This command sets/queries the Burst Capture Range for TX power measurement. See Burst Capture Range | ||
Setting | ||
SING | ALL | ||
*RST Setting | SINGle | |
Requirements | ||
OUTPUT 714;"SETup:TXPower:BURSt:CAPTure SINGle" ! Sets Burst Capture Range to SINGle. |
This command sets/queries the number of TX power measurements made by the test set. This command sets the count state to ON. See Multi-measurement Count |
||
Setting | ||
*RST Setting | 10 | |
Requirements | ||
OUTPUT 714;"SETup:TXPower:COUNt:SNUMber 99" ! Sets the state to on and the multi-measurement count value to 99. |
This command sets/queries the TX power multi-measurement count state. |
||
Setting | ||
*RST Setting | ||
Requirements | ||
OUTPUT 714;"SETup:TXPower:COUNt:STATe ON" ! Sets the multi-measurement count state to on. |
This command is only applicable when the Burst Capture Range is set to All. This command sets/queries the Estimated Carrier Power State. See Estimated Carrier Power |
||
Setting | Range: 0 | OFF | 1 | ON | |
*RST Setting | ||
Requirements | ||
OUTPUT 714;"SETup:TXPower:ECPower:STATe ON" ! Sets the Estimated Carrier Power state to on. |
This command sets/queries the frame qualification for TX power measurements when the operating mode is not Fast Device Tune. See Frame Qualification . |
||
GSM/GPRS LA | ||
Setting | ||
*RST Setting | 0 | |
Requirements | ||
OUTPUT 714;"SETup:TXPower:FRAMe:QUALifier ON" ! Sets TX Power meas Frame Qualification state to on. |
This command sets/queries the measurement results of which Uplink Sequence Step are returned when the FETCH GPIB commands other than FETCh:TXPower:TSEQuence and FETCh:TXP:SSTep are used. This parameter is only applicable when the operating mode is set to Fast Device Tune. See Fast Device Tune Result Set | ||
Setting |
Range: 1 to 50 Resolution: 1 |
|
Range: 1 to 50 Resolution: 1 |
||
*RST Setting | 1 | |
Requirements |
Revision: A.08 and above, with the required feature license. |
|
OUTPUT 714;"SETup:TXPower:GFDTune:RSET 10" ! The results of USS 10 will be returned. |
This command sets/queries the threshold for the first RF trigger in the Fast Device Tune operating mode. See Fast Device Tune Trigger Threshold | ||
Setting |
Range: 10 dB to 35 dB Resolution: 1 dB |
|
Range: 10 dB to 35 dB Resolution: 1 dB |
||
*RST Setting | 35 dB | |
Requirements |
Revision: A.09 and above, with the required feature license. |
|
OUTPUT 714;"SETup:TXPower:GFDTune:THReshold 10" ! Sets the Trigger Threshold to 10dB in the Fast Device Tune operating mode. |
This command is only applicable when the Burst Capture Range is set to All. This command set/queries whether the instrument ranges the hardware to the highest burst or not. See Auto Range to Highest Burst |
||
GSM/GPRS LA | ||
Setting | ||
*RST Setting | ||
Requirements | ||
OUTPUT 714;"SETup:TXPower:RANGe:AUTO ON" ! Sets Auto-Ranging state to on. |
This command queries the number of uplink TDMA frames required in an uplink sequence step. This parameter is only applicable when the operating mode is set to Fast Device Tune. See Uplink Frames per Step . | ||
GSM/GPRS LA | ||
*RST Setting | -1 | |
Requirements |
Revision: A.08 and above, with the required feature license. |
|
OUTPUT 714;"SETup:TXPower:SSTep FRAMes? ! Queries the uplink frames per sequence step in FDT mode. |
This command sets/queries the TX power measurement timeout value. This command also sets the count state to on. The units (S|MS) are optional, if no units are specified then units default to S. See Measurement Timeout . |
||
GSM/GPRS LA | ||
Setting | ||
*RST Setting | ||
Requirements | ||
OUTPUT 714;"SETup:TXPower:TIMeout:STIMe 20" ! Sets the state to on and the timeout value to 20 seconds. |
This command sets/queries the timeout value in seconds that will be used for TX power measurements in the active (that is the selected) format. The units (S|MS) are optional, if no units are specified then units default to S. See Measurement Timeout . |
||
Setting | ||
*RST Setting | ||
Requirements | ||
OUTPUT 714;"SETup:TXPower:TIMeout:TIME 20" ! Sets the TX power measurement timeout to 20 seconds. |
This command selects/queries TX power measurement timeout state in the active (that is the selected) format. See Measurement Timeout . |
||
Setting | ||
*RST Setting | ||
Requirements | ||
OUTPUT 714;"SETup:TXPower:TIMeout:STATe 1" ! Turns the TX power timeout state on. |
This command is only applicable when the Burst Capture Range is set to All. This command sets/queries the trigger qualification for TX power measurements in the active (that is the selected) format. When ON, an automatic trigger re-arm occurs if a measurement is triggered when no valid signal (burst) is present. See Trigger Qualification Description . |
||
GSM/GPRS LA | ||
Setting | ||
*RST Setting | ||
Requirements | ||
OUTPUT 714;"SETup:TXPower:TRIGger:QUALifier ON" ! Sets trigger qualifier state to on. |
This command selects/queries the trigger source for TX power measurements in the active (that is the selected) format. See Triggering of Measurements . |
||
Setting | ||
*RST Setting | ||
Requirements | ||
OUTPUT 714;"SETup:TXPower:TRIGger:SOURce AUTO" ! Sets trigger source to automatic. |