SETup:CTXSpurious

SETup:CTXSpurious

Last updated: November 17, 2008

GPIB Command Conventions

SETup:CTXSpurious:ALTernate:OFFSet:TEST3

Function

Sets/queries the alternate offset of the test 3 since the outer emission powers alternate limit are changed with the offset when the Test Case Setup is set to Test 3.

See TX Spurious Emissions Measurement Parameters , "Test3 Alternate Offset/Test3 Second Alternate Offset"

Requirements Test Application Revision: A.09 and above

Setting

Range: 2.7 MHz to 3.47 MHz

Resolution: 0.01 MHz

Query

Range: Same as setting.

*RST Setting

3.47

Programming Example

 
OUTPUT 714;"SETup:CTXSpurious:ALTernate:OFFSet:TEST3 3.47" 
!Sets alternate offset of the test 3 to 3.47 MHz.

SETup:CTXSpurious:ALTernate2:OFFSet:TEST3

Function

Sets/queries the second alternate offset of the test 3 since the outer emission powers alternate limit are changed with the offset when the Test Case Setup is set to Test 3.

See TX Spurious Emissions Measurement Parameters , "Test3 Alternate Offset/Test3 Second Alternate Offset"

Requirements Lab Application Revision: C.00.00 and above

Setting

Range: 3.50 MHz to 6.50 MHzs

Resolution: 0.01 MHz

Query

Range: Same as setting.

*RST Setting

6.50 MHz

Programming Example

 
OUTPUT 714;"SETup:CTXSpurious:ALTernate2:OFFSet:TEST3 3.5" 
!Sets the second alternate offset of the test 3 to 3.5 MHz.

SETup:CTXSpurious:CONTinuous

Function

Sets/queries the trigger arm state for the tx spurious emissions measurement. The trigger arm state determines whether the tx spurious emissions measurement will be made once then stop (0|OFF (single)), or automatically re-arm upon completion of one measurement and repeat the process (1|ON (continuous)).

See TX Spurious Emissions Measurement Parameters , "Trigger Arm."

Requirements Test Application Revision: A.01.20 and above
Lab Application Revision: A.01 and above

Setting

Range: 1|ON|0|OFF

Query

Range: 1|0

*RST Setting

0 (off)

Programming Example

 
OUTPUT 714;"SETup:CTXSpurious:CONTinuous OFF" 
!Sets measurement to single trigger mode.

SETup:CTXSpurious:COUNt[:SNUMber]

Function

The setting form of this command performs two functions: (1) sets the number of tx spurious emissions measurements to be made in a multi-measurement, and (2) sets the multi-measurement STATe command to 1 (on).

The query form returns the number of tx spurious emissions multi-measurements to be made when multi-measurement state is ON.

See TX Spurious Emissions Measurement Parameters , "Multi-Measurement Count."

Requirements Test Application Revision: A.01.20 and above
Lab Application Revision: A.01 and above

Setting

Range: 1 to 999

Query

Range: 1 to 999

*RST Setting

Off (10 when STATe = On)

Programming Example

 
OUTPUT 714;"SETup:CTXSpurious:COUNt 5" 
!Turns on multi-measurements and sets the number of measurements to 5.

SETup:CTXSpurious:COUNt:NUMBer

Function

Sets/queries the number of tx spurious emissions measurements to be made when the multi-measurement state is ON.

See TX Spurious Emissions Measurement Parameters , "Multi-Measurement Count."

Requirements Test Application Revision: A.01.20 and above
Lab Application Revision: A.01 and above

Setting

Range: 1 to 999

Query

Range: 1 to 999

*RST Setting

10

Programming Example

 
OUTPUT 714;"SETup:CTXSpurious:COUNt:NUMBer 5" 
!Sets the number of tx spurious emissions multi-measurements to 5.

SETup:CTXSpurious:COUNt:STATe

Function

Sets/queries the tx spurious emissions multi-measurement state.

See TX Spurious Emissions Measurement Parameters , "Multi-Measurement Count."

Requirements Test Application Revision: A.01.20 and above
Lab Application Revision: A.01 and above

Setting

Range: 1|ON|0|OFF

Query

Range: 1|0

*RST Setting

0 (off)

Programming Example

 
OUTPUT 714;"SETup:CTXSpurious:COUNt:STATe ON"
! Turns on tx spurious emissions multi-measurements.

SETup:CTXSpurious:MASK:AUTO

Function

Sets/queries the mask control.

ON (1) sets the mask limits to the values specified in C.S0033. OFF (0) allows you to set arbitrary limits (see SETup:CTXSpurious:MASK:MANual:ADJacent[:LIMit] and SETup:CTXSpurious:MASK:MANual:ALTernate[:LIMit] ).

See TX Spurious Emissions Measurement Parameters , "Mask Control."

Requirements Test Application Revision: A.01.20 and above
Lab Application Revision: A.01 and above

Setting

Range: 1|ON|0|OFF

Query

Range: 1|0

*RST Setting

1 (ON)

Programming Example

 
OUTPUT 714;"SETup:CTXSpurious:MASK:AUTO ON" 

SETup:CTXSpurious:MASK:MANual:ADJacent[:LIMit]

Function

Sets/queries the tx spurious emissions manually set measurement limit mask. This limit is activated when SETup:CTXSpurious:MASK:AUTO is set to OFF (0).

See TX Spurious Emissions Measurement Parameters , "Adjacent Limit."

Requirements Test Application Revision: A.01.20 and above
Lab Application Revision: A.01 and above

Setting

Range: -10.00 to -65.00 dBc

Resolution: 0.01 dB

Query

Range/Resolution: Same as setting.

*RST Setting

-42 dBc

Programming Example

 
OUTPUT 714;"SETup:CTXSpurious:MASK:MANual:ADJacent -45" 

SETup:CTXSpurious:MASK:MANual:ALTernate[:LIMit]

Function

Sets/queries the tx spurious emissions manually set measurement limit mask. This limit is activated when SETup:CTXSpurious:MASK:AUTO is set to OFF (0).

See TX Spurious Emissions Measurement Parameters , "Alternate Limit."

Requirements Test Application Revision: A.01.20 and above
Lab Application Revision: A.01 and above

Setting

Range: -10.00 to -65.00 dBc

Resolution: 0.01 dB

Query

Range/Resolution: Same as setting.

*RST Setting

-50 dBc

Programming Example

 
OUTPUT 714;"SETup:CTXSpurious:MASK:MANual:ALTernate -55".

SETup:CTXSpurious:MASK:MANual:ADJacent[:LIMit]:TEST2

Function

Sets/queries the tx spurious emissions manually set measurement limit mask of the test case 2 . This limit is activated when SETup:CTXSpurious:TCASe is set to Test 2 and SETup:CTXSpurious:MASK:AUTO is set to OFF (0).

See TX Spurious Emissions Measurement Parameters , "TestX Adjacent Limit."

Requirements Test Application Revision: A.09 and above

Setting

Range: -40 to 20 dBm

Resolution: 0.01 dBm

Query

Range/Resolution: Same as setting.

*RST Setting

6 dBm/1MHz

Programming Example

 
OUTPUT 714;"SETup:CTXSpurious:MASK:MANual:ADJacent:TEST2 6"
! Sets the Tx Spurious Emissions measurement limit mask of the test case 2 to 6 dBm/1MHz

SETup:CTXSpurious:MASK:MANual:ALTernate[:LIMit]:TEST2

Function

Sets/queries the tx spurious emissions manually set measurement limit mask of the test case 2 . This limit is activated when SETup:CTXSpurious:TCASe is set to Test 2 and SETup:CTXSpurious:MASK:AUTO is set to OFF (0).

See TX Spurious Emissions Measurement Parameters , "TestX Alternate Limit."

Requirements Test Application Revision: A.09 and above

Setting

Range: -40 to 20 dBm

Resolution: 0.01 dBm

Query

Range/Resolution: Same as setting.

*RST Setting

-13 dBm/1MHz

Programming Example

 
OUTPUT 714;"SETup:CTXSpurious:MASK:MANual:ADJacent:TEST2 -13"
! Sets the Tx Spurious Emissions measurement limit mask of the test case 2 to -13 dBm/1MHz

SETup:CTXSpurious:MASK:MANual:ADJacent[:LIMit]:TEST3

Function

Sets/queries the tx spurious emissions manually set measurement limit mask of the test case 3 . This limit is activated when SETup:CTXSpurious:TCASe is set to Test 3 and SETup:CTXSpurious:MASK:AUTO is set to OFF (0).

See TX Spurious Emissions Measurement Parameters , "TestX Adjacent Limit."

Requirements Test Application Revision: A.09 and above

Setting

Range: -40 to 20 dBm

Resolution: 0.01 dBm

Query

Range/Resolution: Same as setting.

*RST Setting

-14 dBm/30kHz

Programming Example

 
OUTPUT 714;"SETup:CTXSpurious:MASK:MANual:ADJacent:TEST3 6"
! Sets the Tx Spurious Emissions measurement limit mask of the test case 3 to 6

SETup:CTXSpurious:MASK:MANual:ALTernate[:LIMit]:TEST3

Function

Sets/queries the tx spurious emissions manually set measurement limit mask of the test case 3 . This limit is activated when SETup:CTXSpurious:TCASe is set to Test 3 and SETup:CTXSpurious:MASK:AUTO is set to OFF (0).

See TX Spurious Emissions Measurement Parameters , "TestX Alternate Limit."

Requirements Test Application Revision: A.09 and above

Setting

Range: -40 to 20 dBm

Resolution: 0.01 dBm

Query

Range/Resolution: Same as setting.

*RST Setting

-25.55 dBm/30kHz

Programming Example

 
OUTPUT 714;"SETup:CTXSpurious:MASK:MANual:ADJacent:TEST3 -13"
! Sets the Tx Spurious Emissions measurement limit mask of the test case 3 to -13

SETup:CTXSpurious:MASK:MANual:ALTernate2[:LIMit]:TEST3

Function

Sets/queries the second manually set measurement limit mask of the test case 3 of the tx spurious emissions measurement. This limit is activated when SETup:CTXSpurious:TCASe is set to Test 3 and SETup:CTXSpurious:MASK:AUTO is set to OFF (0).

See TX Spurious Emissions Measurement Parameters , "Test3 Second Alternate Limit."

Requirements Lab Application Revision: C.00.00 and above

Setting

Range: -40 to 20 dBm

Resolution: 0.01 dBm

Query

Range/Resolution: Same as setting.

*RST Setting

-25.55 dBm/30kHz

Programming Example

 
OUTPUT 714;"SETup:CTXSpurious:MASK:MANual:ADJacent:TEST3 -13"
! Sets the Tx Spurious Emissions measurement limit mask of the test case 3 to -13

SETup:CTXSpurious:TCASe

Function

Sets/queries the tx spurious emissions test case that will be used.

See TX Spurious Emissions Measurement Parameters , "Test Case Setup"

Requirements Test Application Revision: A.09 and above

Setting

TEST1 | TEST2 | TEST3

Query

Range: Same as setting.

*RST Setting

TEST1

Programming Example

 
OUTPUT 714;"SETup:CTXSpurious:TCASe TEST1"
! Sets the Tx Spurious Emissions test case to Test 1

SETup:CTXSpurious:TIMeout[:STIMe]

Function

The setting form of this command performs two functions: (1) sets the tx spurious emissions measurement's timeout (see Measurement Timeouts ) value, and (2) sets the tx spurious emissions measurement's timeout STATe to 1 (on).

The query form returns the tx spurious emissions measurement's timeout value.

See TX Spurious Emissions Measurement Parameters , "Timeout Time/State."

Requirements Test Application Revision: A.01.20 and above
Lab Application Revision: A.01 and above

Setting

Range: 0.1 to 999.9 seconds

Resolution: 0.1 second

Query

Range: 0.1 to 999.9

*RST Setting

10 seconds, on

Programming Example

 
OUTPUT 714;"SETup:CTXSpurious:TIMeout 5" 
!Sets timeout state to ON and sets the timeout time to 5 seconds.

SETup:CTXSpurious:TIMeout:STATe

Function

Sets/queries the tx spurious emissions measurement's timeout (see Measurement Timeouts ) state.

See TX Spurious Emissions Measurement Parameters , "Timeout Time/State."

Requirements Test Application Revision: A.01.20 and above
Lab Application Revision: A.01 and above

Setting

Range: 1|ON|0|OFF

Query

Range: 1|0

*RST Setting

0 (off)

Programming Example

 
OUTPUT 714;"SETup:CTXSpurious:TIMeout:STATe ON"

SETup:CTXSpurious:TIMeout:TIME

Function

Sets/queries the tx spurious emissions measurement's timeout (see Measurement Timeouts ) setting to be used when the timeout state is set to ON.

See TX Spurious Emissions Measurement Parameters , "Timeout Time/State."

Requirements Test Application Revision: A.01.20 and above
Lab Application Revision: A.01 and above

Setting

Range: 0.1 to 999.9 seconds

Resolution: 0.1 second

Query

Range: 0.1 to 999.9

*RST Setting

10 seconds

Programming Example

 
OUTPUT 714;"SETup:CTXSpurious:TIMeout:TIME 5"