Configuring Cell 2 Used for Soft/Softer Handoff

 Last updated: October 28, 2004

 This section is only applicable to the lab application.

 

 Cell 2 Overview

 Supporting soft and softer handoffs in the test set requires simulation of a second CDMA base station simulator (Cell 2). Cell 2 is implemented with only the channels (see Generated Code Channel Information for Cell 1 and Cell 2 ) required to exercise the soft and softer handoff functionality of a CDMA mobile station. For more Cell 2 features in soft/softer handoff, see Cell 1 and Cell 2 Used in Soft/Softer Handoff .

 

 Generated Code Channel Information for Cell 1 and Cell 2

 Cell 1 Code Channels

 Cell 2 Code Channels

 IS-2000 System

 IS-95 System

 IS-2000 System

 IS-95 System

 F-Pilot

F-Pilot F-Pilot F-Pilot

 F-Sync

F-Sync F-FCH/Traffic F-Traffic

 F-Paging

F-Paging F-OCNS F-OCNS

 F-QPCH

F-Traffic    

 F-FCH/Traffic

 F-OCNS

   

 F-SCH

     

 F-OCNS

     
 

 Configuring Cell 2 Power

 Setting Cell 2 Power, such as power range, rules, current level and desired level) are similar to the Cell Power (cell 1) settings. The test set's total RF output power is a combination of the power of cell 1, cell 2 and AWGN power:
Total RF Power (dBm/1.23 MHz) = [Cell 1 Power + Cell 2 Power + AWGN Power] (dBm/1.23 MHz)
(see Configuring Cell Power and AWGN Power Levels for more information about cell power and AWGN power)

 

Front Panel Usage

 For how to change Cell 2 power to verify the mobile station soft/softer handoff behavior, see .

 

 Programming Usage

 The test set includes commands to set the desired Cell 2 Power. You may also query the desired power level settings.

 The test set also includes commands to query the current level of cell 2 power. This CALL:STATus command is query only. It returns a value indicating the test set's actual hardware settings and hardware status. Current hardware settings can differ from the desired level you've set, most frequently due to an error condition. When no error conditions exist, the commands used to set the desired level will return the same results as the commands used to query current level. For example, "CALL:CELL2:POWer?" will return the same value as "CALL:STATus:CELL2:POWer?" when no error conditions exist.

 

 Generated Power Level Information

 Parameter

Current Level Desired Level
Cell 2 Power

 CALL:STATus:CELL2:POWer:[AMPLitude][:SELected]?

 CALL:CELL2:POWer

 

 Configuring Cell 2 PN Offset and Code Channels

 Setting Cell 2 code channel levels (such as current level, desired level, OCNS calculation) are similar to those for Cell 1. See Setting Code Channel Levels and Walsh Codes .

 

Front Panel Usage

 To set the Cell 2 code channels from the front panel, see How Do I Change Generator Information? .

 

Programming Usage

 The following table contains a list of GPIB commands controlling cell 2 code channels.

GPIB Commands for Setting/Querying Cell 2 Parameters

 Parameters

Command Purpose

 Cell 2 PN Offset

 CALL:CELL2:PNOFfset

 Sets/queries the Cell 2 PN (pseudonoise) offset.

 Cell 2 F-Pilot Level (relative to cell)

CALL:CELL2:PILot Sets Cell 2 F-Pilot Level.

 

Cell 2 F-Traffic Level (IS-95 system)

CALL:CELL2:TRAFfic Sets Cell 2 Traffic Level (IS-95).

 Cell 2 F-FCH/Traffic Level (IS-2000 system)

CALL:CELL2:FCHannel Sets Cell 2 F-FCH/Traffic Level (IS-2000).

 Cell 2 Delay

CALL:CELL2:DELay Sets Cell 2 Delay (baseband channel relative to Cell 1).
Cell 2 OCNS Level CALL:CELL2:OCNSource

 The Cell 2 F-OCNS/OCNS level is a calculated value and can only be queried.

 The F-OCNS value is calculated to provide a summing value that results in a total code channel value equalling 100% of cell 2 power.

 

 CALL:STATus

Queries current level and states of Cell 2 Power, Cell 2 F-Pilot code channel, Cell 2 F-FCH/Traffic code channel and Cell 2 OCNS. Queries soft/softer handoff state.
 

 Configuring Cell 2 Closed Loop Power Control

 The Cell 2 closed loop power control, together with cell 1 closed loop power control, is used to verify that the mobile station responds correctly to more than one power control subchannel during soft/softer handoffs. See for details.

 The closed loop power control bit patterns for Cell 2 (Cell 2 Rvs CLPC) are:

 The Cell 2 Rvs CLPC (cell 2 reverse closed loop power control) bit patterns can be set using the CALL:CELL2:CLPControl commands:

 

Related Topics


 

Configuring Cell Power and AWGN Power Levels

 Configuring Forward Channel Power Levels and Walsh Codes