RFANalyzer Subsystem

Last updated: 2014-11-28

For more information on the GPIB Command conventions, see GPIB Command conventions

RFANalyzer:CONTrol:MEASurement:FREQuency:AUTO

Function

Turns on/off the auto Measurement Frequency which determines the measurement downconverter signal path frequency control mode.

ON = automatic

OFF = manual (see RFANalyzer:MANual:MEASurement[:MFRequency])

Setting

Range:0 | OFF | 1 | ON

Query

Range:0 | 1

*RST Setting

1 | ON

Requirements

Test Application Revision: A.01.00 and above

Programming Example

"RFANalyzer:CONTrol:MEASurement:FREQuency:AUTO OFF"

RFANalyzer:CONTrol:POWer:AUTO

Function

Sets/queries power control .

ON = automatic

OFF = manual

Setting

Range:0 | OFF | 1 | ON

Query

Range:0 | 1

*RST Setting

1 |ON

Requirements

Test Application Revision: A.01.00 and above

Programming Example

"RFANalyzer:CONTrol:POWer:AUTO OFF"

RFANalyzer:CW:EXPected:POWer

Function

Sets/queries the expected CW receiver input power. This setting is only used when the operating mode is set to CW.

Setting

Range: -29 dBm to +44 dBm

Query

same as setting

*RST Setting

+13 dBm

Requirements

Test Application Revision: A.01.00 and above

Programming Example

"RFANALYZER:CW:EXPECTED:POWER -20"

RFANalyzer:MANual:MEASurement[:MFRequency]

Function

The setting form of this command performs two functions: (1) sets Measurement Frequency , the tune frequency of the test set's measurement signal path, and (2) selects manual  tune mode for the test set's measurement signal path.

The query form of this command returns the tune frequency of the test set's measurement signal path.

Setting

Range: 292.5 MHz to 2700 MHz

Resolution: 0.1 Hz

Query

same as setting

*RST Setting

896.000 MHz

Requirements

Test Application Revision: A.01.00 and above

Programming Example

"RFANALYZER:MANUAL:MEASUREMENT:MFREQUENCY 896 MHZ"

RFANalyzer:MANual:MEASurement:FREQuency

Function

Sets/queries the frequency of measurement frequency when the frequency mode is manual.

Setting

Range: 292.5 MHz to 2700 MHz

Resolution: 0.1 Hz

Query

same as setting

*RST Setting

896.000 MHz

Requirements

Test Application Revision: A.01.00 and above

Programming Example

"RFANALYZER:MANUAL:MEASUREMENT:FREQUENCY 896 MHZ"

RFANalyzer:MANual:POWer[:SELected]

Function

Sets/queries manual power.

Setting

Range: Test application and system type dependent (approximately-69 to + 31 dBm).

Resolution: 0.01 dBm

Query

same as setting

*RST Setting

0 dBm

Requirements

Test Application Revision: A.01.00 and above

Programming Example

"RFANALYZER:MANUAL:POWER:SELECTED 24 DBM"

RFANalyzer:MANual:POWer:TDSCdma

Function

Sets/queries manual power when TDSCdma is the current system type.

Setting

Range: -70 dBm to +30 dBm actual hardware range with amplitude offset = 0

Resolution: 0.01 dBm

Query

same as setting

*RST Setting

0 dBm

Requirements

Test Application Revision: A.01.00 and above

Programming Example

"RFANalyzer:MANual:POWer:TDSCdma 24 DBM"

RFANalyzer:CONTrol:UPLink:FREQuency:AUTO

Function

Turns on/off the auto Uplink Frequency which determines the test set's demodulation signal path frequency control mode.

ON = automatic

OFF = manual

Setting

Range:0 | OFF | 1 | ON

Query

Range:0 | 1

*RST Setting

1 | ON

Requirements

Test Application Revision: A.01.00 and above

Programming Example

"RFANalyzer:CONTrol:UPLink:FREQuency:AUTO OFF"

RFANalyzer:MANual:UPLink[:MFRequency]

Function

The setting form of this command performs two functions: (1) sets uplink Frequency , the tune frequency of the test set's demodulation signal path, and (2) selects manual  tune mode for the test set's demodulation signal path.

The query form of this command returns the tune frequency of the test set's demodulation signal path.

Setting

Range: 292.5 MHz to 2700 MHz

Resolution: 0.1 Hz

Query

same as setting

*RST Setting

896 MHz

Requirements

Test Application Revision: A.01.00 and above

Programming Example

"RFANALYZER:MANUAL:UPLINK:MFREQUENCY 896 MHZ"

RFANalyzer:MANual:UPLink:FREQuency

Function

Sets/queries the frequency of uplink frequency when the frequency mode is manual.

Setting

Range: 292.5 MHz to 2700 MHz

Resolution: 0.1 Hz

Query

same as setting

*RST Setting

896 MHz

Requirements

Test Application Revision: A.01.00 and above

Programming Example

"RFANALYZER:MANUAL:UPLINK:FREQUENCY 896 MHZ"