CALL:CLPControl

Last updated: 2014-11-28

For more information on the GPIB Command conventions, see GPIB Command conventions

CALL[:CELL]:CLPControl:DOWNlink:DPCHannel:LEVel:MAXimum

Function

Sets/queries Maximum DPCH Level.

Setting

Range:-20.00 dB to 0 dB

Resolution:0.01

Increment: 0.1

Query

Same as setting

*RST Setting

0dB

Requirements

Test Application Revision: A.08.00 and above

Programming Example

"CALL:CELL:CLPCONTROL:DOWNlINK:DPCHANNEL:LEVEL:MAXIMUM -18"

CALL[:CELL]:CLPControl:DOWNlink:DPCHannel:LEVel:MINimum

Function

Sets/queries Minimum DPCH  Level.

Setting

Range:-20.00 dB to 0 dB

Resolution:0.01

Increment: 0.1

Query

Same as setting

*RST Setting

-20.00dB

Requirements

Test Application Revision: A.08.00 and above

Programming Example

"CALL:CELL:CLPCONTROL:DOWNlINK:DPCHANNEL:LEVEL:MINIMUM -19"

CALL[:CELL]:CLPControl:DOWNlink:DPCHannel:LEVel:RECord:COUNt

Function

Sets/queries the DPCH Level Sequence Record Count. This setting can be changed when DPCH power offset sequence record is not started.

Setting

Range: 1  to 30000

Resolution:1

Increment: 1

Query

Same as setting

*RST Setting

1000

Requirements

Test Application Revision: A.08.00 and above

Programming Example

"CALL:CELL:CLPCONTROL:DOWNlINK:DPCHANNEL:LEVEL:RECORD:COUNT 1000"

CALL[:CELL]:CLPControl:DOWNlink:DPCHannel:LEVel:RECord:STARt

Function

This action starts the recording of DPCH power offset sequence.

Requirements

Test Application Revision: A.08.00 and above

Programming Example

"CALL:CELL:CLPCONTROL:DOWNlINK:DPCHANNEL:LEVEL:RECORD:START"

CALL[:CELL]:CLPControl:DOWNlink:DPCHannel:LEVel:RECord:STOP

Function

This action stops the recording of DPCH power offset sequence.

Requirements

Test Application Revision: A.08.00 and above

Programming Example

"CALL:CELL:CLPCONTROL:DOWNlINK:DPCHANNEL:LEVEL:RECORD:STOP"

CALL[:CELL]:CLPControl:DOWNlink:DTCHannel:TARGet:QUALity

Function

Sets/queries the Target Quality  on DTCH. This setting can be changed in Idle state  in Active Cell operating mode or Downlink Closed  Loop  Power  Control  State is set to Off.

Setting

Range: 0.01% to 100%

Resolution:0.01%

Increment: 0.01%

Query

Same as setting

*RST Setting

0.01%

Requirements

Test Application Revision: A.08.00 and above

Programming Example

"CALL:CELL:CLPCONTROL:DOWNlINK:DTCHANNEL:TARGET:QUALITY 0.01"

CALL[:CELL]:CLPControl:DOWNlink:DTCHannel:TARGet:QUALity:STATe

Function

Sets/queries whether the BLER target is to be assigned to a DTCH. This setting can be changed in Idle state  in Active Cell operating mode or Downlink Closed  Loop  Power  Control  State is set to Off.

Setting

Range: 1|ON|0|OFF

Query

Same as setting

*RST Setting

0 (Off)

Requirements

Test Application Revision: A.08.00 and above

Programming Example

"CALL:CELL:CLPCONTROL:DOWNlINK:DTCHANNEL:TARGET:QUALITY:STATE 0"

CALL[:CELL]:CLPControl:DOWNlink:STATe

Function

Sets/queries Downlink Closed  Loop  Power  Control  State.

Setting

Range: ON|OFF

Query

Same as setting

*RST Setting

OFF

Requirements

Test Application Revision: A.08.00 and above

Programming Example

"CALL:CELL:CLPCONTROL:DOWNlINK:STATE ON"

CALL[:CELL]:CLPControl:UPLink:IMMediate:DOWN

Function

This action sends an additional Down TPC command to the UE to decrease its power by one step.

This operation is only allowed when the UL CL Power Ctrl Mode is set to Alternating bits.

Requirements

Test Application Revision: A.02.00 and above

Programming Example

"CALL:CELL:CLPCONTROL:UPLINK:IMMEDIATE:DOWN"

CALL[:CELL]:CLPControl:UPLink:IMMediate:UP

Function

This action sends an additional Up TPC command to the UE to increase its power by one step.

This operation is only allowed when the UL CL Power Ctrl Mode is set to Alternating bits.

Requirements

Test Application Revision: A.02.00 and above

Programming Example

"CALL:CELL:CLPCONTROL:UPLINK:IMMEDIATE:UP"

CALL[:CELL]:CLPControl:UPLink:MODE  

Function

Sets/queries the uplink closed power control mode when in the active cell operating mode. See UL CL Power Ctrl Mode.

Setting

Range: ACTive | ALTernating| UP | DOWN

Query

Range: ACT | ALT| UP | DOWN

*RST Setting

ACT

Requirements

Test Application Revision: A.02.00 and above

Programming Example

"CALL:CELL:CLPControl:UPLink:MODE ACTive"

CALL[:CELL]:CLPControl:UPLink:STEPsize  

Function

Sets/queries UL CL Power Ctrl Step Size when in the active cell operating mode.

Setting

Range: 1|2|3 dB

Query

Range: 1|2|3

*RST Setting

1

Requirements

Test Application Revision: A.02.00 and above

Programming Example

"CALL:CLPControl:UPLink:STEPsize 2"