Queries the TX spurious emissions measurement results.

This measurement returns results in the following order: integrity, TX spurious emissions pass/fail result, lower adjacent channel (-0.885 or -1.25 MHz) average TX spurious emissions, upper adjacent channel (+0.885 or +1.25 MHz) average TX spurious emissions, lower alternate channel (-1.98 MHz) average TX spurious emissions, upper alternate channel (+1.98 MHz) average TX spurious emissions.

See TX Spurious Emissions Measurement Results .

Requirements

Test Application Revision: A.01.20 and above
Lab Application Revision: A.01 and above

Query

Integrity

TX spurious emissions pass/fail result

  • Range: 0 to 1 (0 = passed, 1 = failed), 9.91 E+37

Lower adjacent channel spurious emissions

  • Range: 0.00 to -80.00 dBc, 9.91 E+37
  • Resolution: 0.01 dBc

Upper adjacent channel spurious emissions

  • Range: 0.00 to -80.00 dBc, 9.91 E+37
  • Resolution: 0.01 dBc

Lower alternate channel spurious emissions

  • Range: 0.00 to -80.00 dBc, 9.91 E+37
  • Resolution: 0.01 dBc

Upper alternate channel spurious emissions

  • Range: 0.00 to -80.00 dBc, 9.91 E+37
  • Resolution: 0.01 dBc

Programming Example

 
OUTPUT 714;"FETCh:CTXSpurious?" 

FETCh:CTXSpurious:EPOWer:ALTernate2:TEST3?

Function

Queries the lower second alternate result and upper second alternate result for the tx spurious emissions measurement test case 3. See TX Spurious Emissions Measurement Results .

Requirements

Lab Application Revision: C.00.00 and above

Query

The lower second alternate result

  • Range: -100dBm to 100dBm, 9.91 E+37

The upper second alternate result

  • Range: -100dBm to 100dBm, 9.91 E+37

Programming Example

 
OUTPUT 714;"FETCh:CTXSpurious:EPOWer:ALTernate2:TEST3?" 

FETCh:CTXSpurious:ICOunt?

Function

Queries the intermediate count of measurements completed in a multi-measurement.

The multi-measurement state, by default, is off. To set up multi-measurements, see SETup:CTXSpurious:COUNt[:SNUMber] .

Requirements

Test Application Revision: A.01.20 and above
Lab Application Revision: A.01 and above

Query

Range: 0 to 999

Programming Example

 
OUTPUT 714;"FETCh:CTXSpurious:ICOunt?" 

FETCh:CTXSpurious:MASK[:FAIL]?

Function Queries the pass/fail results. Pass/fail results are based on fixed limit lines on display. See TX Spurious Emissions Measurement Results .

Requirements

Test Application Revision: A.01.20 and above
Lab Application Revision: A.01 and above

Query

Range: 0 (Pass)|1(Fail), 9.91 E+37

Programming Example

 
OUTPUT 714;"FETCh:CTXSpurious:MASK?"