from the Instrument Setup menu on the System Config Screen . The GPIB command for LAN IP Address 2 (Data+) is SYSTem:COMMunicate:LAN[:SELF]:ADDRess2<sp><string> . See SYSTem:COMMunicate:GPIB:[:SELF]:ADDRess for range information.

External Device Connection/Two-Cell Test System Considerations

Additional Information